  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Monroeville",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68226",
      "recalling_firm": "Childrens Medical Ventures",
      "address_1": "191 Wyngate Dr",
      "address_2": "N/A",
      "postal_code": "15146-3045",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including the states of MI, WI, SC, FL, TX, KS, MA, AZ, UT, WV, IN, MN, NV and VA., and the countries of Australia, Belgium and Canada.",
      "recall_number": "Z-1789-2014",
      "product_description": "Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.",
      "product_quantity": "112",
      "reason_for_recall": "The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up.  The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.",
      "recall_initiation_date": "20140429",
      "center_classification_date": "20140613",
      "termination_date": "20150406",
      "report_date": "20140625",
      "code_info": "989805603611 / 1014557, 989805603631 / 1014559, 989805604511 / 1028969, 989805604521 / 1028970, 989805604531 / 1028971, 989805604631 / 1030271, 989805605051 / 1038140, 989805605061 / 1038141, 989805608801 / U1014557, 989805608811 / U1014559, 989805608831 / U1028969, 989805608841 / U1028970, 989805608851 / U1038140, 989805604231 / 1023384",
      "more_code_info": ""
    }
  ]
}