      "status": "Terminated",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85268",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide.  Global distribution.",
      "recall_number": "Z-1797-2020",
      "product_description": "Atellica CH 930 Analyzer, Siemens Material Number 11067000 with software version v1.21.0 SP1 or lower",
      "product_quantity": "1807",
      "reason_for_recall": "Editing an assay Test Definition and switching to another assay s Test Definition screen without saving the changes using the  save  button on the screen, may corrupt the settings of the assay s test definition, and Auto-Rerun or Auto-Dilution results may generate a result of Zero (0) for quantitative assays or Negative for qualitative assays if a well of the reagent pack is not calibrated.",
      "recall_initiation_date": "20200227",
      "center_classification_date": "20200424",
      "termination_date": "20211025",
      "report_date": "20200506",
      "code_info": "UDI: 00630414002163;  All systems with software version v1.21.0 SP1 or lower    Serial Numbers:  CM01702  CM01708  CM00184  CM00196  CM01051  CM01467  CM00770  CM00815  CM00983  CM01053  CM00802  CM01004  CM01903  CM00526  CM00635  CM00809  CM00733  CM00735  CM01045  CM01050  CM01052  CM01059  CM00420  CM00632  CM00633  CM00209  CM00210  CM01799  CM01590  CM01612  CM01047  CM01048  CM00859  CM00865  CM01624  CM01626  CM00829  CM00830  CM00651  CM00654  CM01603  CM01608  CM01695  CM01701  CM00749  CM00786  CM01567  CM01569  CM01682  CM01683  CM00773  CM01431  CM01474  CM01583  CM01598  CM01602  CM00148  CM00151  CM00152  CM01907  CM01738  CM01739  CM01105  CM00799  CM00800  CM01823  CM01917  CM01409  CM01410  CM01588  CM01600  CM01274  CM01306  CM00947  CM00948  CM01660  CM01661  CM00812  CM00828  CM00820  CM01593  CM01594  CM01585  CM01599  CM01058  CM00421  CM00606  CM00331  CM00333  CM01133  CM01149  CM01619  CM01191  CM01200  CM00769  CM00858  CM01237  CM01259  CM00797  CM00798  CM01589  CM01614  CM00426  CM00427  CM00701  CM00704  CM01264  CM01278  CM01391  CM01512  CM00611  CM00612  CM01735  CM00643  CM00646  CM00831  CM00832  CM01761  CM01772  CM01727  CM01740  CM01745  CM00315  CM00316  CM00123  CM00181  CM01031  P-P1-07  CM00147  CM00432  CM00297  CM00430  CM00780  CM01296  CM00268  CM00271  CM00265  CM00267  CM01759  CM01760  CM01579  CM01611  CM01139  CM00607  CM00625  CM01699  CM01716  CM00884  CM00888  CM00577  CM00631  CM01650  CM00459  CM00510  CM01441  CM01343  CM01394  CM00973  CM00977  CM01586  CM01616  CM01620  CM01621  CM00634  CM00637  CM01400  CM01406  CM00626  CM00627  CM01344  CM01345  CM00418  CM00419  CM00724  CM00628  CM00630  CM00958  CM00979  CM00971  CM00481  CM00482  CM01066  CM01118  CM01923  CM01675  CM01241  CM01253  CM01242  CM01255  CM01247  CM01268  CM01801  CM01528  CM01529  CM00864  CM00867  CM00400  CM00407  CM01455  CM01459  CM00851  CM00852  CM01499  CM00330  CM00343  CM00344  CM01513  CM01553  CM01818  CM01819  CM01597  CM00905  CM00915  CM00917  CM00918  CM00457  CM00458  CM01065  CM01239  CM01245  CM00639  CM00640  CM00746  CM00747  CM00748  CM01694  CM01696  CM00182  CM00183  CM01401  CM01423  CM00422  CM00423  CM00424  CM00425  CM00781  CM01576  CM00833  CM00834  CM00957  CM01016  CM00201  CM00621  CM01556  CM01566  CM01580  CM01581  CM01582  CM01800  CM01333  CM00723  CM00734  CM01767  CM01768  CM01318  CM01717  CM01773  CM01399  CM01293  CM01707  CM00768  CM00959  CM00988  CM01577  CM00229  CM00246  CM00715  CM00721  CM01072  CM00613  CM00614  CM01034  CM01068  CM01017  CM01653  CM00743  CM00744  CM01061  CM01062  CM00810  CM01362  CM00853  CM00856  CM01257  CM01258  CM01663  CM01686  CM01176  CM01724  CM00523  CM00524  CM00288  CM00289  CM01432  CM01433  CM01700  CM01617  CM00528  CM00529  CM00756  CM00757  CM00650  CM00653  CM00310  CM00525  CM00545  CM00647  CM00663  CM01395  CM01396  CM00783  CM00784  CM00208  CM00211  CM00272  CM00273  CM01106  CM01107  CM01094  CM01126  CM00449  CM00448  CM01218  CM01297  CM00361  CM00362  CM00202  CM00203  CM00293  CM00693  CM00696  CM00697  CM01645  CM01651  CM01659  CM01662  CM00965  CM00986  CM01657  CM01673  CM00803  CM00804  CM00805  CM00847  CM00962  CM00985  CM00964  CM00970  CM00982  CM01013  CM00963  CM00974  CM01057  CM01069  CM01077  CM01079  CM01081  CM01082  CM01084  CM01085  CM01086  CM01087  CM01090  CM01092  CM01093  CM01096  CM01097  CM01098  CM01099  CM01100  CM01110  CM01120  CM00953  CM01697  CM00163  CM01003  CM00527  CM00729  CM00730  CM00221  CM00222  CM00717  CM01280  CM01666  CM01630  CM01632  CM01520  CM01152  CM01163  CM01300  CM01151  CM00949  CM00664  CM00666  CM00261  CM01162  CM01170  CM00906  CM00903  CM00814  CM00684  CM01631  CM01637  CM01102  CM01165  CM01169  CM01196  CM00886  CM01023  CM01024  CM01369  CM01371  CM01374  CM01387  CM01351  CM01370  CM01372  CM01384  CM00848  CM00414  CM00415  CM00942  CM01073  CM01074  CM00996  CM00328  CM00462  CM00465  CM00467  CM01033  CM00312  CM01456  CM01064  CM01448  CM00304  CM00305  CM00409  CM01028  CM00279  CM00280  CM01019  CM00236  CM00303  CM00314  CM00329  CM00410  CM00413  CM00306  CM00655  CM00659  CM00667  CM00668  CM00755  CM00763  CM00764  CM00767  CM00788  CM00796  CM00807  CM00939  CM01055  CM01720  CM00902  CM00787  CM00821  CM01726  CM01729  CM01018  CM01041  CM01075  CM01536  CM01728  CM01734  CM00284  CM00285  CM01730  CM01185  CM01475  CM00461  CM00808  CM00731  CM00732  CM00762  CM01403  CM01430  CM00555  CM00237  CM00238  CM00239  CM00416  CM01178  CM01054  CM01112  CM00745  CM00455  CM01557  CM00874  CM01208  CM01291  CM01752  CM01043  CM01177  CM01438  CM01485  CM01130  CM01140  CM01314  CM01138  CM01146  CM01894  CM01698  CM00149  CM01470  CM00277  CM00835  CM00840  CM01187  CM01236  CM01355  CM01336  CM01199  CM01897  CM00932  CM00933  CM01325  CM00761  CM00661  CM00676  CM00951  CM00665  CM00765  CM00924  CM00925  CM01453  CM01454  CM01510  CM01584  CM00346  CM00675  CM00677  CM00725  CM00736  CM00896  CM00515  CM00516  CM00518  CM00519  CM00517  CM00509  CM00514  CM01123  CM01125  CM01128  CM01135  CM01416  CM01417  CM00197  CM00198  CM00160  CM00161  CM00167  CM00169  CM01172  CM01295  CM01610  CM01613  CM00992  CM00256  CM01192  CM00885  CM00862  CM00283  CM00278  CM00920  CM00922  CM00923  CM00929  CM01119  CM01121  CM01211  CM01210  CM01207  CM01231  CM01469  CM01483  CM01487  CM01488  CM00994  CM00999  CM01000  CM01030  CM01150  CM01181  CM00997  CM00998  CM01001  CM01002  CM01153  CM01164  CM01180  CM01652  CM01680  CM01691  CM01690  CM01015  CM01755  CM00488  CM00493  CM00494  CM01413  CM01414  CM01269  CM01279  CM00501  CM00556  CM00242  CM00244  CM00934  CM00969  CM01578  CM01605  CM00706  CM00722  CM00679  CM01111  CM01115  CM00817  CM01462  CM00433  CM00946  CM00961  CM00240  CM00258  CM00503  CM00601  CM01326  CM01327  CM01134  CM01142  CM01390  CM01392  CM01412  CM01445  CM00558  CM00935  CM00936  CM00940  CM00497  CM00687  CM00691  CM00928  CM00931  CM00694  CM00703  CM00890  CM01703  CM00170  CM00171  CM00172  CM01472  CM01405  CM01407  CM01415  CM01265  CM01331  CM01440  CM01359  CM01360  CM01672  CM01688  CM01689  CM00569  CM00570  CM00582  CM00485  CM00495  CM00568  CM01209  CM01233  CM01179  CM01232  CM01250  CM01252  CM00499  CM00543  CM00443  CM01020  CM00452  CM00496  CM00135  CM00140  CM01503  CM01537  CM01541  CM00498  CM01622  CM00476  CM00792  CM00557  CM00620  CM01715  CM00154  CM00877  CM00880  CM00921  CM01515  CM01530  CM01532  CM00950  CM00776  CM01526  CM00442  CM00891  CM00894  CM01756  CM01757  CM01029  CM00857  CM00132  CM00134  CM01337  CM01341  CM00816  CM00403  CM00404  CM00405  CM00345  CM00349  CM00350  CM00153  CM00439  CM00311  CM00649  CM01408  CM01411  CM01618  CM01559  CM00137  CM00139  CM00188  CM00189  CM00190  CM00487  CM00926  CM01267  CM01275  CM01288  CM01294  CM00227  CM01032  CM01422  CM00738  CM00739  CM00740  CM00741  CM00742  CM01038  CM01039  CM00259  CM00662  CM01427  CM01439  CM01664  CM01665  CM00843  CM00844  CM01159  CM01160  CM01166  CM01743  CM01744  CM00751  CM00502  CM00276  CM00434  CM01466  CM00536  CM01465  CM00270  CM01812  CM01366  CM01234  CM01719  CM01381  CM01382  CM01385  CM01389  CM00754  CM01136  CM01380  CM01722  CM01647  CM01892  CM01495  CM01816  CM01310  CM01500  CM01171  CM01329  CM01365  CM01025  CM01721  CM01379  CM01501  CM01473  CM00760  CM00294  CM00351  CM01480  CM00165  CM01502  CM01353  CM01540  CM01544  CM01302  CM01644  CM01308  CM01354  CM01494  CM01103  CM01108  CM01010  CM00357  CM00369  CM00381  CM01426  CM01561  CM01504  CM01514  CM01562  CM01324  CM01224  CM00573  CM01320  CM00539  CM00540  CM00537  CM00538  CM00580  CM00603  CM00616  CM00431  CM01251  CM00133  CM00447  CM00456  CM00463  CM00454  CM00388  CM00385  CM00685  CM00753  CM00759  CM00775  CM01342  CM01347  CM00218  CM00219  CM00220  CM00705  CM00383  CM00617  CM00618  CM01542  CM00576  CM01558  CM01217  CM00605  CM00615  CM00166  CM00176  CM00177  CM00178  CM00391  CM00453  CM00444  CM00930  CM00952  CM01350  CM01358  CM00402  CM00440  CM00450  CM00563  CM00597  CM00254  CM01349  CM00897  CM00900  CM00910  CM00911  CM00912  CM00913  CM00466  CM00469  CM00777  CM00823  CM00260  CM01442  CM01443  CM01014  CM00195  CM00225  CM00226  CM00224  CM00882  CM00854  CM00253  CM00252  CM00384  CM00214  CM00217  CM00207  CM00145  CM00619  CM00600  CM00484  CM00486  CM01220  CM00870  CM00806  CM00468  CM01229  CM01436  CM01446  CM01450  CM01452  CM01063  CM01071  CM01076  CM00771  CM00868  CM00872  CM00608  CM00836  CM01303  CM01281  CM01311  CM01322  CM01299  CM01373  CM01364  CM01282  CM01388  CM01340  CM00689  CM00726  CM00130  CM00871  CM01533  CM00472  CM00473  CM00839  CM01709  CM00173  CM00174  CM00175  CM00441  CM01490  CM01491  CM00406  CM01713  CM00282  CM00281  CM01143  CM01145  CM01486  CM01157  CM01156  CM00247  CM01357  CM01867  CM00863  CM00893  CM01868  CM00670  CM01056  CM01778  CM01779  CM01361  CM01460  CM01204  CM00699  CM00483  CM00710  CM01027  CM01078  CM01060  CM01080  CM01091  CM01088  CM01710  CM00474  CM00475  CM00674  CM01841  CM00199  CM00200  CM01363  CM01316  CM00711  CM00712  CM00714  CM00899  CM00901  CM00908  CM00909  CM00919  CM00938  CM01629  CM00602  CM00883  CM00895  CM00194  CM00191  CM00876  CM00873  CM00308  CM00309  CM00146  CM01309  CM01313  CM00801  CM01554  CM01564  CM00464  CM01420  CM01421  CM01457  CM01731  CM01732  CM01733  CM00125  CM01008  CM01298  CM01451  CM01129  CM01522  CM00993  CM01009  CM01026  CM01228  CM01175  CM01681  CM01687  CM01437  CM01428  CM01429  CM01383  CM01386  CM00117  CM00673  CM00412  CM00193  CM01338  CM01036  CM01005  CM01748  CM01012  CM01212  CM01284  CM01292  CM01011  CM01286  CM01201  CM00968  CM01260  CM01346  CM00916  CM01367  CM01213  CM01287  CM01434  CM01435  CM01646  CM01339  CM01477  CM01676  CM01677  CM01678  CM01424  CM01425  CM01718  CM01725  CM01193  CM01198  CM00943  CM00955  CM00954  CM01202  CM01147  CM01131  CM01418  CM01419  CM01007  CM01221  CM00249  CM01285  CM01658  CM01604  CM00966  CM01393  CM00914  CM00713  CM00230  CM00551  CM01116  CM00347  CM00334  CM00945  CM01307  CM01315  CM01305  CM00698  CM00778  CM01623  CM01633  CM00287  CM00286  CM00335  CM00505  CM00504  CM01227  CM00506  CM00507  CM00508  CM00987  CM01021  CM01534  CM01538  CM00332  CM00336  CM00337  CM00708  CM00709  CM00554  CM00559  CM00561  CM00562  CM00838  CM00648  CM01506  CM01498  CM01492  CM01489  CM00491  CM00492  CM01493  CM00489  CM00758  CM00234  CM00233  CM00547  CM00549  CM00550  CM00552  CM00553  CM00560  CM00530  CM00629  CM00359  CM00360  CM00490  CM00941  CM00967  CM01197  CM01206  CM00317  CM00318  CM00319  CM00320  CM00321  CM00323  CM00324  CM00325  CM00326  CM00150  CM00136  CM01335  CM01356  CM00215  CM00216  CM00720  CM01158  CM01194  CM01319  CM01640  CM01328  CM01484  CM01625  CM00564  CM00824  CM00825  CM01124  CM00386  CM00363  CM00396  CM01674  CM00248  CM00257  CM00251  CM00366  CM01458  CM01519  CM01517  CM00245  CM01521  CM01919  CM00658  CM00813  CM00512  CM00904  CM01668  CM01669  CM01188  CM01189  CM00541  CM00669  CM00671  CM00678  CM00542  CM01270  CM01277  CM01317  CM01330  CM00688  CM00875  CM00398  CM00429  CM00428  CM00477  CM00478  CM00479  CM00752  CM00975  CM01044  CM01109  CM01114  CM00989  CM00990  CM00389  CM00390  CM01321  CM01497  CM01524  CM01525  CM00340  CM00339  CM00520  CM00521  CM00716  CM00991  CM01035  CM00793  CM00794  CM00937  CM00131  CM00500  CM00338  CM00342  CM01545  CM00826  CM00827  CM00795  CM00255  CM00275  CM00300  CM00263  CM00264  CM00692  CM00737  CM00451  CM00927  CM01552  CM01565  CM00566  CM00587  CM00594  CM00599  CM00604  CM00686  CM00690  CM00700  CM00702  CM00707  CM00978  CM00980  CM01225  CM01226  CM01246  CM01203  CM01248  CM01254  CM01261  CM01273  CM00179  CM01348  CM01468  CM01447  CM00159  CM00156  CM00158  CM01776  CM01777  CM01154  CM01161  CM01332  CM01334  CM00879  CM00887  CM01792  CM00341  CM00575  CM00204  CM00791  CM00822  CM01655  CM01746  CM00298  CM00984  CM01574  CM00850  CM00790  CM00842  CM00845  CM01527  CM01543  CM01368  CM00683  CM01186  CM01271  CM00356  CM00355  CM00438  CM00437  CM00435  CM00436  CM00574  CM00585  CM00589  CM00586  CM00590  CM00583  CM01670  CM01671  CM01615  CM01654  CM01667  CM01679  CM00682  CM00262  CM00241  CM00866  CM01083  CM00638  CM00206  CM01648  CM01649  CM01518  CM01523  CM01183  CM01205  CM01040  CM01042  CM00141  CM00143  CM00565  CM00571  CM00598  CM00591  CM01222  CM00186  CM00187  GGCCEP23  GGCCEP4  CM00578  CM00588  CM00296  CM00960  CM00972  CM01022  CM00595  CM01132  CM01290  CM01266  CM01549  CM00727  CM01535  CM00785  CM01046  CM00511  CM00728  CM01693  CM01737  CM00513  CM01505  CM01141  CM00353  CM00354"
    }
  ]
}