  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74038",
      "recalling_firm": "Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) and globally to Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecaudor, The Netherlands, Japan, Korea, Malaysia, Mexico, Peru.",
      "recall_number": "Z-1819-2016",
      "product_description": "Various trauma and sports medicine instruments and implants.  Rod, fixation, intramedullary and accessories.",
      "product_quantity": "N/A",
      "reason_for_recall": "Insufficient seal strength on sterile barrier packaging.  The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile.  If the product becomes non-sterile and is used or implanted, it may lead to an infection.",
      "recall_initiation_date": "20160401",
      "center_classification_date": "20160526",
      "termination_date": "20170523",
      "report_date": "20160601",
      "code_info": "Part #'s  28234  28338  28811  28813  28821  29260  29822  805065080  14-400428  14-400436  14-405032  14-405040  14-440118  14-440121  14-441280  14-440066S    Lot #'s  590970  591080  591090  665210  692040  692250  692280  692290  695900  610470  665230  665240  692440  692450  692480  692620  692660  692670  692690  721680  721720  721730  729080  722730  621840  621860  678300  664740  664820  664870  664880  503760  503770  693840  476230"
    }
  ]
}