  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Richmond",
      "state": "VA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90731",
      "recalling_firm": "Mckesson Medical-Surgical Inc. Corporate Office",
      "address_1": "9954 Maryland Drive",
      "address_2": "Deep Run Iii Ste. 4000",
      "postal_code": "23233",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1792-2022",
      "product_description": "a. SPS MEDICAL SUPPLY Steam Indicator Strip SPSmedical Sterilization Chemical Indicator Strip Steam 4 Inch Model Number: SIS-250.    b. SPS MEDICAL SUPPLY SteamPlus Sterilization Integrator Sterilization Steam Plus Monitor Model Number: SSI-1000.",
      "product_quantity": "a. 1 box, b. 19 cases.",
      "reason_for_recall": "Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.",
      "recall_initiation_date": "20220525",
      "center_classification_date": "20220926",
      "report_date": "20221005",
      "code_info": "GTIN:  a. 00732224710174; 10732224710171.   b. 00732224710433; 10732224710430."
    }
  ]
}