  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Saint Paul",
      "address_1": "1265 Grey Fox Rd",
      "reason_for_recall": "Smiths Medical has become aware of an issue with an intermittent occurrence of binding of the locking assembly on some CADD¿-Solis pumps. Binding of the Cassette/ Keypad Lock can occur after latching the CADD¿-Solis Medication Cassette Reservoir or Administration Set \u001cdisposable\u001d to the pump. When binding occurs, it can prevent the key from fully rotating the Cassette/ Keypad Lock to the locked (e",
      "address_2": "",
      "product_quantity": "4426 (2427 US, 1999 OUS)",
      "code_info": "Serial Number US  1061044  1061058  1061072  1061075  1061082  1061083  1061093  1061104  1061105  1061106  1061107  1061108  1061109  1061110  1061111  1061112  1061113  1061114  1061115  1061116  1061117  1061118  1061119  1061120  1061121  1061122  1061123  1061124  1061125  1061126  1061127  1061128  1061129  1061130  1061131  1061132  1061133  1061134  1061135  1061136  1061137  1061138  1061139  1061140  1061141  1061142  1061143  1061144  1061145  1061146  1061147  1061148  1061149  1061150  1061151  1061152  1061153  1061154  1061155  1061156  1061157  1061158  1061159  1061160  1061161  1061162  1061163  1061164  1061165  1061166  1061167  1061168  1061169  1061170  1061171  1061172  1061173  1061174  1061175  1061176  1061177  1061178  1061179  1061180  1061181  1061182  1061183  1061184  1061185  1061186  1061187  1061188  1061189  1061190  1061191  1061192  1061193  1061194  1061195  1061196  1061197  1061198  1061199  1061200  1061201  1061202  1061203  1061204  1061205  1061206  1061207  1061208  1061209  1061210  1061211  1061212  1061213  1061214  1061215  1061216  1061217  1061218  1061219  1061220  1061221  1061222  1061223  1061224  1061225  1061226  1061227  1061228  1061229  1061230  1061231  1061232  1061233  1061234  1061235  1061236  1061243  1061244  1061245  1061246  1061247  1061248  1061249  1061250  1061251  1061261  1061262  1061263  1061264  1061265  1061266  1061267  1061268  1061269  1061270  1061271  1061272  1061274  1061275  1061276  1061277  1061278  1061279  1061280  1061281  1061282  1061283  1061284  1061285  1061286  1061287  1061288  1061289  1061290  1061291  1061292  1061293  1061294  1061295  1061296  1061297  1061298  1061299  1061311  1061313  1061314  1061315  1061316  1061317  1061318  1061319  1061320  1061321  1061322  1061323  1061324  1061325  1061326  1061327  1061328  1061329  1061330  1061331  1061332  1061333  1061334  1061335  1061336  1061337  1061338  1061339  1061340  1061341  1061342  1061343  1061344  1061345  1061346  1061347  1061348  1061349  1061350  1061351  1061352  1061353  1061354  1061355  1061356  1061357  1061358  1061359  1061360  1061361  1061362  1061363  1061364  1061365  1061366  1061367  1061368  1061369  1061370  1061371  1061372  1061373  1061374  1061375  1061376  1061377  1061378  1061379  1061380  1061381  1061382  1061383  1061384  1061385  1061386  1061387  1061388  1061389  1061390  1061391  1061392  1061393  1061394  1061396  1061397  1061398  1061399  1061400  1061401  1061402  1061403  1061404  1061405  1061406  1061407  1061408  1061409  1061410  1061411  1061412  1061413  1061414  1061415  1061416  1061417  1061418  1061419  1061420  1061421  1061422  1061423  1061424  1061425  1061426  1061427  1061428  1061451  1061462  1061482  1061507  1061509  1061522  1061530  1061544  1061546  1061549  1061552  1061564  1061565  1061566  1061567  1061568  1061569  1061570  1061571  1061572  1061573  1061574  1061575  1061576  1061577  1061578  1061579  1061580  1061581  1061582  1061583  1061584  1061585  1061586  1061587  1061588  1061589  1061590  1061591  1061592  1061593  1061594  1061595  1061596  1061597  1061598  1061599  1061600  1061601  1061602  1061603  1061604  1061605  1061606  1061607  1061608  1061609  1061610  1061611  1061612  1061613  1061614  1061615  1061616  1061617  1061618  1061619  1061620  1061621  1061622  1061623  1061624  1061625  1061626  1061627  1061628  1061629  1061630  1061631  1061632  1061633  1061634  1061635  1061636  1061637  1061638  1061639  1061640  1061641  1061642  1061643  1061644  1061645  1061646  1061647  1061648  1061649  1061650  1061651  1061652  1061653  1061654  1061655  1061656  1061657  1061658  1061659  1061660  1061661  1061662  1061663  1061664  1061665  1061666  1061667  1061668  1061669  1061670  1061671  1061672  1061673  1061674  1061675  1061676  1061677  1061678  1061679  1061680  1061681  1061682  1061683  1061684  1061685  1061686  1061687  1061688  1061689  1061690  1061691  1061692  1061693  1061694  1061695  1061696  1061697  1061698  1061699  1061700  1061701  1061702  1061703  1061704  1061705  1061706  1061707  1061708  1061709  1061710  1061711  1061712  1061713  1061714  1061715  1061716  1061717  1061718  1061719  1061720  1061721  1061722  1061723  1061724  1061725  1061726  1061727  1061728  1061729  1061730  1061731  1061732  1061733  1061734  1061735  1061736  1061737  1061738  1061739  1061740  1061741  1061742  1061753  1061768  1061770  1061772  1061787  1061793  1061794  1061795  1061796  1061797  1061798  1061799  1061800  1061801  1061802  1061803  1061804  1061805  1061806  1061807  1061808  1061809  1061810  1061811  1061812  1061813  1061814  1061815  1061816  1061817  1061818  1061819  1061820  1061821  1061822  1061823  1061824  1061825  1061826  1061827  1061828  1061829  1061830  1061831  1061832  1061833  1061834  1061835  1061836  1061837  1061838  1061839  1061840  1061841  1061842  1061843  1061844  1061845  1061846  1061847  1061848  1061849  1061850  1061851  1061852  1061853  1061854  1061855  1061856  1061857  1061858  1061859  1061860  1061861  1061862  1061863  1061864  1061865  1061866  1061867  1061868  1061869  1061872  1061874  1061891  1061917  1061918  1061924  1061925  1061926  1061927  1061928  1061929  1061930  1061931  1061932  1061933  1061934  1061935  1061936  1061937  1061938  1061939  1061940  1061941  1061942  1061943  1061944  1061945  1061946  1061947  1061948  1061949  1061950  1061951  1061952  1061953  1061954  1061955  1061956  1061957  1061958  1061959  1061960  1061961  1061962  1061963  1061964  1061965  1061966  1061967  1061968  1061969  1061970  1061971  1061972  1061973  1061974  1061975  1061976  1061977  1061978  1061979  1061980  1061981  1061982  1061983  1061984  1061985  1061986  1061987  1061988  1061989  1061990  1061991  1061992  1061993  1061994  1061995  1061996  1061997  1061998  1061999  1062000  1062001  1062002  1062003  1062004  1062005  1062006  1062007  1062008  1062009  1062010  1062011  1062012  1062013  1062014  1062015  1062016  1062017  1062018  1062019  1062020  1062021  1062022  1062023  1062024  1062025  1062026  1062027  1062028  1062029  1062030  1062031  1062032  1062033  1062034  1062035  1062036  1062037  1062038  1062039  1062040  1062041  1062042  1062043  1062044  1062045  1062046  1062047  1062048  1062049  1062050  1062051  1062052  1062053  1062054  1062055  1062056  1062057  1062058  1062059  1062060  1062061  1062062  1062063  1062064  1062065  1062066  1062067  1062068  1062069  1062070  1062071  1062072  1062073  1062074  1062075  1062076  1062077  1062078  1062079  1062080  1062081  1062082  1062083  1062084  1062085  1062086  1062087  1062088  1062089  1062090  1062091  1062092  1062093  1062094  1062095  1062096  1062097  1062098  1062099  1062100  1062101  1062102  1062103  1062104  1062105  1062106  1062107  1062108  1062109  1062110  1062111  1062112  1062113  1062114  1062115  1062116  1062117  1062118  1062119  1062120  1062121  1062122  1062123  1062124  1062125  1062126  1062127  1062128  1062129  1062130  1062131  1062132  1062133  1062134  1062135  1062136  1062137  1062138  1062139  1062140  1062141  1062142  1062143  1062144  1062145  1062146  1062147  1062148  1062149  1062150  1062151  1062152  1062153  1062154  1062155  1062156  1062157  1062158  1062159  1062160  1062161  1062162  1062163  1062164  1062165  1062166  1062167  1062168  1062169  1062170  1062171  1062172  1062173  1062174  1062175  1062176  1062177  1062178  1062179  1062180  1062181  1062182  1062183  1062184  1062185  1062186  1062187  1062188  1062189  1062190  1062191  1062192  1062193  1062194  1062195  1062196  1062197  1062198  1062199  1062200  1062202  1062203  1062204  1062206  1062207  1062208  1062209  1062210  1062211  1062212  1062213  1062214  1062215  1062216  1062218  1062219  1062220  1062221  1062223  1062224  1062225  1062226  1062227  1062228  1062230  1062231  1062232  1062233  1062234  1062235  1062236  1062237  1062238  1062239  1062240  1062241  1062242  1062243  1062244  1062245  1062246  1062247  1062248  1062249  1062250  1062251  1062252  1062253  1062254  1062255  1062256  1062257  1062258  1062259  1062260  1062261  1062262  1062263  1062264  1062265  1062266  1062267  1062268  1062269  1062271  1062272  1062273  1062274  1062275  1062276  1062277  1062278  1062279  1062280  1062281  1062282  1062283  1062284  1062285  1062286  1062287  1062288  1062289  1062290  1062291  1062292  1062293  1062294  1062295  1062296  1062297  1062298  1062299  1062300  1062301  1062302  1062303  1062304  1062305  1062306  1062307  1062313  1062320  1062323  1062340  1062341  1062342  1062343  1062344  1062345  1062346  1062347  1062348  1062349  1062350  1062351  1062352  1062353  1062354  1062355  1062356  1062357  1062358  1062359  1062360  1062361  1062362  1062363  1062364  1062365  1062366  1062367  1062368  1062369  1062370  1062371  1062372  1062373  1062374  1062375  1062376  1062377  1062378  1062379  1062380  1062381  1062382  1062383  1062384  1062385  1062386  1062387  1062388  1062389  1062390  1062391  1062392  1062393  1062394  1062395  1062396  1062397  1062399  1062400  1062401  1062402  1062403  1062404  1062405  1062406  1062407  1062409  1062410  1062411  1062412  1062413  1062414  1062415  1062416  1062419  1062420  1062421  1062423  1062424  1062425  1062426  1062427  1062428  1062429  1062430  1062431  1062432  1062433  1062434  1062435  1062436  1062437  1062438  1062439  1062440  1062441  1062442  1062443  1062444  1062445  1062446  1062447  1062448  1062449  1062450  1062451  1062457  1062458  1062459  1062460  1062461  1062462  1062463  1062464  1062465  1062466  1062467  1062468  1062469  1062470  1062471  1062472  1062473  1062474  1062475  1062476  1062477  1062478  1062479  1062480  1062481  1062482  1062483  1062484  1062485  1062486  1062487  1062488  1062489  1062490  1062491  1062492  1062493  1062494  1062495  1062496  1062497  1062525  1062529  1062553  1062565  1062573  1062644  1062667  1062679  1062680  1062681  1062682  1062683  1062684  1062685  1062686  1062687  1062688  1062689  1062690  1062691  1062692  1062700  1062714  1062715  1062741  1062775  1062785  1062800  1062818  1062820  1062821  1062822  1062823  1062824  1062826  1062827  1062828  1062829  1062830  1062831  1062832  1062833  1062834  1062835  1062836  1062837  1062840  1062841  1062878  1062879  1062880  1062882  1062883  1062884  1062885  1062886  1062887  1062888  1062889  1062890  1062891  1062892  1062893  1062894  1062895  1062896  1062897  1062898  1062900  1062901  1062902  1062903  1062904  1062905  1062906  1062907  1062908  1062909  1062910  1062911  1062912  1062913  1062914  1062915  1062916  1062917  1062918  1062919  1062920  1062921  1062922  1062923  1062924  1062925  1062926  1062928  1062929  1062930  1062931  1062932  1062933  1062934  1062935  1062936  1062937  1062938  1062939  1062940  1062941  1062944  1062945  1062946  1062947  1062948  1062949  1062950  1062951  1062952  1062953  1062954  1062955  1062956  1062957  1062958  1062959  1062960  1062962  1062963  1062964  1062965  1062966  1062967  1062968  1062969  1062970  1062971  1062972  1062973  1062974  1062975  1062977  1062978  1062979  1062980  1062981  1062982  1062983  1062984  1062985  1062987  1062988  1062989  1062990  1062991  1062992  1062993  1062998  1063017  1063034  1063035  1063036  1063037  1063038  1063039  1063040  1063041  1063042  1063043  1063044  1063045  1063046  1063047  1063048  1063049  1063050  1063051  1063052  1063053  1063054  1063055  1063056  1063057  1063058  1063059  1063060  1063061  1063062  1063064  1063065  1063066  1063067  1063068  1063069  1063070  1063071  1063072  1063073  1063074  1063075  1063076  1063077  1063078  1063079  1063080  1063081  1063082  1063084  1063085  1063086  1063087  1063088  1063089  1063090  1063091  1063092  1063093  1063094  1063095  1063096  1063098  1063099  1063100  1063101  1063103  1063105  1063106  1063107  1063108  1063109  1063110  1063111  1063112  1063113  1063114  1063116  1063117  1063118  1063119  1063121  1063123  1063124  1063125  1063126  1063127  1063128  1063129  1063130  1063131  1063132  1063133  1063134  1063135  1063136  1063137  1063138  1063139  1063140  1063141  1063143  1063144  1063146  1063147  1063148  1063149  1063150  1063151  1063152  1063153  1063154  1063155  1063156  1063157  1063160  1063161  1063162  1063164  1063165  1063166  1063167  1063168  1063169  1063170  1063171  1063172  1063173  1063174  1063175  1063176  1063177  1063180  1063181  1063182  1063183  1063184  1063185  1063186  1063187  1063188  1063189  1063190  1063191  1063192  1063193  1063194  1063195  1063196  1063197  1063198  1063199  1063200  1063201  1063202  1063203  1063204  1063205  1063206  1063207  1063208  1063209  1063210  1063211  1063212  1063213  1063214  1063215  1063216  1063217  1063218  1063220  1063221  1063222  1063223  1063224  1063225  1063226  1063227  1063228  1063229  1063230  1063231  1063232  1063233  1063234  1063235  1063236  1063237  1063238  1063239  1063240  1063241  1063242  1063243  1063244  1063245  1063410  1063867  1063884  1063936  1063957  1063969  1063985  1063986  1064001  1064015  1064047  1064091  1064116  1064120  1064147  1064158  1064162  1064182  1064255  1064262  1064263  1064264  1064269  1064271  1064273  1064290  1064295  1064296  1064298  1064299  1064308  1064315  1064316  1064317  1064318  1064319  1064320  1064321  1064322  1064323  1064324  1064325  1064326  1064327  1064328  1064329  1064330  1064331  1064332  1064333  1064334  1064335  1064336  1064337  1064338  1064339  1064340  1064341  1064342  1064343  1064344  1064345  1064346  1064347  1064348  1064349  1064350  1064351  1064352  1064353  1064354  1064355  1064356  1064357  1064358  1064359  1064360  1064361  1064362  1064363  1064364  1064365  1064366  1064367  1064368  1064369  1064370  1064371  1064372  1064373  1064374  1064375  1064376  1064377  1064378  1064379  1064380  1064381  1064382  1064383  1064384  1064385  1064386  1064387  1064388  1064389  1064390  1064391  1064392  1064393  1064394  1064395  1064396  1064397  1064398  1064399  1064400  1064401  1064402  1064403  1064404  1064405  1064406  1064407  1064408  1064409  1064410  1064411  1064412  1064413  1064414  1064415  1064416  1064417  1064418  1064419  1064420  1064421  1064422  1064423  1064424  1064425  1064426  1064427  1064428  1064429  1064430  1064431  1064432  1064433  1064434  1064435  1064436  1064437  1064438  1064439  1064440  1064441  1064442  1064446  1064447  1064448  1064449  1064450  1064451  1064452  1064453  1064454  1064455  1064456  1064457  1064458  1064459  1064460  1064461  1064462  1064463  1064464  1064465  1064466  1064467  1064468  1064469  1064470  1064471  1064472  1064473  1064474  1064475  1064476  1064477  1064478  1064479  1064480  1064481  1064482  1064483  1064484  1064485  1064486  1064487  1064488  1064489  1064490  1064491  1064492  1064493  1064494  1064495  1064496  1064497  1064498  1064499  1064500  1064501  1064502  1064503  1064504  1064505  1064506  1064507  1064508  1064509  1064510  1064511  1064512  1064513  1064514  1064515  1064516  1064517  1064518  1064519  1064520  1064521  1064522  1064523  1064524  1064525  1064526  1064527  1064528  1064529  1064530  1064531  1064532  1064533  1064534  1064535  1064536  1064537  1064538  1064543  1064567  1064580  1064587  1064588  1064601  1064606  1064618  1064623  1064624  1064625  1064626  1064627  1064628  1064629  1064630  1064631  1064632  1064633  1064634  1064635  1064636  1064637  1064638  1064639  1064640  1064641  1064642  1064643  1064644  1064645  1064646  1064647  1064648  1064649  1064650  1064651  1064652  1064653  1064654  1064655  1064656  1064657  1064658  1064659  1064660  1064661  1064662  1064663  1064664  1064665  1064666  1064667  1064668  1064669  1064670  1064671  1064672  1064673  1064674  1064675  1064676  1064677  1064678  1064679  1064680  1064681  1064682  1064683  1064684  1064685  1064686  1064687  1064688  1064689  1064690  1064691  1064692  1064693  1064694  1064695  1064696  1064697  1064698  1064699  1064700  1064701  1064702  1064703  1064704  1064705  1064706  1064707  1064708  1064709  1064710  1064711  1064712  1064713  1064714  1064715  1064716  1064717  1064718  1064719  1064720  1064721  1064722  1064723  1064724  1064725  1064726  1064727  1064728  1064729  1064730  1064731  1064732  1064733  1064734  1064735  1064736  1064737  1064738  1064739  1064740  1064741  1064742  1064743  1064744  1064745  1064746  1064747  1064748  1064749  1064750  1064751  1064752  1064753  1064754  1064755  1064756  1064757  1064758  1064759  1064760  1064761  1064762  1064763  1064764  1064765  1064766  1064767  1064768  1064769  1064770  1064771  1064772  1064773  1064774  1064775  1064776  1064777  1064778  1064779  1064780  1064781  1064782  1064783  1064784  1064785  1064786  1064787  1064788  1064789  1064790  1064791  1064792  1064793  1064794  1064795  1064796  1064797  1064798  1064799  1064800  1064801  1064802  1064803  1064804  1064805  1064806  1064807  1064808  1064809  1064810  1064811  1064812  1064813  1064814  1064815  1064816  1064817  1064839  1064840  1064841  1064842  1064843  1064844  1064845  1064846  1064847  1064848  1064849  1064850  1064851  1064852  1064853  1064854  1064855  1064856  1064857  1064858  1064859  1064860  1064861  1064862  1064863  1064864  1064865  1064866  1064867  1064868  1064869  1064870  1064871  1064872  1064873  1064874  1064875  1064876  1064877  1064878  1064879  1064880  1064881  1064882  1064883  1064884  1064885  1064886  1064887  1064888  1064889  1064890  1064891  1064892  1064893  1064894  1064895  1064896  1064897  1064898  1064899  1064900  1064901  1064902  1064903  1064904  1064905  1064906  1064907  1064908  1064909  1064910  1064911  1064912  1064913  1064914  1064915  1064916  1064917  1064918  1064919  1064920  1064921  1064922  1064923  1064924  1064925  1064926  1064927  1064928  1064929  1064930  1064931  1064932  1064933  1064934  1064935  1064936  1064937  1064938  1064939  1064940  1064941  1064942  1064943  1064944  1064945  1064946  1064947  1064948  1064949  1064950  1064951  1064952  1064953  1064954  1064955  1064956  1064957  1064958  1064959  1064960  1064965  1064966  1064967  1064968  1064969  1064970  1064971  1064972  1064973  1064974  1064975  1064976  1064977  1064978  1064979  1064980  1064981  1064982  1064983  1064984  1064985  1064986  1064987  1064988  1064989  1064990  1065003  1065004  1065005  1065006  1065007  1065008  1065009  1065010  1065011  1065012  1065013  1065014  1065015  1065016  1065017  1065018  1065019  1065020  1065021  1065022  1065023  1065024  1065026  1065027  1065028  1065029  1065030  1065031  1065032  1065033  1065034  1065035  1065036  1065037  1065038  1065039  1065040  1065041  1065042  1065043  1065044  1065045  1065046  1065047  1065048  1065049  1065050  1065051  1065052  1065053  1065054  1065055  1065056  1065057  1065059  1065060  1065061  1065062  1065063  1065064  1065065  1065066  1065067  1065068  1065069  1065070  1065071  1065072  1065073  1065078  1065079  1065080  1065081  1065082  1065083  1065084  1065085  1065086  1065087  1065088  1065089  1065091  1065092  1065093  1065095  1065096  1065097  1065098  1065099  1065100  1065101  1065102  1065103  1065104  1065105  1065106  1065107  1065108  1065109  1065110  1065111  1065112  1065113  1065114  1065115  1065116  1065117  1065118  1065119  1065120  1065121  1065122  1065123  1065124  1065125  1065126  1065127  1065128  1065129  1065130  1065131  1065133  1065134  1065135  1065136  1065137  1065138  1065139  1065140  1065141  1065142  1065143  1065144  1065145  1065146  1065147  1065148  1065149  1065150  1065151  1065152  1065153  1065154  1065155  1065156  1065157  1065158  1065159  1065160  1065161  1065162  1065163  1065164  1065165  1065166  1065167  1065168  1065169  1065170  1065171  1065172  1065173  1065174  1065176  1065177  1065178  1065179  1065180  1065181  1065182  1065183  1065184  1065185  1065186  1065187  1065188  1065189  1065190  1065191  1065192  1065193  1065194  1065195  1065196  1065197  1065198  1065199  1065201  1065202  1065203  1065204  1065205  1065206  1065207  1065208  1065209  1065210  1065211  1065212  1065213  1065214  1065215  1065216  1065217  1065218  1065219  1065220  1065754  1065756  1067084  1067087  1067088  1067089  1067090  1067091  1067092  1067093  1067094  1067095  1067096  1067097  1067098  1067099  1067100  1067101  1067102  1067103  1067104  1067105  1067106  1067107  1067108  1067109  1067110  1067111  1067112  1067113  1067114  1067115  1067116  1067117  1067118  1067119  1067120  1067121  1067122  1067123  1067124  1067125  1067126  1067127  1067128  1067129  1067130  1067131  1067132  1067133  1067134  1067135  1067136  1067137  1067138  1067139  1067140  1067141  1067142  1067143  1067144  1067145  1067146  1067147  1067148  1067149  1067150  1067151  1067152  1067153  1067154  1067155  1067156  1067157  1067158  1067159  1067160  1067161  1067162  1067163  1067164  1067165  1067166  1067167  1067168  1067169  1067170  1067171  1067172  1067173  1067174  1067175  1067176  1067177  1067178  1067179  1067180  1067181  1067182  1067183  1067184  1067185  1067186  1067187  1067188  1067189  1067190  1067191  1067192  1067193  1067194  1067195  1067196  1067197  1067198  1067199  1067200  1067201  1067202  1067203  1067204  1067205  1067206  1067207  1067208  1067209  1067210  1067211  1067212  1067213  1067214  1067215  1067216  1067217  1067218  1067219  1067220  1067222  1067223  1067224  1067225  1067226  1067227  1067228  1067232  1067233  1067427  1067434  1067443  1067449  1067456  1067462  1067477  1067478    Serial Number OUS  1061043  1061045  1061046  1061047  1061048  1061049  1061050  1061051  1061052  1061053  1061054  1061055  1061056  1061057  1061059  1061060  1061061  1061062  1061063  1061064  1061065  1061066  1061067  1061068  1061069  1061070  1061071  1061073  1061074  1061076  1061077  1061078  1061079  1061080  1061081  1061084  1061085  1061086  1061087  1061088  1061089  1061090  1061091  1061092  1061094  1061095  1061096  1061097  1061098  1061099  1061100  1061101  1061102  1061103  1061237  1061238  1061239  1061240  1061241  1061255  1061258  1061300  1061301  1061302  1061303  1061304  1061305  1061306  1061307  1061309  1061310  1061312  1061395  1061429  1061430  1061432  1061433  1061434  1061435  1061436  1061437  1061438  1061439  1061440  1061441  1061442  1061443  1061444  1061445  1061446  1061447  1061448  1061449  1061450  1061452  1061453  1061454  1061455  1061456  1061457  1061459  1061460  1061463  1061464  1061465  1061466  1061467  1061468  1061469  1061470  1061471  1061472  1061473  1061474  1061475  1061476  1061477  1061478  1061479  1061480  1061481  1061483  1061484  1061485  1061486  1061487  1061488  1061489  1061490  1061491  1061492  1061493  1061494  1061495  1061496  1061497  1061498  1061499  1061500  1061501  1061502  1061503  1061504  1061505  1061506  1061508  1061510  1061511  1061512  1061513  1061514  1061515  1061516  1061517  1061518  1061519  1061520  1061521  1061523  1061524  1061525  1061526  1061527  1061528  1061529  1061531  1061532  1061533  1061534  1061535  1061536  1061537  1061538  1061539  1061540  1061541  1061542  1061543  1061545  1061547  1061548  1061550  1061551  1061553  1061554  1061555  1061556  1061557  1061558  1061559  1061560  1061561  1061562  1061563  1061743  1061744  1061745  1061746  1061747  1061748  1061749  1061750  1061751  1061752  1061754  1061755  1061756  1061757  1061758  1061759  1061760  1061761  1061762  1061763  1061764  1061765  1061766  1061767  1061769  1061771  1061775  1061776  1061777  1061778  1061779  1061780  1061781  1061782  1061783  1061784  1061785  1061786  1061788  1061789  1061790  1061791  1061792  1061870  1061871  1061873  1061875  1061876  1061877  1061878  1061879  1061880  1061881  1061882  1061883  1061884  1061885  1061886  1061887  1061888  1061889  1061890  1061892  1061893  1061894  1061895  1061896  1061897  1061898  1061899  1061900  1061901  1061902  1061903  1061904  1061905  1061906  1061907  1061908  1061909  1061910  1061911  1061912  1061913  1061914  1061915  1061916  1061919  1061920  1061921  1061922  1061923  1062201  1062205  1062217  1062222  1062229  1062308  1062309  1062310  1062311  1062312  1062314  1062315  1062316  1062336  1062337  1062338  1062339  1062398  1062408  1062417  1062418  1062422  1062452  1062453  1062454  1062455  1062456  1062498  1062499  1062500  1062501  1062502  1062503  1062504  1062505  1062506  1062507  1062508  1062509  1062510  1062511  1062512  1062513  1062514  1062515  1062516  1062517  1062518  1062519  1062520  1062521  1062522  1062523  1062524  1062526  1062527  1062528  1062530  1062531  1062532  1062533  1062534  1062535  1062536  1062537  1062538  1062539  1062540  1062541  1062542  1062543  1062544  1062545  1062546  1062547  1062548  1062549  1062550  1062551  1062552  1062554  1062555  1062556  1062557  1062558  1062559  1062560  1062561  1062562  1062563  1062564  1062566  1062567  1062568  1062569  1062570  1062571  1062572  1062574  1062575  1062576  1062577  1062578  1062579  1062580  1062581  1062582  1062583  1062584  1062585  1062586  1062587  1062588  1062589  1062590  1062591  1062592  1062593  1062594  1062595  1062596  1062597  1062598  1062599  1062600  1062601  1062602  1062603  1062604  1062605  1062606  1062607  1062608  1062609  1062610  1062611  1062612  1062613  1062614  1062615  1062616  1062617  1062618  1062619  1062620  1062621  1062622  1062623  1062624  1062625  1062626  1062627  1062628  1062629  1062630  1062631  1062632  1062633  1062634  1062635  1062636  1062637  1062638  1062647  1062650  1062651  1062652  1062653  1062654  1062655  1062656  1062657  1062658  1062659  1062660  1062661  1062662  1062663  1062664  1062665  1062666  1062668  1062669  1062670  1062671  1062672  1062673  1062674  1062675  1062676  1062677  1062678  1062693  1062694  1062695  1062696  1062697  1062698  1062699  1062701  1062702  1062703  1062704  1062705  1062706  1062707  1062708  1062709  1062710  1062711  1062712  1062713  1062716  1062717  1062718  1062719  1062720  1062721  1062722  1062723  1062724  1062725  1062726  1062727  1062728  1062729  1062730  1062731  1062732  1062733  1062734  1062735  1062736  1062737  1062738  1062739  1062740  1062742  1062743  1062744  1062745  1062746  1062747  1062748  1062749  1062750  1062751  1062752  1062753  1062754  1062755  1062756  1062757  1062758  1062759  1062760  1062761  1062762  1062763  1062764  1062765  1062766  1062767  1062768  1062769  1062770  1062771  1062772  1062773  1062774  1062776  1062777  1062778  1062781  1062782  1062783  1062786  1062787  1062788  1062789  1062790  1062791  1062792  1062793  1062794  1062797  1062798  1062799  1062801  1062802  1062803  1062804  1062805  1062806  1062807  1062808  1062809  1062810  1062811  1062812  1062813  1062814  1062815  1062816  1062819  1062825  1062838  1062839  1062842  1062843  1062844  1062845  1062846  1062847  1062848  1062851  1062852  1062853  1062854  1062855  1062856  1062857  1062858  1062859  1062860  1062861  1062862  1062863  1062864  1062865  1062866  1062867  1062868  1062869  1062870  1062871  1062872  1062873  1062874  1062875  1062876  1062877  1062881  1062899  1062927  1062942  1062943  1062961  1062976  1062986  1062994  1062995  1062996  1062997  1062999  1063000  1063001  1063002  1063003  1063004  1063005  1063006  1063007  1063008  1063009  1063010  1063011  1063012  1063013  1063014  1063015  1063016  1063018  1063019  1063020  1063021  1063022  1063023  1063024  1063025  1063026  1063027  1063028  1063029  1063030  1063031  1063032  1063033  1063083  1063097  1063102  1063104  1063115  1063120  1063122  1063142  1063145  1063158  1063159  1063163  1063178  1063179  1063219  1063246  1063247  1063248  1063249  1063250  1063251  1063252  1063253  1063255  1063256  1063257  1063260  1063262  1063263  1063265  1063266  1063268  1063269  1063270  1063273  1063278  1063279  1063280  1063281  1063282  1063283  1063284  1063285  1063286  1063287  1063288  1063289  1063290  1063291  1063292  1063293  1063294  1063295  1063296  1063297  1063298  1063299  1063300  1063301  1063302  1063303  1063305  1063306  1063307  1063308  1063309  1063310  1063311  1063312  1063313  1063314  1063315  1063316  1063317  1063318  1063319  1063320  1063321  1063322  1063323  1063324  1063325  1063326  1063327  1063328  1063329  1063330  1063331  1063332  1063333  1063334  1063335  1063336  1063337  1063338  1063339  1063340  1063341  1063342  1063343  1063344  1063345  1063346  1063347  1063348  1063349  1063350  1063351  1063352  1063354  1063355  1063356  1063357  1063358  1063359  1063360  1063361  1063362  1063363  1063364  1063365  1063366  1063367  1063368  1063369  1063370  1063371  1063372  1063373  1063374  1063375  1063376  1063377  1063378  1063379  1063380  1063381  1063382  1063383  1063384  1063385  1063386  1063387  1063388  1063389  1063390  1063391  1063392  1063393  1063394  1063395  1063396  1063397  1063398  1063399  1063400  1063401  1063402  1063403  1063404  1063405  1063406  1063407  1063408  1063409  1063411  1063412  1063413  1063414  1063415  1063416  1063417  1063418  1063419  1063420  1063421  1063422  1063423  1063424  1063425  1063426  1063427  1063428  1063429  1063430  1063431  1063432  1063433  1063434  1063435  1063436  1063437  1063438  1063439  1063440  1063441  1063442  1063443  1063444  1063445  1063446  1063447  1063448  1063449  1063450  1063451  1063452  1063453  1063454  1063455  1063456  1063457  1063458  1063459  1063460  1063461  1063462  1063463  1063464  1063465  1063466  1063467  1063468  1063469  1063470  1063471  1063472  1063473  1063474  1063475  1063476  1063477  1063478  1063479  1063480  1063481  1063482  1063483  1063484  1063485  1063486  1063487  1063488  1063489  1063490  1063491  1063492  1063493  1063494  1063495  1063496  1063497  1063498  1063499  1063500  1063501  1063502  1063503  1063504  1063505  1063506  1063507  1063508  1063509  1063510  1063511  1063512  1063513  1063514  1063515  1063516  1063520  1063521  1063522  1063523  1063524  1063525  1063526  1063527  1063528  1063529  1063530  1063531  1063532  1063533  1063534  1063535  1063536  1063537  1063538  1063539  1063540  1063541  1063542  1063543  1063544  1063545  1063546  1063547  1063548  1063549  1063550  1063551  1063552  1063553  1063554  1063555  1063556  1063557  1063558  1063559  1063560  1063561  1063562  1063563  1063564  1063565  1063566  1063567  1063568  1063569  1063570  1063571  1063572  1063573  1063574  1063575  1063576  1063577  1063578  1063579  1063580  1063581  1063582  1063583  1063584  1063585  1063586  1063587  1063588  1063589  1063590  1063591  1063592  1063593  1063594  1063595  1063596  1063597  1063598  1063599  1063600  1063601  1063602  1063603  1063604  1063605  1063606  1063607  1063608  1063609  1063610  1063611  1063612  1063613  1063614  1063615  1063616  1063617  1063618  1063619  1063620  1063621  1063622  1063623  1063624  1063625  1063626  1063627  1063628  1063629  1063630  1063631  1063632  1063633  1063634  1063635  1063636  1063637  1063638  1063639  1063640  1063641  1063642  1063643  1063644  1063645  1063646  1063647  1063648  1063649  1063650  1063651  1063652  1063653  1063654  1063655  1063656  1063657  1063658  1063659  1063660  1063661  1063662  1063663  1063664  1063665  1063666  1063667  1063668  1063669  1063670  1063671  1063672  1063673  1063674  1063675  1063676  1063677  1063678  1063679  1063680  1063681  1063682  1063683  1063684  1063685  1063686  1063687  1063688  1063689  1063690  1063691  1063692  1063693  1063694  1063695  1063696  1063697  1063699  1063700  1063701  1063702  1063703  1063704  1063705  1063706  1063707  1063708  1063709  1063710  1063711  1063712  1063713  1063714  1063715  1063716  1063717  1063718  1063719  1063720  1063721  1063722  1063723  1063724  1063725  1063726  1063727  1063728  1063729  1063730  1063731  1063732  1063733  1063734  1063735  1063736  1063737  1063738  1063739  1063740  1063741  1063742  1063743  1063744  1063745  1063746  1063747  1063748  1063749  1063750  1063751  1063752  1063753  1063754  1063755  1063756  1063757  1063758  1063759  1063760  1063761  1063762  1063763  1063764  1063765  1063766",
      "center_classification_date": "20150618",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) including the states of  AL, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IA, KS, LA, MA, MI, MN, MS, MO, NE, NM, NY, NC, ND, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, and WI, and the countries of AT (Austria), AU (Australia), BE (Belgium), CA (Canada), CH (Switzerland), DE (Germany), DK (Denmark), FR (France), GB (Great Britain), HK (Hong Kong), LB (Lebanon), NL (Netherlands), NZ (New Zealand), QA (Qatar), SA (Saudi Arabia), and SE (Sweden).",
      "state": "MN",
      "product_description": "CADD¿- Solis Ambulatory Infusion Pump, Model 2111 & 2112, Version 1.0, Version 2.0 and Version 3.0 &  Model 2120, Version 1.0.     Provide measured drug therapy to patients in hospital or outpatient settings.",
      "report_date": "20150624",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Smiths Medical ASD, Inc.",
      "recall_number": "Z-1813-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71304",
      "termination_date": "20160509",
      "more_code_info": "1063768  1063769  1063770  1063771  1063772  1063773  1063774  1063775  1063776  1063777  1063778  1063779  1063780  1063781  1063782  1063783  1063784  1063785  1063786  1063787  1063788  1063789  1063790  1063791  1063792  1063793  1063794  1063795  1063796  1063797  1063798  1063799  1063800  1063801  1063802  1063803  1063804  1063805  1063806  1063807  1063808  1063809  1063810  1063811  1063812  1063813  1063814  1063815  1063817  1063818  1063819  1063820  1063821  1063822  1063823  1063824  1063825  1063826  1063827  1063828  1063829  1063830  1063831  1063832  1063833  1063834  1063835  1063836  1063837  1063838  1063839  1063840  1063841  1063842  1063843  1063844  1063845  1063846  1063847  1063848  1063849  1063850  1063851  1063852  1063853  1063854  1063855  1063856  1063857  1063858  1063859  1063860  1063861  1063862  1063863  1063864  1063865  1063866  1063868  1063869  1063870  1063871  1063872  1063873  1063874  1063875  1063876  1063877  1063878  1063879  1063880  1063881  1063882  1063883  1063885  1063886  1063887  1063888  1063889  1063890  1063891  1063892  1063893  1063894  1063895  1063896  1063897  1063898  1063899  1063900  1063901  1063902  1063903  1063904  1063905  1063906  1063907  1063908  1063909  1063910  1063911  1063912  1063913  1063914  1063915  1063916  1063917  1063918  1063919  1063920  1063921  1063922  1063923  1063924  1063925  1063926  1063927  1063928  1063929  1063930  1063931  1063932  1063933  1063934  1063935  1063937  1063938  1063939  1063940  1063941  1063942  1063943  1063944  1063945  1063946  1063947  1063948  1063949  1063950  1063951  1063952  1063953  1063954  1063955  1063956  1063958  1063959  1063960  1063961  1063962  1063963  1063964  1063965  1063966  1063967  1063968  1063970  1063971  1063972  1063973  1063974  1063975  1063976  1063977  1063978  1063979  1063980  1063981  1063982  1063983  1063984  1063987  1063988  1063989  1063990  1063991  1063992  1063993  1063994  1063995  1063996  1063997  1063998  1063999  1064000  1064002  1064003  1064004  1064005  1064006  1064007  1064009  1064010  1064012  1064013  1064014  1064016  1064017  1064018  1064019  1064020  1064021  1064022  1064023  1064024  1064025  1064026  1064027  1064028  1064029  1064030  1064031  1064032  1064033  1064034  1064035  1064036  1064041  1064042  1064043  1064044  1064045  1064046  1064048  1064049  1064050  1064051  1064052  1064053  1064054  1064055  1064056  1064057  1064058  1064059  1064060  1064061  1064062  1064063  1064064  1064065  1064066  1064067  1064068  1064069  1064070  1064071  1064072  1064073  1064074  1064075  1064076  1064077  1064078  1064079  1064080  1064081  1064082  1064083  1064084  1064085  1064086  1064087  1064088  1064089  1064090  1064092  1064093  1064094  1064095  1064096  1064097  1064098  1064099  1064100  1064101  1064102  1064103  1064104  1064105  1064106  1064107  1064108  1064109  1064110  1064111  1064112  1064113  1064114  1064115  1064117  1064118  1064119  1064121  1064122  1064123  1064124  1064125  1064126  1064127  1064128  1064129  1064130  1064131  1064132  1064133  1064134  1064135  1064136  1064137  1064138  1064139  1064140  1064141  1064142  1064143  1064144  1064145  1064146  1064148  1064149  1064150  1064151  1064152  1064153  1064154  1064155  1064156  1064157  1064159  1064160  1064161  1064163  1064164  1064165  1064167  1064168  1064169  1064170  1064171  1064172  1064173  1064174  1064175  1064176  1064177  1064178  1064179  1064180  1064181  1064183  1064184  1064185  1064186  1064187  1064188  1064189  1064190  1064191  1064192  1064193  1064194  1064195  1064196  1064197  1064198  1064199  1064200  1064201  1064202  1064203  1064204  1064205  1064206  1064207  1064208  1064209  1064210  1064211  1064212  1064213  1064214  1064215  1064216  1064217  1064218  1064219  1064220  1064221  1064222  1064223  1064224  1064225  1064226  1064227  1064228  1064229  1064230  1064231  1064232  1064233  1064234  1064235  1064236  1064237  1064238  1064239  1064240  1064241  1064242  1064243  1064244  1064245  1064246  1064247  1064248  1064249  1064250  1064251  1064252  1064253  1064254  1064256  1064257  1064258  1064259  1064260  1064261  1064265  1064266  1064267  1064268  1064270  1064272  1064274  1064275  1064276  1064277  1064278  1064279  1064280  1064281  1064282  1064283  1064284  1064285  1064286  1064287  1064288  1064289  1064291  1064292  1064293  1064294  1064297  1064300  1064301  1064302  1064303  1064304  1064305  1064306  1064307  1064309  1064310  1064311  1064312  1064313  1064314  1064443  1064444  1064445  1064539  1064540  1064541  1064542  1064544  1064545  1064546  1064547  1064548  1064549  1064550  1064551  1064552  1064553  1064554  1064555  1064556  1064557  1064558  1064559  1064560  1064561  1064562  1064563  1064564  1064565  1064566  1064568  1064569  1064570  1064571  1064572  1064573  1064574  1064575  1064576  1064577  1064578  1064579  1064581  1064582  1064583  1064584  1064585  1064586  1064589  1064590  1064591  1064592  1064593  1064594  1064595  1064596  1064597  1064598  1064599  1064600  1064602  1064603  1064604  1064605  1064607  1064608  1064609  1064610  1064611  1064612  1064613  1064614  1064615  1064616  1064617  1064619  1064620  1064621  1064622  1065271  1065331  1065332  1065333  1065334  1065335  1065336  1065338  1065339  1065340  1065341  1065342  1065343  1065344  1065345  1065346  1065347  1065348  1065349  1065350  1065351  1065352  1065353  1065354  1065356  1065469  1065529  1065537  1065538  1065539  1065540  1065541  1065542  1065549  1065550  1065552  1065553  1065554  1065555  1065556  1065557  1065558  1065559  1065560  1065561  1065562  1065563  1065564  1065565  1065566  1065567  1065568  1065569  1065570  1065571  1065572  1065573  1065574  1065575  1065576  1065577  1065578  1065579  1065580  1065581  1065582  1065583  1065584  1065585  1065586  1065587  1065588  1065589  1065590  1065592  1065593  1065594  1065595  1065671  1065672  1065674  1065675  1065676  1065677  1065678  1065683  1065685  1065686  1065689  1065690  1065739  1065740  1065741  1065743  1065746  1065747  1065748  1065749  1065750  1065751  1065752  1065753  1065755  1065757  1065758  1065759  1065760  1065761  1065762  1065763  1065764  1065765  1065766  1065767  1065768  1065769  1065770  1065771  1067072  1067073  1067074  1067075  1067076  1067077  1067078  1067079  1067080  1067081  1067082  1067083  1067085  1067086  1067221  1067229  1067230  1067231  1067234  1067235  1067236  1067237  1067238  1067390  1067391  1067392  1067393  1067394  1067395  1067396  1067397  1067398  1067399  1067400  1067401  1067402  1067403  1067404  1067405  1067406  1067407  1067408  1067409  1067410  1067411  1067412  1067413  1067414  1067415  1067416  1067417  1067418  1067419  1067420  1067421  1067422  1067423  1067424  1067425  1067426  1067428  1067429  1067430  1067431  1067432  1067435  1067436  1067437  1067438  1067439  1067440  1067441  1067442  1067444  1067445  1067446  1067447  1067448  1067450  1067451  1067452  1067453  1067454  1067455  1067457  1067458  1067459  1067460  1067461  1067464  1067465  1067466  1067467  1067468  1067469  1067470  1067471  1067472  1067473  1067474  1067475  1067476",
      "recall_initiation_date": "20150526",
      "postal_code": "55112-6929",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}