{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90628",
      "recalling_firm": "Tandem Diabetes Care Inc",
      "address_1": "11075 Roselle St",
      "address_2": "N/A",
      "postal_code": "92121-1204",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: NC, OK, ME, NY, TX, LA, NE, FL, AZ, IL, OH, CA, AL, MI, UT, VA, MA, SC, MS, GA, NV, HI, MN, WA, IN, AR, ID, CO, IA, MO, MT, PA, WI, NH, MD, WV, KS, NM, OR, KY, DE, ND, CT, NJ, TN, WY, AK, SD, DC, RI, IA, PR, VT, PT, VI, NC.   OUS: Australia, New Zealand, South Africa",
      "recall_number": "Z-1812-2022",
      "product_description": "t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ technology; t: slim x2 insulin pump with the Control-IQ Technology",
      "product_quantity": "8399",
      "reason_for_recall": "Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.",
      "recall_initiation_date": "20220524",
      "center_classification_date": "20220926",
      "report_date": "20221005",
      "code_info": "Insulin Pump/Software Versions: t:slim X2 insulin pump/6.4.1 and earlier; t:slim X2 insulin pump with Dexcom G5/5.2.1, 5.2.2; t:slim X2 insulin pump with Basal-IQ technology/6.3.0.1, 6.4, 6.4.1; t: slim x2 insulin pump with the Control-IQ Technology/7.4, 7.4.3"
    }
  ]
}