  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Batesville",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65259",
      "recalling_firm": "Hill-Rom, Inc.",
      "address_1": "1069 State Road 46 E",
      "address_2": "N/A",
      "postal_code": "47006-7520",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) and countries of: Canada, Colombia, Nigeria, Kuwait and Saudi Arabia.",
      "recall_number": "Z-1794-2013",
      "product_description": "TotalCare Bed used with AccuMax Quantum VPC mattress  AccuMax Quantum Complete mattress, and AccuMax Quantum Convertible mattress    The TotalCare Bed System is intended to provide a patient support ideally suited to be used in health care environments. The TotalCare Bed System may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU), and sections of the emergency department (ED). The TotalCare Bed System is capable of being used with a broad patient population as determined appropriate by the caregiver or institution.",
      "product_quantity": "4,121 Units",
      "reason_for_recall": "When the bed is put into the chair egress  position, the magnets on the mattress may not hold the bottom of the mattress flat to the bed frame, and the mattress foot section will extend away from the bed at an angle. A patient attempting to get back into the bed may miss the seat and sit on the foot section instead. The patient can slide down to the floor, or fall, with possible injury.",
      "recall_initiation_date": "20130530",
      "center_classification_date": "20130723",
      "termination_date": "20140417",
      "report_date": "20130731",
      "code_info": "Bed model numbers: PAU001010784, PAU005010784, P1900, PAK169N10184, PAK169N10384, PA8171210184, PA8175210184, PA8175210384, PA8171210384, PA8171310384, and PA8173210384.     The AccuMax Quantum VPC, Complete and Convertible mattresses are not serialized, but were distributed between September 2005 and April 2013",
      "more_code_info": ""
    }
  ]
}