  "results": [
    {
      "status": "Terminated",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87945",
      "recalling_firm": "Smiths Medical ASD Inc.",
      "address_1": "6000 Nathan Ln N",
      "address_2": "N/A",
      "postal_code": "55442-1690",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of UK,  Spain.",
      "recall_number": "Z-1798-2021",
      "product_description": "JELCO HYPODERMIC NEEDLE-PRO FIXED NEEDLE INSULIN SYRINGE:  a) 1.0mL 28G X 1/2\" (0.33mm x 13mm), REF 4428-1  b) 1.0mL 29G X 1/2\" (0.33mm x 13mm), REF 4429-1",
      "product_quantity": "1,432,200 units",
      "reason_for_recall": "Specific models and lot numbers of the Jelco Hypodermic Needle-Pro Fixed Needle Insulin Syringe 29Gx1/2 odd number line graduation markings on the syringe barrel were skewed approximately 20 degrees upward.",
      "recall_initiation_date": "20210517",
      "center_classification_date": "20210610",
      "termination_date": "20231011",
      "report_date": "20210616",
      "code_info": "a) REF 4428-1, Lot Numbers:  4046543 and 4062235  b) REF 4429-1, Lot Numbers:  4031846, 4031845, 4040734, 4043536, 4046545, 4046546, 4062239, 4062240, 4062238, 4062242, 4014096"
    }
  ]
}