{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Bethlehem",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98552",
      "recalling_firm": "B Braun Medical Inc",
      "address_1": "824 12th Ave",
      "address_2": "N/A",
      "postal_code": "18018-3524",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the country of Canada.",
      "recall_number": "Z-1798-2026",
      "product_description": "B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG DR.   Model Number: SL-2010M2096A.",
      "product_quantity": "3,158,104 units",
      "reason_for_recall": "Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.",
      "recall_initiation_date": "20260319",
      "center_classification_date": "20260415",
      "report_date": "20260422",
      "code_info": "Model Number SL-2010M2096A; UDI-DI Primary: 04046955674992; UDI-DI Unit: 04046955735280; All lots manufactured since 09JUN2025"
    }
  ]
}