  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Louisville",
      "address_1": "826 Coal Creek Cir",
      "reason_for_recall": "Passive Planar Blunt Probe (Lot# 150223) was manufactured with a sharp tip rather than a blunt tip.",
      "address_2": "",
      "product_quantity": "18",
      "code_info": "lot number 150223",
      "center_classification_date": "20150612",
      "distribution_pattern": "US Distribution to the states of :  CA, MO, LA and PA., and Internationally to Ireland.",
      "state": "CO",
      "product_description": "Medtronic Passive Planar Blunt Probe Part Number 960-556, also included in Part Number Kits: 9732540, 9732622, 9733174, 9733935. As part of kits, the product is either packaged with other instruments or into a holding tray.  The Passive Planar Blunt Probe is an optically tracked, reusable instrument used in navigated surgery and is included with Medtronic cranial instrument kits. It is designed for use in patient registration and palpating brain tissue during surgical procedures.",
      "report_date": "20150624",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Navigation, Inc.",
      "recall_number": "Z-1794-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71278",
      "termination_date": "20150702",
      "more_code_info": "",
      "recall_initiation_date": "20150505",
      "postal_code": "80027-9710",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}