  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Brea",
      "address_1": "250 S Kraemer Blvd",
      "reason_for_recall": "Beckman Coulter has received customer complaints of lowered minimum inhibitory concentration (MIC) for antimicrobial agents Cefaclor (Ctr) and Penicillin (P) with control organism Streptococcus pneumoniae (ATCC 49619) during quality control (QC) testing on MICroSTREP plus(R) Panel Type 1 hydrated with Mueller-Hinton Broth with 3% lysed horse blood part number B1015-25 lot number 1005986140.",
      "address_2": "",
      "product_quantity": "523 boxes",
      "code_info": "Lot Number 1005986140, Expiration date April 1, 2018",
      "center_classification_date": "20180511",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) Distribution and to the countries of :  Canada, Kuwait, South Korea and Saudi Arabia.",
      "state": "CA",
      "product_description": "BECKMAN COULTER MicroScan MICroSTREP Plus Panel, MicroScan Mueller-Hinton Broth with 3% lysed horse blood, Catalog Number 81015    MicroScan Mueller Hinton Broth with 3% Lysed Horse Blood is used for inoculation of MicroScan MICroSTREP plus and MicroFAST panels. MICroSTREP plus panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae.",
      "report_date": "20180523",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Beckman Coulter Inc.",
      "recall_number": "Z-1804-2018",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "79800",
      "termination_date": "20181219",
      "more_code_info": "",
      "recall_initiation_date": "20180326",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}