      "status": "Terminated",
      "city": "Marcy L'Etoile",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76727",
      "recalling_firm": "BioMerieux SA",
      "address_1": "Chemin De L'Orme",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including AZ, CA, CO, FL, GA, IN, IA, LA, MI, MS, NY, NC, OH, OK, OR, PA, TX, WA",
      "recall_number": "Z-1801-2017",
      "product_description": "VIDAS Estradiol II, Ref 30431-01, 60 tests  The VIDAS Estradiol II (E2 II) assay is intended for use on the instruments of the VIDAS family (VITEK ImmunoDiagnostic Assay System) as an automated quantitative enzyme-linked fluorescent immunoassay (ELFA) to the determination of total estradiol concentration in human serum or plasma (heparin).",
      "product_quantity": "834 kits",
      "reason_for_recall": "Fulvestrant cross reacts with the VIDAS Estradiol II immunoassay and may cause falsely elevated estradiol results in patients treated with Fulvestrant.",
      "recall_initiation_date": "20170317",
      "center_classification_date": "20170412",
      "termination_date": "20220216",
      "report_date": "20170419",
      "code_info": "Lots 1005081700, 1005172850 & 1005359170",
      "more_code_info": ""
    }
  ]
}