  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76911",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 100",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Only - one location each in AL, CO, KS, NV, NY, and 2 in TX",
      "recall_number": "Z-1819-2017",
      "product_description": "da Vinci Xi EndoWrist Suction Irrigator, 8 mm instrument;  PN 480299-03; and   SOFTWARE, EMBEDDED RLS, IS4000, A70_P6_B440; PN 610092-440.      General and Plastic Surgery: The EndoWrist¿ Suction Irrigator is designed to be used in conjunction with an Intuitive Surgical da Vinci Surgical System and compatible suction and irrigation sources and tubing sets for delivering fluid to the surgical site and for evacuation and aspiration of fluids. The instrument may also be used for retraction and blunt dissection of tissue. The instrument tip is blunt and intended to contact tissue.",
      "product_quantity": "15 x 6 packs (90 units)",
      "reason_for_recall": "Intuitive Surgical has become aware that in specific scenarios with system software P6 and the da Vinci Xi  Suction Irrigator (PN 480299-03), users can experience unexpected motion of a system arm.",
      "recall_initiation_date": "20170331",
      "center_classification_date": "20170413",
      "termination_date": "20170728",
      "report_date": "20170419",
      "code_info": "All",
      "more_code_info": ""
    }
  ]
}