{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Stockholm",
      "state": "N/A",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68162",
      "recalling_firm": "RAYSEARCH LABORATORIES AB",
      "address_1": "Sveavaegen 9",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including the states of  AZ, CA, FL, IL, MA, MD, MO, MS, NJ, NV, NY, OH, PA, TN, VA, and WA, and the countries of Australia, Belgium, Germany, France, Italy, New Zealand, Republic of Korea, Switzerland, Spain, and United Kingdom.",
      "recall_number": "Z-1806-2014",
      "product_description": "RayStation software version 4.0.3.4    RayStation is a software system designed for treatment planning and analysis of radiation therapy.",
      "product_quantity": "25 sites/245 units (US), 25 sites/124 units (foreign)",
      "reason_for_recall": "This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution.",
      "recall_initiation_date": "20140425",
      "center_classification_date": "20140616",
      "termination_date": "20170626",
      "report_date": "20140625",
      "code_info": "Software version 4.0.3.4",
      "more_code_info": ""
    }
  ]
}