  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73959",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide distribution, including  Puerto Rico and Guam.",
      "recall_number": "Z-1812-2016",
      "product_description": "Draeger D-Vapor/ D-Vapor 3000 Vaporizers with Baxter Desflurane Agent and Bottles.   Catalog # M35500 (D-Vapor) and M36700 (D-Vapor 3000).  Anesthesia Vaporizer",
      "product_quantity": "16078 units distributed within the United States",
      "reason_for_recall": "Operators noticed unwanted release of evaporating anesthetic agent at the end of the filling procedure.    Eye contact with volatile anesthetics can result in serious irritation. Immediate eye flushing and consultation of a physician is recommended.  Updated Material Safety Data Sheets (MSDS) for Desflurane recommend wearing safety glasses when handling these substances.",
      "recall_initiation_date": "20160421",
      "center_classification_date": "20160525",
      "termination_date": "20161207",
      "report_date": "20160601",
      "code_info": "Catolog Number M35500 (D-Vapor)  M36700 (D-Vapor 3000)  Distributed October 2004 \u0013 April 2016"
    }
  ]
}