  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Goleta",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65600",
      "recalling_firm": "Medtronic Neurosurgery",
      "address_1": "125 Cremona Dr",
      "address_2": "N/A",
      "postal_code": "93117-3083",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution in the US.",
      "recall_number": "Z-1810-2013",
      "product_description": "Medtronic Preimplantation Test Kit, Catalog No. 21047. The kit is indicated for use with Medtronic CSF-Flow Control Valves and Delta Valves, when a simple preimplantation test is desired to verify that the valve conforms to labeled product specifications.",
      "product_quantity": "915 units",
      "reason_for_recall": "Medtronic Neurosurgery is recalling the Medtronic Preimplantation Test Kit because the outer carton had expiration dates that were incorrect.  The date on the outer carton is later than the expiration date for one or more of the individually packaged sterile components in the kit.",
      "recall_initiation_date": "20130628",
      "center_classification_date": "20130725",
      "termination_date": "20130923",
      "report_date": "20130731",
      "code_info": "C22929 C40543 C56465 C80034 D16291  C25329 C42910 C57924 C80787 D16693  C25475 C43115 C62128 D00052 D20967  C25662 C43512 C65929 D01033 D22544  C28396 C45615 C66120 D07250 D25802  C29229 C46351 C68906 D07373 D30597  C33946 C46583 C69701 D07507  C36893 C49439 C71003 D10462  C36904 C49706 C73622 D11409  C38718 C51418 C76639 D13208  C39484 C52417 C78186 D14099  C40150 C53058 C79051 D16157",
      "more_code_info": ""
    }
  ]
}