  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90809",
      "recalling_firm": "Baxter Healthcare Corporation",
      "address_1": "1 Baxter Pkwy",
      "address_2": "N/A",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution and government/military distribution.",
      "recall_number": "Z-1817-2022",
      "product_description": "Starling Monitor, Product code CMMST5 and Starling Monitor SV, Product code CMM-ST5 (same products);  and accessories/spare parts Starling Battery Pack, Product Code CMABATT; and Starling Patient Cable, Product code CMASC10.",
      "product_quantity": "9,671 units",
      "reason_for_recall": "Labeling has incorrect or incomplete Unique Device Identification (UDI) information; and the barcodes are not compliant with GS-1 standards.",
      "recall_initiation_date": "20220902",
      "center_classification_date": "20220926",
      "report_date": "20221005",
      "code_info": "All serial numbers after 6/2016.  Product code CMMST5 - Incorrect UDI-DI: 7290013030269; correct UDI-DI: 07290013030269.  Product code CMM-ST5 - Incorrect UDI-DI: 7290013030030; correct UDI-DI: 07290013030030.  Product code CMABATT - Incorrect UDI-DI: 7290013030160; correct UDI-DI: 07290013030160.  Product code CMASC10 - Incorrect UDI-DI: 7290013030054; correct UDI-DI: 07290013030054."
    }
  ]
}