  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "61639",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution - including the states of AL, FL, GA, MO, and VA.",
      "recall_number": "Z-1787-2012",
      "product_description": "GE Healthcare, Aestiva/5 7900 SmartVent, anesthesia machine.  Used in Datex-Ohmeda Aestiva/5 Anesthesia Systems.",
      "product_quantity": "8",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential safety issue due to the interlock slide mechanism missing from the 3-vaporizer manifold of your Aestiva 7900 Anesthesia Device. This Potential for two (2) vaporizers delivering agent at the same time.  This could manifest as over delivery of a single agent if both vaporizers contain the same agent or as delivery of more than one agent.  This m",
      "recall_initiation_date": "20120410",
      "center_classification_date": "20120617",
      "termination_date": "20120709",
      "report_date": "20120627",
      "code_info": "Serial Number  AMRP01031  AMRP00966  AMRP01030  AMRP00968  AMRP00967  AMRP01033  AMRP00970  AMRP00969"
    }
  ]
}