  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "61926",
      "recalling_firm": "Philips Healthcare Inc.",
      "address_1": "3000 Minuteman Road",
      "address_2": "N/A",
      "postal_code": "01810",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and the countries of:  Nationwide  Foreign: Canada  AFGHANISTAN  ALBANIA  ALGERIA  ANGOLA  ARGENTINA  ARMENIA  ARUBA  AUSTRALIA  AUSTRIA  AZERBAIJAN  BAHAMAS  BAHRAIN  BANGLADESH  BARBADOS  BELARUS  BELGIUM  BENIN  BHUTAN  BOLIVIA  BOSNIA&HERZEGOVINA  BRAZIL  BRUNEI DARUSSALAM  BULGARIA  BURUNDI  CAMBODIA  CAMEROON  CANARY ISLANDS  CHAD  CHILE  CHINA  COLOMBIA  CONGO  CONGO,DEMOCRATIC  COSTA RICA  COTE D'IVOIRE  CROATIA  CUBA  CYPRUS  CZECH REPUBLIC  DENMARK  DJIBOUTI  DOMINICAN REPUBLIC  ECUADOR  EGYPT  EL SALVADOR  EQUATORIAL GUINEA  ERITREA  ESTONIA  ETHIOPIA  fake test ssd  FINLAND  FRANCE  FRENCH GUIANA  GABON  GAMBIA  GEORGIE  GERMANY  GHANA  GREECE  GUADELOUPE  GUAM  GUATEMALA  GUINEA  GUYANA  HONDURAS  HONG KONG  HUNGARY  ICELAND  INDIA  INDONESIA  IRAN  IRAQ  IRELAND  ISRAEL  ITALY  JAMAICA  JAPAN  JORDAN  KAZAKSTAN  KENYA  KUWAIT  KYRGYZSTAN  LAO PEOPLE'S DEM.REP  LATVIA  LEBANON  LIBERIA  LIBYAN ARAB JAMAHIRI  LIECHTENSTEIN  LITHUANIA  LUXEMBOURG  MACAU  MACEDONIA  MALAWI  MALAYSIA  MALI  MALTA  MARTINIQUE  MAURITANIA  MAURITIUS  METRACOM  MEXICO  MOLDOVA  MONGOLIA  MOROCCO  MOZAMBIQUE  NAMIBIA  NEPAL  NETHERLANDS  NETHERLANDS ANTILLES  NEW CALEDONIA  NEW ZEALAND  NICARAGUA  NIGERIA  NORTH KOREA  NORWAY  OMAN  PAKISTAN  PALESTINE  PANAMA  PARAGUAY  PERU  PHC BEST  PHILIPPINES  PMCC BU Canada  PMCC BU USA  POLAND  PORTUGAL  PUERTO RICO  QATAR  REUNION  ROMANIA  RUSSIA  RWANDA  SAUDI ARABIA  SENEGAL  SERBIA&MONTE NEGRO  SIERRA LEONE  SINGAPORE  SLOVAKIA  SLOVENIA  SOMALIA  SOUTH AFRICA  SOUTH KOREA  SPAIN  SRI LANKA  SUDAN  SURINAME  SWEDEN  SWITZERLAND  SYRIAN ARAB REPUBLIC  TAIWAN  TAJIKISTAN  TANZANIA  THAILAND  TRINIDAD AND TOBAGO  TUNISIA  TURKEY  TURKMENISTAN  UGANDA  UKRAINE  Unidentified Country  UNITED ARAB EMIRATES  UNITED KINGDOM   URUGUAY  UZBEKISTAN  VENEZUELA  VIET NAM  YEMEN  ZAMBIA  ZIMBABWE",
      "recall_number": "Z-1818-2012",
      "product_description": "Philips Digital Diagnost Software, software release 2.0.2 and 2.0.2SP1 including mirror icon    Stationary Fluoroscopic system",
      "product_quantity": "752 units",
      "reason_for_recall": "With Digital Diagnost (DIDI) R 2.0.2 a mirrored \"R in a circle\" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored \"R in a circle\" appears in the lower left corner. This can be mistaken for a \"right patient side\" marker, although this can appear on the left patient side",
      "recall_initiation_date": "20120522",
      "center_classification_date": "20120615",
      "termination_date": "20170224",
      "report_date": "20120627",
      "code_info": "Software release 2.0.2 and 2.0.2SP1 including mirror icon"
    }
  ]
}