      "country": "United States",
      "city": "Roswell",
      "address_1": "1009 Mansell Rd Ste F",
      "reason_for_recall": "Potential failure in the glued seal between the stem and the tip of the handpiece leading to the possible leakage of a very small drop of distilled water coolant from the handpiece in use.",
      "address_2": "",
      "product_quantity": "36 units",
      "code_info": "Lot Number 40041, exp. date February 2015 and Lot Number 40043, exp. date February 2015",
      "center_classification_date": "20120611",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) including states of: CA and NY; and country of: United Kingdom.",
      "state": "GA",
      "product_description": "PlasmaJet Open Surgery Handpiece (Part of the PlasmaJet Neutral Plasma Surgery System)    The PlasmaJet system is a neutral plasma surgery system that is designed for cutting and coagulation in open surgery and laparoscopic surgery.",
      "report_date": "20120620",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Plasma Surgical Inc.",
      "recall_number": "Z-1794-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "61951",
      "termination_date": "20121101",
      "more_code_info": "",
      "recall_initiation_date": "20120427",
      "postal_code": "30076-4806",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}