{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76810",
      "recalling_firm": "Philips Electronics North  America Corporation",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide, worldwide including Canada",
      "recall_number": "Z-1821-2017",
      "product_description": "UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR Table 722034; UNIQ FD20/10 OR Table 722036; UNIQ FD20/20 722038; UNIQ FD20/20 biplane OR Table 722039",
      "product_quantity": "6992 (total for all devices)",
      "reason_for_recall": "Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.",
      "recall_initiation_date": "20170322",
      "center_classification_date": "20170414",
      "termination_date": "20210517",
      "report_date": "20170426",
      "code_info": "Devices with a Laird Chiller."
    }
  ]
}