  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83002",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide",
      "recall_number": "Z-1781-2019",
      "product_description": "GE Centricity PACS-IW versions 3.6.0 through 3.7.3.9 SP1, SP2, SP3 and 3.7.3 SPa10, Model Numbers:  a) 2052831-00x  b) 2050578-0xx  c) 2090228-002  d) 2052782-00x",
      "product_quantity": "620 units",
      "reason_for_recall": "There is a potential that one or more  image series  (i.e., all images within an image set) may be missing from an exam without a warning displayed in the viewer.",
      "recall_initiation_date": "20190522",
      "center_classification_date": "20190613",
      "termination_date": "20240920",
      "report_date": "20190619",
      "code_info": "a) 2052831-00x, Lot or Serial Numbers:  100839-1-Centricity PACS-IW-00558122  100893-1-Centricity PACS-IW-00558171  4320-1-Centricity PACS-IW-00558202  4474-1-Centricity PACS-IW-00558169  2455-1-Centricity PACS-IW-00558139  100863-1-Centricity PACS-IW-00558143  100794-1-Centricity PACS-IW-00558193  4346-1-Centricity PACS-IW-00558136  100909-1-Centricity PACS-IW-00558198  4355-1-Centricity PACS-IW-00558204  100809-1-Centricity PACS-IW-00558124  4300-1-Centricity PACS-IW-00558207  4484-1-Centricity PACS-IW-00558213  4479-1-Centricity PACS-IW-00558214  13240-2-Centricity PACS-IW-01225875  4270-1-Centricity PACS-IW-00558129  11059-2-Centricity PACS-IW-00558101  101598-1-Centricity PACS-IW-00667133  4273-1-Centricity PACS-IW-01225873  4357-1-Centricity PACS-IW-00558444  4293-1-Centricity PACS-IW-00558131  100835-1-Centricity PACS-IW-00558568  100900-1-Centricity PACS-IW-00558170  4344-1-Centricity PACS-IW-00558196  3836-1-Centricity PACS-IW-00558179  4428-1-Centricity PACS-IW-00558205  100609-1-Centricity PACS-IW-00558130  4297-1-Centricity PACS-IW-00558134  100975-1-Centricity PACS-IW-00558137  100667-1-Centricity PACS-IW-00558163  4316-1-Centricity PACS-IW-00558210  4251-1-Centricity PACS-IW-00558120  4254-1-Centricity PACS-IW-00558211  100981-1-Centricity PACS-IW-00558132  100871-1-Centricity PACS-IW-00558146  2553-1-Centricity PACS-IW-00558206  4283-1-Centricity PACS-IW-00558187  4469-1-Centricity PACS-IW-00558128  100020-1-Centricity PACS-IW-00558147  4443-1-Centricity PACS-IW-00558167  4317-1-Centricity PACS-IW-00558208  100775-1-Centricity PACS-IW-00558209  4323-1-Centricity PACS-IW-00558164  100825-1-Centricity PACS-IW-00558172  100830-1-Centricity PACS-IW-00558166  4304-1-Centricity PACS-IW-00558145  4721-1-Centricity PACS-IW-00558138  4104-1-Centricity PACS-IW-00558133  4292-1-Centricity PACS-IW-00558203  100980-1-Centricity PACS-IW-00558140  4343-1-Centricity PACS-IW-00558115  4298-1-Centricity PACS-IW-00558116  4252-1-Centricity PACS-IW-00558119  4287-1-Centricity PACS-IW-00558117  4291-1-Centricity PACS-IW-00558194  100978-1-Centricity PACS-IW-00558141  4448-1-Centricity PACS-IW-00558118  100979-1-Centricity PACS-IW-00558142  100865-1-Centricity PACS-IW-01225871  100807-1-Centricity PACS-IW-00558121  100807-1-Centricity PACS-IW-00558212  4424-1-Centricity PACS-IW-00558429  339142-Centricity PACS-IW-01431503  13759-Centricity PACS-IW-01431508  320090-Centricity PACS-IW-01444604  320092-Centricity PACS-IW-01444618  297958-Centricity PACS-IW-01444903  329997-Centricity PACS-IW-01444898  AiKang-Centricity PACS-IW-01824737  SiteID-Centricity PACS-IW-01572187  SiteID-Centricity PACS-IW-01797297  339734-Centricity PACS-IW-01444878  313825-Centricity PACS-IW-01444348  332562-Centricity PACS-IW-01444345  338295-Centricity PACS-IW-01443383  SiteID-Centricity PACS-IW-01444354  SiteID-Centricity PACS-IW-02003770  276233-Centricity PACS-IW-01444355  331973-Centricity PACS-IW-01434505  279961-Centricity PACS-IW-01444358  SiteID-Centricity PACS-IW-00161903  SiteID-Centricity PACS-IW-00161906  186588-Centricity PACS-IW-01461366  SiteID-Centricity PACS-IW-02095351  RU9038-Centricity PACS-IW-01273313  RU1107-Centricity PACS-IW-01273202  MPX431-Centricity PACS-IW-00044783    b) 2050578-0xx  Lot or Serial Numbers:  100664-1-Centricity PACS-IW-00558437  100639-1-Centricity PACS-IW-00558450  4261-1-Centricity PACS-IW-00558480  101587-1-Centricity PACS-IW-00558369  4614-1-Centricity PACS-IW-00558393  4376-1-Centricity PACS-IW-00558399  4436-1-Centricity PACS-IW-00558417  4350-13-Centricity PACS-IW-00558296  4680-1-Centricity PACS-IW-00558360  100666-1-Centricity PACS-IW-00558465  101171-1-Centricity PACS-IW-00558310  101792-1-Centricity PACS-IW-00558370  4264-1-Centricity PACS-IW-00558254  4286-1-Centricity PACS-IW-00558361  4284-1-Centricity PACS-IW-00558453  4605-1-Centricity PACS-IW-00558350  4446-2-Centricity PACS-IW-00558106  4280-1-Centricity PACS-IW-00558240  12830-1-Centricity PACS-IW-00558312  100580-1-Centricity PACS-IW-00558339  100858-1-Centricity PACS-IW-00558426  100674-1-Centricity PACS-IW-00558315  102206-1-Centricity PACS-IW-00558447  4684-1-Centricity PACS-IW-00558245  4256-1-Centricity PACS-IW-00558250  4707-1-Centricity PACS-IW-00558434  100783-1-Centricity PACS-IW-00558255  4263-1-Centricity PACS-IW-00558279  100970-1-Centricity PACS-IW-00558303  4689-1-Centricity PACS-IW-00558242  100767-1-Centricity PACS-IW-00558377  100766-1-Centricity PACS-IW-00558355  100764-1-Centricity PACS-IW-00558452  100765-1-Centricity PACS-IW-00558331  4350-1-Centricity PACS-IW-00558301  4350-1-Centricity PACS-IW-00558307  4350-1-Centricity PACS-IW-00558318  4350-1-Centricity PACS-IW-00558432  4109-1-Centricity PACS-IW-00558389  4485-1-Centricity PACS-IW-00558479  100485-1-Centricity PACS-IW-00558258  100759-1-Centricity PACS-IW-00558268  4687-1-Centricity PACS-IW-00558334  1005-1-Centricity PACS-IW-00558282  1005-1-Centricity PACS-IW-00558290  100836-1-Centricity PACS-IW-00558256  4311-1-Centricity PACS-IW-00558321  100757-1-Centricity PACS-IW-00558261  4679-1-Centricity PACS-IW-00558412  100718-1-Centricity PACS-IW-00558424  13332-1-Centricity PACS-IW-00558390  4294-1-Centricity PACS-IW-00558379  13360-1-Centricity PACS-IW-00558263  12908-1-Centricity PACS-IW-00558302  4359-1-Centricity PACS-IW-00558410  13039-1-Centricity PACS-IW-00558385  100166-1-Centricity PACS-IW-00558448  9757-1-Centricity PACS-IW-00558247  1412-1-Centricity PACS-IW-00558228  4274-1-Centricity PACS-IW-00558436  100156-1-Centricity PACS-IW-00558243  100156-1-Centricity PACS-IW-00558475  4723-1-Centricity PACS-IW-00558367  4539-1-Centricity PACS-IW-00558378  101042-1-Centricity PACS-IW-00558386  3189-1-Centricity PACS-IW-00558314  4495-1-Centricity PACS-IW-00558476  4734-1-Centricity PACS-IW-00558375  100701-1-Centricity PACS-IW-00558322  100722-1-Centricity PACS-IW-00558407  2035-1-Centricity PACS-IW-00558411  101742-1-Centricity PACS-IW-00558470  104648-2-Centricity PACS-IW-00558406  4670-1-Centricity PACS-IW-00558329  100855-1-Centricity PACS-IW-00558298  4385-1-Centricity PACS-IW-00558336  4260-1-Centricity PACS-IW-00558244  100065-1-Centricity PACS-IW-00558317  4295-1-Centricity PACS-IW-00558341  13209-1-Centricity PACS-IW-00558357  4676-1-Centricity PACS-IW-00558449  13295-1-Centricity PACS-IW-00558346  9791-1-Centricity PACS-IW-00558418  4273-1-Centricity PACS-IW-01225873  4269-1-Centricity PACS-IW-00558252  100589-1-Centricity PACS-IW-00558384  13337-1-Centricity PACS-IW-00558401  13337-1-Centricity PACS-IW-00558478  100876-1-Centricity PACS-IW-00558440  11764-1-Centricity PACS-IW-00558388  101689-1-Centricity PACS-IW-00558326  9764-1-Centricity PACS-IW-00558394  100835-1-Centricity PACS-IW-00558287  100831-1-Centricity PACS-IW-00558232  4744-1-Centricity PACS-IW-00558373  100772-1-Centricity PACS-IW-00558269  4281-1-Centricity PACS-IW-00558354  12960-1-Centricity PACS-IW-00558289  9978-3-Centricity PACS-IW-00558108  1402-1-Centricity PACS-IW-00558291  4321-1-Centricity PACS-IW-00558371  4381-1-Centricity PACS-IW-00558423  1451-1-Centricity PACS-IW-00558294  13175-1-Centricity PACS-IW-00558383  10467-1-Centricity PACS-IW-00558405  10911-1-Centricity PACS-IW-00558471  100821-1-Centricity PACS-IW-00558325  4425-1-Centricity PACS-IW-00558472  4268-1-Centricity PACS-IW-00558416  9844-1-Centricity PACS-IW-00558332  9767-1-Centricity PACS-IW-00558265  4305-1-Centricity PACS-IW-00558337  4367-1-Centricity PACS-IW-00558415  4128-1-Centricity PACS-IW-00558234  101098-1-Centricity PACS-IW-00558387  4322-1-Centricity PACS-IW-00558285  4352-1-Centricity PACS-IW-00558316  4374-1-Centricity PACS-IW-00558422  100827-1-Centricity PACS-IW-00558233  4360-1-Centricity PACS-IW-00558237  4335-1-Centricity PACS-IW-00558267  4349-1-Centricity PACS-IW-00558284  4211-1-Centricity PACS-IW-00558277  3545-1-Centricity PACS-IW-00558239  10527-3-Centricity PACS-IW-00558344  4348-1-Centricity PACS-IW-00558458  4362-1-Centricity PACS-IW-00558395  4576-1-Centricity PACS-IW-00558306  10916-1-Centricity PACS-IW-00558299  4288-1-Centricity PACS-IW-00558235  4266-1-Centricity PACS-IW-00558304  100511-1-Centricity PACS-IW-00558549  1565-1-Centricity PACS-IW-00603058  4277-1-Centricity PACS-IW-00558474  4553-1-Centricity PACS-IW-00558327  101216-1-Centricity PACS-IW-00558356  4531-1-Centricity PACS-IW-00558376  101240-1-Centricity PACS-IW-00558425  100021-1-Centricity PACS-IW-00558260  4660-1-Centricity PACS-IW-00558288  2481-1-Centricity PACS-IW-00558381  4447-1-Centricity PACS-IW-00558398  4078-1-Centricity PACS-IW-00558462  100806-1-Centricity PACS-IW-00558441  4361-1-Centricity PACS-IW-00558343  13164-1-Centricity PACS-IW-00558439  4504-1-Centricity PACS-IW-00558438  4413-1-Centricity PACS-IW-00558433  4416-1-Centricity PACS-IW-00558292  4416-1-Centricity PACS-IW-00558293  4417-1-Centricity PACS-IW-00558364  4419-1-Centricity PACS-IW-00558483  4420-1-Centricity PACS-IW-00558404  100378-1-Centricity PACS-IW-00558238  4508-1-Centricity PACS-IW-00558271  100603-1-Centricity PACS-IW-00558266  10462-1-Centricity PACS-IW-00558272  12953-1-Centricity PACS-IW-00558443  10483-1-Centricity PACS-IW-00558098  1331-1-Centricity PACS-IW-00558451  12949-1-Centricity PACS-IW-00558442  4296-1-Centricity PACS-IW-00558477  100088-1-Centricity PACS-IW-00558320  4351-1-Centricity PACS-IW-00558485  1584-1-Centricity PACS-IW-00558297  1173-1-Centricity PACS-IW-00558473  4450-1-Centricity PACS-IW-00558319  100972-1-Centricity PACS-IW-00558278  4276-1-Centricity PACS-IW-00558274  101647-1-Centricity PACS-IW-00558515  11711-2-Centricity PACS-IW-00558262  4542-1-Centricity PACS-IW-00558430  13265-1-Centricity PACS-IW-00558482  12974-1-Centricity PACS-IW-00558300  100770-1-Centricity PACS-IW-00558097  100769-1-Centricity PACS-IW-00558427  4393-1-Centricity PACS-IW-00558328  13372-1-Centricity PACS-IW-00558309  13197-1-Centricity PACS-IW-00558251  11042-1-Centricity PACS-IW-00558368  100468-1-Centricity PACS-IW-00558330  4575-1-Centricity PACS-IW-00558342  12931-1-Centricity PACS-IW-00558419  4482-1-Centricity PACS-IW-00558281  4568-1-Centricity PACS-IW-00558280  3571-1-Centricity PACS-IW-00558468  100645-1-Centricity PACS-IW-00558408  100771-1-Centricity PACS-IW-00558249  4632-1-Centricity PACS-IW-00558311  4681-1-Centricity PACS-IW-00558431  4573-1-Centricity PACS-IW-00558420  100729-1-Centricity PACS-IW-00558455  4456-1-Centricity PACS-IW-00558295  4506-1-Centricity PACS-IW-00558345  BA4080-Centricity PACS-IW-00038769  324462-Centricity PACS-IW-01431554  293035-Centricity PACS-IW-01401532  280801-Centricity PACS-IW-01431490  335279-Centricity PACS-IW-01431638  306930-Centricity PACS-IW-01401530  296619-Centricity PACS-IW-01401506  330917-Centricity PACS-IW-01401529  SiteID-Centricity PACS-IW-01422991  313741-Centricity PACS-IW-01422998  238110-Centricity PACS-IW-01431521  246838-Centricity PACS-IW-01421153  134265-Centricity PACS-IW-01431539  307319-Centricity PACS-IW-01444609  230048-Centricity PACS-IW-01442018  SiteID-Centricity PACS-IW-01444879  209545-Centricity PACS-IW-01460552  EG1856-Centricity PACS-IW-00118541  EG1009-Centricity PACS-IW-00118535  EG1871-Centricity PACS-IW-00118532  EG1905-Centricity PACS-IW-00117725  EG1771-Centricity PACS-IW-00118550  EG1858-Centricity PACS-IW-00118544  EG1808-Centricity PACS-IW-00118538  EG1857-Centricity PACS-IW-00118547  M41678-Centricity PACS-IW-00043697  M41741-Centricity PACS-IW-00043705  M42172-Centricity PACS-IW-00165182  683198-Centricity PACS-IW-00000826  600791-Centricity PACS-IW-00000776  UHAM02-Centricity PACS-IW-00001927  RMUE01-Centricity PACS-IW-00001801  PC0125-Centricity PACS-IW-00001683  HU1025-Centricity PACS-IW-00038776  HU1469-Centricity PACS-IW-00038785  HU1038-Centricity PACS-IW-00038778  HU1093-Centricity PACS-IW-00038784  HU1397-Centricity PACS-IW-00038782  HU1185-Centricity PACS-IW-00038781  HU1433-Centricity PACS-IW-00038783  HU1008-Centricity PACS-IW-00038775  HU1036-Centricity PACS-IW-00038777  HU1041-Centricity PACS-IW-00038780  HU1039-Centricity PACS-IW-00038779  ApolloBSR-Centricity PACS-IW-01819452  StThomas-Centricity PACS-IW-01819447  HOLY-FAMILY-THODU-Centricity PACS-IW-01821493  SODANI-Centricity PACS-IW-01819445  WOCKHARDT-Centricity PACS-IW-01819449  DAKE-Centricity PACS-IW-01819448  SIMS-Centricity PACS-IW-01819453  MARTHANDAM-Centricity PACS-IW-01819446  CTSI-Centricity PACS-IW-01819444  CONTINENTAL-Centricity PACS-IW-01819443  197240-Centricity PACS-IW-01442043  KW1112-Centricity PACS-IW-00168151  LB1331-Centricity PACS-IW-00116959  REGENCY-Centricity PACS-IW-01286180  BEVERLY-Centricity PACS-IW-01288659  MAHKOTA-Centricity PACS-IW-01286179  UMUM-Centricity PACS-IW-01288666  5791-Centricity PACS-IW-01442181  324271-Centricity PACS-IW-01970711  AMS39-Centricity PACS-IW-00044104  PL1731-Centricity PACS-IW-00038835  PL1517-Centricity PACS-IW-00038837  PL1515-Centricity PACS-IW-00038839  PL1394-Centricity PACS-IW-00038841  PL1639-Centricity PACS-IW-00038843  PL1426-Centricity PACS-IW-00038845  PL1491-Centricity PACS-IW-00038847  PL2619-Centricity PACS-IW-00038814  PL1631-Centricity PACS-IW-00038849  PL1684-Centricity PACS-IW-00038851  PL1942-Centricity PACS-IW-00038853  PL1825-Centricity PACS-IW-00038819  PL1505-Centricity PACS-IW-00038855  PL3249-Centricity PACS-IW-00038865  PL3249-Centricity PACS-IW-00038868  PL3249-Centricity PACS-IW-00038870  PL3249-Centricity PACS-IW-00038871  PL1530-Centricity PACS-IW-00038822  PL2762-Centricity PACS-IW-00038828  PL1544-Centricity PACS-IW-00038859  PL1644-Centricity PACS-IW-00038861  PL1650-Centricity PACS-IW-00038857  PL1508-Centricity PACS-IW-00038863  QA1015-Centricity PACS-IW-00115165  RU3382-Centricity PACS-IW-01273247  RU3680-Centricity PACS-IW-01273260  RU3380-Centricity PACS-IW-01273246  SA2187-Centricity PACS-IW-00065875  SA2294-Centricity PACS-IW-00065843  SA1198-Centricity PACS-IW-00065868  SA1175-Centricity PACS-IW-00038187  SiteID-Centricity PACS-IW-02266272  MPX352-Centricity PACS-IW-01973993  DP7603-Centricity PACS-IW-00045307  UP506-Centricity PACS-IW-00044759  529010-Centricity PACS-IW-00000342  ONCOLOGY-Centricity PACS-IW-01288658  115-Centricity PACS-IW-01288657    c) 2090228-002  Lot or Serial Numbers:  3640-1-Centricity PACS-IW-00558511  4345-1-Centricity PACS-IW-00558548  4430-1-Centricity PACS-IW-00558500  4378-1-Centricity PACS-IW-00558498  100663-1-Centricity PACS-IW-00558514  100674-1-Centricity PACS-IW-00558486  9754-1-Centricity PACS-IW-00558490  4429-1-Centricity PACS-IW-00558499  4328-1-Centricity PACS-IW-00558518  4552-1-Centricity PACS-IW-00558510  4387-1-Centricity PACS-IW-00558493  100702-1-Centricity PACS-IW-00558539  100717-1-Centricity PACS-IW-00558517  101625-1-Centricity PACS-IW-00558531  4593-1-Centricity PACS-IW-00558488  100009-1-Centricity PACS-IW-00558560  4354-1-Centricity PACS-IW-00558519  100838-1-Centricity PACS-IW-00558532  4551-1-Centricity PACS-IW-00558508  100199-1-Centricity PACS-IW-00558503  12968-1-Centricity PACS-IW-00558506  13102-1-Centricity PACS-IW-00558512  4379-1-Centricity PACS-IW-00558496  4366-1-Centricity PACS-IW-00558542  4694-1-Centricity PACS-IW-00558556  1635-1-Centricity PACS-IW-00558557  100881-1-Centricity PACS-IW-00558546  4299-2-Centricity PACS-IW-00558545  10884-2-Centricity PACS-IW-00558100  101883-1-Centricity PACS-IW-00558487  100852-1-Centricity PACS-IW-00558494  4368-1-Centricity PACS-IW-00558513  4668-1-Centricity PACS-IW-00558529  100811-1-Centricity PACS-IW-00558535  4497-1-Centricity PACS-IW-00558497  4682-1-Centricity PACS-IW-00558270  100458-1-Centricity PACS-IW-00558538  4358-1-Centricity PACS-IW-00558564  13095-1-Centricity PACS-IW-00558534  13064-2-Centricity PACS-IW-00558102  4372-1-Centricity PACS-IW-00558523  4326-4-Centricity PACS-IW-00558105  101932-1-Centricity PACS-IW-00558495  4301-1-Centricity PACS-IW-00558525  10750-1-Centricity PACS-IW-00558536  100761-1-Centricity PACS-IW-00558527  10749-1-Centricity PACS-IW-00558509  2387-1-Centricity PACS-IW-00558489  10250-1-Centricity PACS-IW-00558504  100802-1-Centricity PACS-IW-00558540  101193-1-Centricity PACS-IW-00558551  1475-1-Centricity PACS-IW-00558558  12821-1-Centricity PACS-IW-00558526  4661-1-Centricity PACS-IW-00558528  2481-1-Centricity PACS-IW-00558524  3943-1-Centricity PACS-IW-00558501  4299-2-Centricity PACS-IW-02180909  13134-1-Centricity PACS-IW-00558492  4157-1-Centricity PACS-IW-00558516  4383-1-Centricity PACS-IW-00558533  13233-1-Centricity PACS-IW-00558554  4389-1-Centricity PACS-IW-00558563  100582-1-Centricity PACS-IW-00558530  4179-1-Centricity PACS-IW-00558543  DZ1004-Centricity PACS-IW-00116835  BH1038-Centricity PACS-IW-00162931  80083-Centricity PACS-IW-00043981  SiteID-Centricity PACS-IW-01431505  332921-Centricity PACS-IW-01442042  313689-Centricity PACS-IW-01431524  SiteID-Centricity PACS-IW-01401523  000341097-Centricity PACS-IW-01431636  191244-Centricity PACS-IW-01422989  298983-Centricity PACS-IW-01401548  293834-Centricity PACS-IW-01422975  302144-Centricity PACS-IW-01421157  32125-Centricity PACS-IW-01444621  13628-Centricity PACS-IW-01422992  213537-Centricity PACS-IW-01431510  000282202-Centricity PACS-IW-01422973  319674-Centricity PACS-IW-01431535  369946-Centricity PACS-IW-01264977  218314-Centricity PACS-IW-01434509  248553-Centricity PACS-IW-01442041  EG1017-Centricity PACS-IW-00117723  X2575258-Centricity PACS-IW-00043952  M43994-Centricity PACS-IW-00043773  M60364-Centricity PACS-IW-00043899  X1413896-Centricity PACS-IW-00043945  TESTSITE-Centricity PACS-IW-01986381  714235-Centricity PACS-IW-00000842  972439-Centricity PACS-IW-00000859  363691-Centricity PACS-IW-00000520  424881-Centricity PACS-IW-00000556  A2513980-Centricity PACS-IW-00040355  A56306-Centricity PACS-IW-00040439  PUSANPAIK-Centricity PACS-IW-01819440  CHILGOK-Centricity PACS-IW-01819441  KANGBUK-Centricity PACS-IW-01819442  1971565-YUHSSC-RAD-Centricity PACS-IW-01819454  KW1003-Centricity PACS-IW-01798078  KW1019-Centricity PACS-IW-00115173  KW1018-Centricity PACS-IW-01422962  KW1001-Centricity PACS-IW-00164288  KW1096-Centricity PACS-IW-00166324  SSH-Centricity PACS-IW-01288660  NG1043-IW w UW Base SW PVP-00158638  MYHEALTH-Centricity PACS-IW-01819534  StLukes-Centricity PACS-IW-01286216  PL2059-Centricity PACS-IW-00038789  PL2997-Centricity PACS-IW-00038793  PL1803-Centricity PACS-IW-00038812  PL2823-Centricity PACS-IW-00038813  PL2342-Centricity PACS-IW-00038815  PL1666-Centricity PACS-IW-00038816  PL2257-Centricity PACS-IW-00038817  PL1961-Centricity PACS-IW-00038818  PL3092-Centricity PACS-IW-00038830  PL1006-Centricity PACS-IW-00038821  PL1822-Centricity PACS-IW-00038825  PL3362-Centricity PACS-IW-00038833  PL1951-Centricity PACS-IW-00038827  RU1016-Centricity PACS-IW-01273193  RU1241-Centricity PACS-IW-01273216  RU1493-Centricity PACS-IW-01273220  RU2967-Centricity PACS-IW-01273244  RU5302-Centricity PACS-IW-01273274  RU2420-Centricity PACS-IW-01273234  RU3910-Centricity PACS-IW-01273270  RU3262-Centricity PACS-IW-01273245  RU6248-Centricity PACS-IW-01273279  RU1648-Centricity PACS-IW-01273222  RU8071-Centricity PACS-IW-01273295  RU6340-Centricity PACS-IW-01273280  RU6340-Centricity PACS-IW-01273282  RU7576-Centricity PACS-IW-01273287  RU3880-Centricity PACS-IW-01273266  RU3894-Centricity PACS-IW-01273269  RU3564-Centricity PACS-IW-01273252  RU8162-Centricity PACS-IW-01273299  RU2527-Centricity PACS-IW-01273235  RU5310-Centricity PACS-IW-01273276  RU3959-Centricity PACS-IW-01273273  RU2878-Centricity PACS-IW-01273243  RU8370-Centricity PACS-IW-01273306  RU3572-Centricity PACS-IW-01273255  600160-Centricity PACS-IW-01273191  RU3625-Centricity PACS-IW-01273258  RU5499-Centricity PACS-IW-01273278  RU1469-Centricity PACS-IW-01273219  RU1199-Centricity PACS-IW-01273215  RU1717-Centricity PACS-IW-01273225  RU9000-Centricity PACS-IW-01273312  RU8041-Centricity PACS-IW-01273294  RU1857-Centricity PACS-IW-01273229  RU8219-Centricity PACS-IW-01273302  RU1279-Centricity PACS-IW-01273217  RU1095-Centricity PACS-IW-01273200  RU7000-Centricity PACS-IW-01273281  RU8321-Centricity PACS-IW-01273303  RU7563-Centricity PACS-IW-01273285  RU1130-Centricity PACS-IW-01273210  RU2371-Centricity PACS-IW-01273233  RU7426-Centricity PACS-IW-01273284  RU2167-Centricity PACS-IW-01273231  RU3890-Centricity PACS-IW-01273268  RU3828-Centricity PACS-IW-01273262  RU1669-Centricity PACS-IW-01273224  RU5308-Centricity PACS-IW-01273275  RU8953-Centricity PACS-IW-01273311  RU2367-Centricity PACS-IW-01273232  RU1026-Centricity PACS-IW-01273195  RU3476-Centricity PACS-IW-01273251  RU1050-Centricity PACS-IW-01273198  RU1050-Centricity PACS-IW-00167296  RU7564-Centricity PACS-IW-01273286  RU8036-Centricity PACS-IW-01273293  RU3570-Centricity PACS-IW-01273254  RU3610-Centricity PACS-IW-01273257  RU3794-Centricity PACS-IW-01273261  RU8192-Centricity PACS-IW-01273300  RU8192-Centricity PACS-IW-01273301  RU3832-Centricity PACS-IW-01273263  RU1144-Centricity PACS-IW-01273211  600100-Centricity PACS-IW-01273188  RU3878-Centricity PACS-IW-01273265  RU1028-Centricity PACS-IW-01273197  RU3925-Centricity PACS-IW-01273272  RU1055-Centricity PACS-IW-01273199  RU3919-Centricity PACS-IW-01273271  RU1022-Centricity PACS-IW-01273194  RU5442-Centricity PACS-IW-01273277  RU3883-Centricity PACS-IW-01273267  RU3467-Centricity PACS-IW-01273250  RU3663-Centricity PACS-IW-01273259  RU7905-Centricity PACS-IW-01273291  RU3593-Centricity PACS-IW-01273256  RU1298-Centricity PACS-IW-01273218  RU8151-Centricity PACS-IW-01273296  RU7838-Centricity PACS-IW-01273289  RU7838-Centricity PACS-IW-01273290  RU1152-Centricity PACS-IW-01273212  RU8607-Centricity PACS-IW-01273308  RU3569-Centricity PACS-IW-01273253  SA2128-Centricity PACS-IW-00065800  SA1126-Centricity PACS-IW-00065887  SA2235-Centricity PACS-IW-00038164  SA2221-Centricity PACS-IW-00065837  ZA2408-Centricity PACS-IW-00116590  ZA2339-Centricity PACS-IW-00116589  ZA2356-Centricity PACS-IW-00116576  ZA2355-Centricity PACS-IW-00116575  ZA2338-Centricity PACS-IW-00116588  ZA2331-Centricity PACS-IW-00116443  ZA2335-Centricity PACS-IW-00116447  ZA2330-Centricity PACS-IW-00116442  ZA2334-Centricity PACS-IW-00116446  ZA2409-Centricity PACS-IW-00116591  ZA2410-Centricity PACS-IW-00116592  ZA2336-Centricity PACS-IW-00116448  ZA2470-Centricity PACS-IW-00116572  ZA2352-Centricity PACS-IW-00116570  ZA2351-Centricity PACS-IW-00116569  ZA2474-Centricity PACS-IW-00116763  ZA2471-Centricity PACS-IW-00116573  YP5673-Centricity PACS-IW-00045220  941804-Centricity PACS-IW-00000474    d) 2052782-00x  Lot or Serial Numbers:  100884-1-Centricity PACS-IW-00558222  100832-1-Centricity PACS-IW-00558224  100828-1-Centricity PACS-IW-00558231  100954-1-Centricity PACS-IW-00558225  100826-1-Centricity PACS-IW-00558229  4595-1-Centricity PACS-IW-00558226  4290-1-Centricity PACS-IW-00558217  4364-1-Centricity PACS-IW-00558219  4524-1-Centricity PACS-IW-00558218  4536-1-Centricity PACS-IW-00558216"
    }
  ]
}