  "results": [
    {
      "status": "Terminated",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87887",
      "recalling_firm": "Alphatec Spine, Inc.",
      "address_1": "1950 Camino Vida Roble",
      "address_2": "N/A",
      "postal_code": "92008-6505",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution in the states of FL, CA, IL, OK, TX, CT, IN, NC, WA,  HI, ID, MA, SC, NM, and NY.",
      "recall_number": "Z-1816-2021",
      "product_description": "Atec Insignia Anterior Cervical Plate System, REF 136-0350 Insignia, ACP, 3-Level, 50 mm, Rx Only, Non-Sterile, UDI: (01)00190376268569",
      "product_quantity": "N/A",
      "reason_for_recall": "There is a potential for the screw blocking mechanism on the  anterior cervical plate system  to disassociate intraoperatively/postoperatively or unlock postoperatively.",
      "recall_initiation_date": "20210426",
      "center_classification_date": "20210607",
      "termination_date": "20221117",
      "report_date": "20210616",
      "code_info": "8609322R"
    }
  ]
}