  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76805",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "United Kingdom, United Arab Emirates, Turkey, Thailand, Switzerland, Sweden,  Spain, South Africa, Slovenia, Saudi Arabia, Russia, Romania, Portugal, Poland,  Philippines, Norway, New Zealand, New Caledonia, Netherlands, Myanmar,   Kuwait, Korea (Republic of), Japan, Italy, ISRAEL, Ireland, Iran, India, Hungary,   Hong Kong, Guatemala, Greece, Great Britain, Germany, France, Finland,   Estonia, Egypt, Ecuador, Denmark, Czech Republic, Croatia, Colombia,   Chile, Canada, Bulgaria, Brunei Darussalam, Brazil, Belgium, Austria,   Australia, Argentina, Vietnam.  Uruguay, Taiwan, Slovakia, Singapore, Reunion, Qatar, Panama, Pakistan, Jordan,   Indonesia, Guadeloupe, Czech Republic",
      "recall_number": "Z-1780-2017",
      "product_description": "Vivid S60/S70/S60N/S70N, H45041SU, H45041SW, H45581MS, H45581PD.",
      "product_quantity": "396 (67 US; 329 OUS)",
      "reason_for_recall": "GE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by the operator in situations where DICOM Worklist search response time is slow. This issue is limited to certain Vivid Ultrasound systems and can result in the incorrect patient information showing on the screen during the exam. If not detected by the operator, images from the actual (intended) patient will be stored under the incorrect patient after the exam. This issue could lead to misdiagnosis. There have been no injuries reported because of this issue.",
      "recall_initiation_date": "20170310",
      "center_classification_date": "20170411",
      "termination_date": "20180312",
      "report_date": "20170419",
      "code_info": "Mfg. Lot or Serial # System ID  000955S60 VS60000955  000973S70 VS70000973  000974S70 VS7000974  000997S70 VS70000997  001022S70 VS70001022  001020S70 VS70001020  000941S70 VS70000941  000951S70 VS70000951  000982S70 VS70000982  000981S70 VS70000981  000994S70 VS70000994  000976S70 VS70R000976  000993S70 VS70000993  000980S70 VS70000980  000998S70 VS70R000998  000996S70 VS70000996  001005S70 VS70001005  001010S70 VS70001010  001007S70 VS70001007  001008S70 VS70001008  001021S70 VS70001021  001029S70 VS70001029  001040S60 VS60001040  001052S70 VS70001052  001048S70 VS70001048  001051S70 VS70001051  001066S60 VS60001066  001054S60 VS60001054  001059S70 VS70001059  001064S70 VS70001064  001067S70 VS70001067  001062S70 VS70001062  001047S70 VS70001047  001050S70 VS70001050  001073S70 VS70001073  001076S70 VS70001076  001078S70 VS70001078  001074S70 VS70001074  001129S70 VS70001129  001120S60 VS6001120  001137S70 VS70001137  001157S70 VS70001157  001230S60 VS60001230  001246S60 VS60001246  001253S60 VS6001253  001336S60 VS6001336  001345S60 VS60001345  000948S70 VS70000948  000995S70 VS70000995  001053S70 VS70001053  001055S60 VS60001055  001056S70 VS70001056  001039S70 VS70001039  001060S70 VS70001060  001069S70 VS70001069  001112S60 VS60001112  001128S70 VS70001128  001138S70 VS70001138  001163S70 VS70001163  001165S70 VS70001165  001238S60 Not Available  001256S60 VS60001256  001003S70 VS70001003  001332S60 VS60001332  001347S60 VS60001347  001329S60 VS60001329  001310S60 Not Available  001250S70 Not Available  000969S70 VS70969S70  000971S70 00123ULT34  000966S70 00111ULT15  001023S70 00111ULT14  001046S70 00982ULT02  001036S70 00506ULT19  001072S70 20347ULT03  001117S60 00058ULT15  001175S70 00115ULT16  001298S70 00149ULT44  001340S70 00004ULT17  001150S70 00199ULT36  001264S70 00199ULT39  001260S70 00199ULT37  001254S70 00199ULT36  000978S70 00459ULT17  001322S70 00624ULT30  001307S70 00136ULT11  001309S70 21367ULT01  001354S60 00195ULT26  001337S70 00296ULT13  001002S70 0853380377  001182S70 529040US21  001325S70 Not Available  001114S70 A004US71  000960S70 O009US94  001113S70 A042US13  001140S70 BD003US38  001171S70 A042US14  001190S70 MPX43159  001174S60 Not Available  001146S60 GON4457540  001172S60 GON4422820  001143S70 C99999284  1104S70N ZA2785US01  1106S70N ZA2790US01  1103S70N ZA2820US01  1198S60N ZA2945US01  592S70N ZA2820US02  1107S70N Not Available  001417S60 GON4445496  001400S70 6073UL0319  1366S60N SA2517US01  1194S60N  SA2511US01   001242S70 RU7950US05  001231S70 RU1173US06  001257S70 RU6396US07  001393S70 600152US07  001042S60 Not Available  001034S60 RO4186US12  001151S60 RE1110US01  1105S60N QA1046US02  001115S60 PER46108  001167S70 PL1698US08  001080S70 PL7475US03  001215S70 PL1334US09  001192S60 1557UL0338  000977S70 VS70977S70  1368S60N GON4436983  001173S70 Not Available  001279S70 Not Available  001099S70 Not Available  001179S60 Not Available  001234S60 Not Available  001224S60 Not Available  001324S60 Not Available  1200S70N Not Available  000953S60 NO4005U120  000970S70 NO4031US24  001000S70 NO4068US85  001169S70 NO4088US60  001218S60 NO4202US03  001316S60 NO4442US01  001221S60 8315344197  001012S60 0850372938  001093S70 0850372991  001108S70 0850372993  001098S70 0850372992  001265S70 0850373034  001284S60 0850373074  001311S60 0850373460  001312S70 0850373120  001313S70 0850373117  001326S70 0850373109  1195S60N JO4041US03  1196S70N JO4041US04  000944S60 QEA060  001011S60 QEA061  001014S60 QEA062  001016S60 QEA064  001015S60 QE0063  001085S60 QE0065  001092S60 QE0066  001203S70 QE0067  001239S60 UX9652  001247S60 QE0069  001233S60 QE0073  001248S60 QE0075  001245S60 QE0074  001244S70 QE0071  001249S70 QE0072  001227S70 QE0070  001261S60 QE0076  001286S70 QE0078  001275S70 QE0077  001282S60 QE0080  001258S70 QE0079  001304S70 QE0084  001341S60 QE0083  001328S60 QE0082  001299S60 QE0081  001292S70 Not Available  001177S70 A470456401  001369S70 A5935902  001394S70 A5289906  000029S70 A5113139  001033S70 Not Applicable  907S60N IR2197US01  1367S60N 5512UL0905  001152S70 083038820115216  001152S70 083038820115216  001281S70 083038800060616  001213S70 083038800148317/083038801699017  001327S70 083038804905016  986S70N 083038604420116  001083S70 HUDUS0525  001084S70 HUDUS0527  001095S70 HUDUS0526  001094S60 HUDUS0123  001243S70 HUDUS0420  001241S70 HUDUS0421  001396S70 HUDUS0489  001387S60 HUDUS0183  001412S60 HUDUS0184  001407S70 HUDUS0182  001131S70 HUDUS0354  001330S70 0828380483  001097S60 GP1112US01  001216S60 H2105US04  001216S60 H2105US04  001305S70 H22421US02  001270S70 Not Available  001331S70 Not Available  000975S70 142541U076  001018S60 110012U092  001025S60 110012U093  001028S60 TP1098U001  001043S60 110012U090  001086S60 4683234U01  001090S70 4633351U01  001135S60 802641U005  001136S60 2431497U02  001142S70 110826U019  001160S60 1771688U04  001291S60 149712U007  001306S60 2551720U03  001252S60 145424U104  001274S60 104462U006  001333S70 TP9048U024  001350S70 151176U012  001395S70 110916U011  001403S60 110459U036  001164S60 148015U004  001166S60 147739U005  001259S70 106018U004  001263S70 2103790U24  001024S60 3650585U02  001026S60 1279706U02  001049S60 356393U005  001041S60 138050U008  001070S70 4209769U01  001013S70 120139U054  001081S60 1702988U05  001087S60 4697185U01  001071S70 143083U004  001089S70 4678439U01  001088S70 112132U011  001100S70 141114U056  001096S70 138792U029  001110S70 100007U027  001119S70 138251U024  001130S70 110910U050  001134S70 145905U014  001121S60 148265U004  001132S60 122009U079  001133S70 3686201U04  001159S60 4731568U01  001161S70 110173U024  001158S60 124001U002  001154S70 1220242U05  001149S70 TP0575US04  001162S70 4678439U02  001214S60 4731568U01  001212S60 2088742U03  001236S60 800524U005  001297S60 4208447U03  001294S60 140141U034  001301S60 110596U013  001251S70 145766U006  001288S70 145766U007  001293S70 1898210U03  001295S60 149847U024  001303S60 TP2467U039  001289S60 1893276U01  001302S60 4700381U01  001296S60 4208447U02  001318S60 104096U004  001300S60 4701723U03  001323S60 105843U003  001321S60 4701723U01  001365S70 1951144U28  001374S70 110997U013  001356S70 110146U008  001383S70 1951144U27  001397S70 122006U085  001382S70 4695527U01  001401S70 110855U016  001402S60 110632U004  001406S60 4762251U01  001408S60 4762250U01  001398S60 145430U007  001413S60 804380U006  001418S70 110499U032  001201S60 4711307U01  000934S70 110162U025  000963S70 145924U004  001272S70 120053U024  001268S70 2017571U33  000947S60 Not Available  000946S60 Not Available  000956S60 Not Available  000958S60 Not Available  001148S60 GON4423471  001183S60 U437830301  001185S60 U437830302  001220S60 US30003301  001222S70 U464557301  001315S60 US31001301  001314S60 U470803601  000937S60 US78044607  001038S70 Not Applicable  000957S60 U287989802  000961S60 USM4198037  000959S60 USM4198036  000967S60 US29066206  000968S60 US88657318  001031S70 US26187813  001035S70 USM2505607  001109S60 US50272418  001127S70 US88657317  001126S70 USM4177308  001111S60 US50272417  001144S70 USN4165408  001147S70 U151064604  001153S70 U151064605  001189S60 USM2505317  001186S70 U276410807  001178S70 USN4165409  001204S70 USM4166011  001168S60 001168S60  001219S70 USM3183518  001228S70 USM4329426  001232S60 USM4477110  001240S60 USN4213405  001225S70 U243851504  001271S60 USM4497217  001267S70 USM2607019  001278S70 USM2607020  001280S70 USM4482808  001283S70 US67382203  001287S60 US41892022  001290S60 U282272013  001392S70 USN4165410  001379S70 USM4490608  001376S70 USM4199322  001364S70 USM4497218  001391S60 U222842502  001421S70 US78034314  001191S60 U239390103  001206S60 UN88016207  000962S60 US98957510  001141S60 US60815807  000938S70 US29072605  000933S70 GON4457410  000939S60 Not Available  001217S70 FI1063US29  001223S70 FI1032US07  001229S70 FI1063US28  000979S70 DK1022US54  000954S70 Not Available  001255S70 DK1090US21  001320S70 DK1058US42  001262S70 DK1022US58  001027S60 CZ6403US02  001317S70 CZ6001US09  000999S70 416778VS70  001037S70 416494VS70  001335S70 416446VS70  001370S70 Not Available  001202S60 VS601202S60  000972S70 030042US13  001102S70 030392US15  001235S60 030419US12  001237S60 030419US13  001386S70 030392US17  001030S70 947667US02  001032S60 568080US02  000753S70 100029US11  001205S70 220166US25  001208S70 190105US28  001416S70 701697US01  001207S60 190171US13  001170S70 701708US01  001388S60 Not Available  001385S70 Not Available  001410S70 Not Available  001226S70 AH1912US32  001273S70 AH5045US11  000952S70 83150625712  001017S70 83151375114  001123S70 83152818712  001180S70 83152648280  001277S70 83154415380  001269S70 83153232163  001276S70 83154414988  001285S70 83154415943  001101S70 Not Available  001193S60 Not Available",
      "more_code_info": ""
    }
  ]
}