  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76805",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "United Kingdom, United Arab Emirates, Turkey, Thailand, Switzerland, Sweden,  Spain, South Africa, Slovenia, Saudi Arabia, Russia, Romania, Portugal, Poland,  Philippines, Norway, New Zealand, New Caledonia, Netherlands, Myanmar,   Kuwait, Korea (Republic of), Japan, Italy, ISRAEL, Ireland, Iran, India, Hungary,   Hong Kong, Guatemala, Greece, Great Britain, Germany, France, Finland,   Estonia, Egypt, Ecuador, Denmark, Czech Republic, Croatia, Colombia,   Chile, Canada, Bulgaria, Brunei Darussalam, Brazil, Belgium, Austria,   Australia, Argentina, Vietnam.  Uruguay, Taiwan, Slovakia, Singapore, Reunion, Qatar, Panama, Pakistan, Jordan,   Indonesia, Guadeloupe, Czech Republic",
      "recall_number": "Z-1779-2017",
      "product_description": "Vivid E95/E90/E80, H45581DC, H45581LB and H45581DA.",
      "product_quantity": "971(412 US; 559 OUS)",
      "reason_for_recall": "GE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by the operator in situations where DICOM Worklist search response time is slow. This issue is limited to certain Vivid Ultrasound systems and can result in the incorrect patient information showing on the screen during the exam. If not detected by the operator, images from the actual (intended) patient will be stored under the incorrect patient after the exam. This issue could lead to misdiagnosis. There have been no injuries reported because of this issue.",
      "recall_initiation_date": "20170310",
      "center_classification_date": "20170411",
      "termination_date": "20180312",
      "report_date": "20170419",
      "code_info": "Mfg. Lot or Serial # System ID  1 VE95R00100  AU00104 VE95R00104  AU00107 VE95R00107  AU00109 VE95R000109  AU00110 VE95R000110  AU00111 VE95R000111  AU00113 VE95R000113  AU00117 VE95R000117  AU00118 VE95R000118  AU00120 VE95000120  AU00121 VE95R000121  AU00123 VE95R000123  AU00124 VE95R000124  AU00125 VE95R000125  AU00126 VE90000126  AU00127 VE95R000127  AU00128 VE95R000128  AU00132 VE90000132  AU00133 VE95R000133  AU00135 VE90000135  AU00138 VE95R000138  AU00159 VE95R000159  AU00160 VE9500160  AU00162 VE95000162  AU00165 VE95000165  AU00168 VE95000168  AU00171 VE90000171  AU00172 VE95R000172  AU00173 VE90000173  AU00174 VE90000174  AU00176 VE95000176  AU00182 VE95000182  AU00183 VE90000183  AU00184 VE95R000184  AU00206 VE9500206  AU00207 VE95U00207  AU00211 VE9500211  AU00224 VE90000224  AU00230 VE95000230  AU00231 VE90000231  AU00232 VE90000232  AU00240 VE95R000240  AU00242 VE95000242  AU00243 VE90000243  AU00246 VE95V000246  AU00258 VE90000258  AU00277 VE95R000277  AU00281 VE95000281  AU00287 VE95000287  AU00300 VE95000300  AU00302 VE95R000302  AU00321 VE9500321  AU00329 VE95000329  AU00330 VE95000330  AU00331 VE9500331  AU00332 VE95000332  AU00335 VE95R000335  AU00339 VE9500339  AU00347 VE95000347  AU00348 VE95000348  AU00352 VE95000352  AU00359 VE9500359  AU00360 VE95000360  AU00361 VE95000361  AU00363 VE95000363  AU00367 VE95000367  AU00368 VE95000368  AU00369 VE95000369  AU00370 VE95000370  AU00372 VE95000372  AU00373 VE95000373  AU00374 VE9500374  AU00376 VE95R00376  AU00388 VE95000388  AU00396 VE95000396  AU00398 VE95000398  AU00401 VE95000401  AU00407 VE95R000407  AU00417 VE95000417  AU00426 VE95000426  AU00429 VE95000429  AU00430 VE95000430  AU00434 VE95000434  AU00438 VE9500438  AU00439 VE95000439  AU00441 VE95000441  AU00446 VE90000446  AU00448 VE95000448  AU00450 VE9500450  AU00454 VE95000454  AU00457 VE95000457  AU00460 VE95000460  AU00465 VE95000465  AU00466 VE95000466  AU00468 VE95000468  AU00474 VE95000474  AU00478 VE95000478  AU00482 VE95000482  AU00486 VE95000486  AU00488 VE90000488  AU00494 VE95000494  AU00499 VE95000499  AU00501 VE90000501  AU00505 VE95000505  AU00509 VE95000509  AU00512 VE95000512  AU00517 VE90000517  AU00521 VE90000521  AU00527 VE95000527  AU00531 VE9500531  AU00532 VE95000532  AU00550 VE95000550  AU00551 VE90000551  AU00552 VE95000552  AU00554 VE95000554  AU00565 VE95000565  AU00568 VE90000568  AU00571 VE9500571  AU00574 VE95000574  AU00575 VE95000575  AU00577 VE95000577  AU00580 VE95000580  AU00581 VE95000581  AU00582 VE90000582  AU00586 VE95000586  AU00590 VE95000590  AU00592 VE95000592  AU00593 VE90000593  AU00595 VE95000595  AU00598 VE95000598  AU00601 VE90000601  AU00602 VE95000602  AU00605 VE90000605  AU00608 VE9500608  AU00610 VE90000610  AU00616 VE90000616  AU00618 VE9500618  AU00621 VE95000621  AU00624 VE95000624  AU00625 VE95000625  AU00628 VE9000628  AU00630 VE95000630  AU00633 VE95000633  AU00661 VE95000661  AU00667 VE90000667  AU00684 VE95000684  AU00685 VE95R000685  AU00694 VE90000694  AU00700 VE95000700  AU00707 VE95000707  AU00714 VE95000714  AU00720 VE95000720  AU00722 VE95000722  AU00728 VE95000728  AU00733 VE95000733  AU00734 VE95000734  AU00736 VE95000736  AU00744 VE90000744  AU00747 VE95000747  AU00757 VE95000757  AU00761 VE95000761  AU00762 VE95000762  AU00767 VE95000767  AU00768 VE95000768  AU00769 VE90000769  AU00770 VE95000770  AU00773 VE95000773  AU00775 VE95000775  AU00779 VE95000779  AU00780 VE95000780  AU00783 VE95000783  AU00784 VE95000784  AU00789 VE95000789  AU00795 VE90000795  AU00808 VE90000808  AU00817 VE95000817  AU00820 VE95000820  AU00821 VE95000821  AU00836 VE95000836  AU00838 VE90000838  AU00849 VE90000849  AU00852 VE95000852  AU00853 VE90000853  AU00858 VE95000858  AU00862 VE95000862  AU00863 VE95000863  AU00864 VE95000864  AU00865 VE95000865  AU00866 VE95000866  AU00872 VE95000872  AU00873 VE95000873  AU00875 VE95000875  AU00878 VE95000878  AU00884 VE95000884  AU00888 VE95000888  AU00895 VE95000895  AU00913 VE95000913  AU00924 VE95000924  AU00942 VE95R000942  AU00944 VE95000944  AU00950 VE95000950  AU00952 VE95000952  AU00953 VE95000953  AU00959 VE95R000959  AU00965 VE95000965  AU00970 VE95000970  AU00972 VE95000972  AU00977 VE95000977  AU00979 VE95000979  AU00981 VE95000981  AU00982 VE95000982  AU00988 VE95R000988  AU00993 VE95R000993  AU00994 VE95000994  AU00995 VE95000995  AU00996 VE95000996  AU00997 VE95R000997  AU01001 VE95001001  AU01008 VE95001008  AU01010 VE95001010  AU01011 VE95001011  AU01012 VE95001012  AU01016 VE95001016  AU01017 VE95001017  AU01022 VE95001022  AU01024 VE9501024  AU01025 VE95001025  AU01028 VE95001028  AU01030 VE95001030  AU01032 VE95001032  AU01033 VE95001033  AU01041 VE95001041  AU01042 VE95001042  AU01043 VE95001043  AU01046 VE95001046  AU01050 VE95V001050  AU01053 VE95001053  AU01057 VE95001057  AU01058 VE95001058  AU01063 VE95001063  AU01070 VE95001070  AU01071 VE95001071  AU01079 VE95001079  AU01091 VE90001091  AU01094 VE95001094  AU01096 VE95001096  AU01097 VE95001097  AU01100 VE95001100  AU01106 VE95001106  AU01108 VE95001108  AU01111 VE95001111  AU01113 VE95001113  AU01115 VE95001115  AU01116 VE95001116  AU01119 VE90V001119  AU01122 VE95001122  AU01124 VE95001124  AU01129 VE9001129  AU01136 VE95001136  AU01138 VE90001138  AU01140 VE90R001140  AU01141 VE9001141  AU01145 VE90V01145  AU01147 VE90001147  AU01148 VE90R001148  AU01149 VE90001149  AU01158 VE95001158  AU01161 VE95001161  AU01165 VE95001165  AU01173 VE95001173  AU01174 AM155RY647  AU01175 VE95001175  AU01176 VE95001176  AU01177 VE95001177  AU01178 VE95001178  AU01179 VE95R001179  AU01181 VE95001181  AU01184 VE95001184  AU01189 VE95001189  AU01190 VE95001190  AU01191 VE95001191  AU01192 VE95R001192  AU01194 VE95001194  AU01195 VE95001195  AU01197 VE90001197  AU01199 VE95001199  AU01206 VE95001206  AU01207 VE95001207  AU01208 VE95001208  AU01211 VE95001211  AU01214 VE95001214  AU01217 VE95001217  AU01245 VE90V01245  AU01249 VE95001249  AU01256 VE95001256  AU01257 VE95001257  AU50001 VE95050001  AU50002 VE95050002  AU50022 VE95050022  AU50033 VE90050033  AU50037 VE95R050037  AU50041 VE95050041  AU50046 VE95050046  AU50049 VE95050049  AU50052 VE95050052  AU50053 VE95050053  AU50054 VE95050054  AU50056 VE95050056  AU50057 VE95050057  AU50059 VE95050059  AU50062 VE95050062  AU50064 VE95050064  AU50072 VE95050072  AU50074 VE95050074  AU50077 VE95050077  AU50090 VE95050090  AU50097 VE95050097  AU50107 VE95050107  AU50107 VE95050107  AU50108 VE95050108  AU50114 VE95050114  AU50116 VE95050116  AU50117 VE90R00050117  AU50119 VE95050119  AU50126 VE95R050126  AU50128 VE95050128  AU50131 VE90050131  AU50136 VE95R050136  AU50137 VE90050137  AU50138 VE90R050138  AU50139 VE95050139  AU50140 VE95050140  AU50141 VE90050141  AU50143 VE95050143  AU50146 VE95050146  AU50148 VE95050148  AU50151 VE95050151  AU50153 VE95050153  AU50157 VE95050157  AU50162 VE95050162  AU50166 VE95050166  AU50167 VE95050167  AU50168 VE90050168  AU50170 VE95050170  AU50174 VE90050174  AU50177 VE95050177  AU50178 VE95050178  AU50180 VE95050180  AU50181 VE95050181  AU50182 VE95050182  AU50184 VE95050184  AU50188 VE95050188  AU50189 VE95050189  AU50190 VE95050190  AU50191 VE950050191  AU50192 VE95050192  AU50195 VE95050195  AU50198 VE90050198  AU50199 VE9050199  AU50200 VE95050200  AU50201 VE95050201  AU50202 VE90R050202  AU50203 VE95050203  AU50204 VE95050204  AU50205 VE95050205  AU50207 VE95050207  AU50208 VE90050208  AU50209 VE90050209  AU50212 VE95050212  AU50214 VE95050214  AU50216 VE95050216  AU50218 VE95050218  AU50228 VE95050228  AU50248 VE95050248  AU50256 VE95050256  AU50275 VE9050275  AU50292 VE90050292  AU50306 VE95050306  AU50307 VE95050307  AU50308 VE95050308  AU50312 VE95R050312  AU50320 VE95050320  AU50321 VE95R0050321  AU50333 VE95050333  AU50351 VE9050351  AU50358 VE95050358  AU50360 VE9050360  AU50361 VE95050361  AU50362 VE90050362  AU50363 VE95050363  AU50364 VE95050364  AU50365 VE95050365  AU50366 VE95050366  AU50367 VE95050367  AU50371 VE95050371  AU50372 VE95050372  AU50373 VE95050373  AU50374 VE95050374  AU50375 VE95050375  AU50376 VE95050376  AU50379 VE95050379  AU50382 VE95050382  AU50383 VE90050383  AU50384 VE95050384  AU50386 VE95050386  AU50389 VE95050389  AU50421 VE95000421  AU50463 VE95000463  AU50469 VE95000469  AU50471 VE95000471  AU50473 VE95000473  AU50484 VE95000484  AU50370 5973UL0647  AU00436 20208ULT03       AU00698 IE1560US17       AU01052 AE1028US08  AU01077 AE1028US09  AU01226 AE1045US09  AU01080 33043ULT01  AU50111 P06118UL13  AU50155 1573UL0555  AU00270 795020US08       AU00389 502010US31       AU00399 534551US03       AU00496 110010US16       AU00900 533911US03       AU00960 502010US32  AU00966 504020US11  AU00973 150180US01  AU01009 121383US18  AU01036 124057US03  AU01044 224677US01  AU01054 274070US04  AU50068 900010U102       AU50154 900010U100  AU50186 115025US03  AU50345 GON4428693  AU00303 E001US90         AU00771 O033US03         AU00912 O020US09         AU50008 O009US90         AU50016 O009US89         AU50134 O010US19  AU50135 N001US51  AU50197 A204US01  AU50206 A207US07  AU50210 P001US33  AU50211 A207US06  AU50225 L006US03  AU00304 DEN00745         AU00400 MP230732         AU00612 CEG94641         AU00654 DE528490         AU00746 UE277524         AU00750 DEF28622         AU00816 YP263715         AU00954 YP603500  AU01171 MPX43158  AU50341 MP599325  AU00964 ZA2086US01  AU00969 ZA2846US02  AU00983 ZA2812US10  AU00841 SI4067US45       AU00992 SA1012US62  AU01019 SA1012US65  AU01027 SA1012US63  AU01034 SA1012US64  AU50039 SA1052US24  AU01202 RU5264US05  AU01252 DEMO__US75  AU00416 GEDEM2US34       AU00908 RO4547US05  AU50175 RO4766US04  AU00276 PPR35413         AU00850 PER18316         AU00904 PER06648         AU01049 PER06649  AU01068 PES05822  AU01209 PER19023  AU50065 PER03525         AU50171 PPR23107         AU50387 PPR40828  AU00583 600043US06       AU00656 PL1023US25       AU50285 PL4430US20  AU50310 PL4176US01  AU50377 1557UL0337  AU01037 NO4031US23  AU01060 NO4014US59  AU01143 NO4088US63  AU50130 NO4043US23  AU50236 NO4064US21  AU50247 NO4064US22  AU50353 NO4068US92  AU00365 85140386500  AU00642 85142791102  AU01066 85150626441  AU01230 NC1071US01  AU00971 LDD0449  AU01051 STK01402  AU01064 STK01403  AU01243 LDN01427  AU50244 EMM0540  AU50246 EMM0541  AU50435 LDN01428  AU00500 1573UL0306       AU00613 1573UL0325       AU00462 KW1011US34       AU00533 KW1011US35       AU00555 KW1011US36       AU00560 KW1011US37       AU005622 KW1011US38       AU00883 KW1004US36       AU00911 KW1004US37       AU01205 KW1004US36  AU01258 KW1004US38  AU50419 KW1004US37  AU00216 850372173  AU00262 850372176  AU00290 850372224  AU00307 850372246  AU00315 850372225  AU00392 850372307  AU00578 850372380  AU00599 850372454  AU00620 850372413  AU00627 850372504  AU00649 850372476  AU00948 850372640  AU01067 0850372844  AU01090 0850372872  AU01123 0850372868  AU50105 0850372651  AU50271 0850373066  AU00247 AR0017           AU00564 AR0027           AU00606 AR0029           AU00615 AR0031           AU00710 AR0041           AU00752 AR0042           AU00756 AR0045           AU00809 AR0051           AU00899 AR0054           AU00951 AR0061  AU00962 AR0064  AU01031 AR0067  AU01055 AR0072  AU01084 AR0073  AU01099 AR0080  AU01114 AR0077  AU01126 AR0083  AU01150 AR0078  AU01236 AR0090  AU01238 AR0095  AU01240 AR0091  AU01241 AR0094  AU01242 AR0093  AU01247 AR0092  AU50042 AR0062  AU50098 AR0059  AU50103 AR0060  AU50133 AR0063  AU50158 AR0068  AU50163 AR0066  AU50164 ARA065  AU50165 AR0069  AU50183 ARA070  AU50217 AR0076  AU50220 AR0071  AU50227 AR0075  AU50250 AR0074  AU50288 AR0079  AU50299 AR0082  AU50313 AR0081  AU50402 AR0084  AU50404 AR0085  AU50405 AR0089  AU50410 AR0086  AU50413 AR0087  AU50420 AR0088  AU00916 CONSIP4314  AU01039 A5943804  AU01069 CONSIP4308  AU01073 CONSIP4342  AU01078 CONSIP4309  AU01082 CONSIP4310  AU01088 CONSIP4301  AU01089 CONSIP4304  AU01092 CONSIP4340  AU01121 A51205101  AU01135 CONSIP4392  AU01144 CONSIP4334  AU01146 CONSIP4394  AU01151 CONSIP4395  AU01153 CONSIP4345  AU01155 CONSIP4323  AU01157 CONSIP4319  AU01159 CONSIP4360  AU01164 CONSIP4361  AU01166 CONSIP4313  AU01169 CONSIP4317  AU01182 CONSIP4418  AU01183 GON4421348  AU01185 CONSIP4357  AU01186 B5120740  AU01193 CONSIP4316  AU01198 CONSIP4312  AU01204 CONSIP4359  AU01210 CONSIP4389  AU01213 CONSIP4376  AU01218 CONSIP4372  AU01219 CONSIP4374  AU01221 CONSIP4379  AU01222 CONSIP4369  AU50113 A5985911  AU50129 CONSIP4353  AU50222 CONSIP4307  AU50224 CONSIP4335  AU50231 CONSIP4305  AU50234 CONSIP4346  AU50239 CONSIP4318  AU50260 CONSIP4306  AU50261 CONSIP4303  AU50266 CONSIP4347  AU50268 CONSIP4344  AU50272 CONSIP4344  AU50274 CONSIP4336  AU50281 CONSIP4337  AU50290 CONSIP4302  AU50291 CONSIP4391  AU50294 CONSIP4328  AU50295 CONSIP4326  AU50304 CONSIP4330  AU50305 CONSIP4315  AU50309 B5810128  AU50311 CONSIP4329  AU50315 CONSIP4300  AU50318 CONSIP4321  AU50322 CONSIP4348  AU50323 CONSIP4333  AU50324 CONSIP4339  AU50325 CONSIP4325  AU50326 CONSIP4320  AU50327 CONSIP4349  AU50328 CONSIP4332  AU50329 CONSIP4324  AU50330 CONSIP4341  AU50331 CONSIP4331  AU50332 CONSIP4338  AU50334 CONSIP4351  AU50336 CONSIP4352  AU50338 CONSIP4322  AU50344 CONSIP4409  AU50346 CONSIP4350  AU50349 CONSIP4356  AU50352 CONSIP4358  AU50356 CONSIP4405  AU50359 CONSIP4355  AU50380 CONSIP4378  AU50385 A51014135  AU50388 CONSIP4370  AU50391 CONSIP4373  AU50392 CONSIP4371  AU50393 CONSIP4404  AU50394 CONSIP4375  AU50396 CONSIP4387  AU50398 CONSIP4377  AU50399 CONSIP4380  AU50400 CONSIP4399  AU50401 CONSIP4385  AU50406 CONSIP4396  AU50407 CONSIP4381  AU50408 CONSIP4400  AU50409 CONSIP4390  AU50411 CONSIP4367  AU50412 CONSIP4402  AU50414 CONSIP4393  AU50415 CONSIP4427  AU50417 CONSIP4403  AU50418 CONSIP4398  AU50421 CONSIP4364  AU50422 CONSIP4368  AU50424 CONSIP4401  AU50425 CONSIP4362  AU50426 CONSIP4365  AU50427 CONSIP4384  AU50429 CONSIP4397  AU50431 CONSIP4366  AU50432 CONSIP4388  AU50434 CONSIP4383  AU50436 CONSIP4363  AU50438 CONSIP4386  AU01047 GON4397886  AU00874 IE1164US11  AU01062 GON4404707  AU50176 083038825017616  AU50428 083038825042816  AU00855 HUDUS0502  AU50099 HUDUS0091  AU50100 HUDUS0092  AU00683 828380443  AU01117 GON4423007  AU00415 828380478  AU50237 VE9AU50237  AU01018 H2026US17  AU50433 H2063US11  AU00876 00257ULT31  AU00929 00257ULT32  AU00945 00257ULT33  AU01112 00190ULT01  AU01118 00192ULT43  AU01125 00082ULT15  AU01128 IE1504US12  AU01231 00526ULT26  AU50086 00091ULT15  AU50092 00040ULT31  AU50093 00257ULT34  AU50094 00091ULT16  AU50102 00257ULT35  AU50109 00023ULT38  AU50118 00040ULT32  AU50120 10448ULT64  AU50122 00040ULT33  AU50124 10448ULT65  AU50243 00638ULT16  AU50273 00115ULT10  AU50280 00309ULT16  AU50454 00123ULT39  AU50456 00316ULT19  AU00105 104046U012       AU00106 109529U087       AU00205 151308U009       AU00338 149870U013       AU00425 113310U003       AU00485 149657U055       AU00502 120111U028       AU00548 100011U026       AU00623 109818U013       AU00635 112099U015       AU00640 140032U025       AU00641 1685347U33       AU00646 100011U027       AU00651 110499U028       AU00689 122109U092       AU00827 110499U029       AU00892 TP1103U024       AU00978 2678902U05  AU00980 145471U005  AU00984 2678902U06  AU00986 110477U020  AU00987 110128U019  AU00990 109725U043       AU00991 1906476U49  AU00998 117946U008  AU01004 TP1098U003  AU01005 2296936U04  AU01006 2432783U02  AU01007 1607646U13  AU01013 110056U006  AU01015 138907U027  AU01020 121982U014  AU01021 123008U006  AU01045 1874464U05  AU01048 141825U021  AU01093 140559U008  AU01098 109026U008  AU01109 1282968U02  AU01132 110810U021  AU01163 138778U015  AU01216 145746U006  AU50047 802809U004  AU50088 120008U026  AU50145 109853U003  AU50149 802682U005  AU50149 802682U005       AU50152 121982U013  AU50161 121982U012  AU50253 800218U008  AU50263 138955U019  AU50269 141825U020  AU50270 138726U008  AU50283 145032U025  AU50286 1951144U25  AU50289 148270U008  AU50293 148119U006  AU50300 123017U024  AU50302 104077U037  AU50335 145228U016  AU50339 110988U014  AU50340 TP8351U003  AU50343 3686201U03  AU50347 142541U077  AU50348 4122339U03  AU50350 110910U049  AU50354 138560U012  AU00356 USM4143713       AU00458 USM4487804       AU00941 U4459201  AU00956 U235798707  AU00967 U235798708  AU00975 USM1470601  AU01038 USM4496914  AU01056 U280444801  AU01072 US92967801  AU01075 USN4023004  AU01081 U292967802  AU01085 USM4479609  AU01087 USM3168009  AU01101 U101514704  AU01167 US88657320  AU01203 USM2637301  AU01223 US88638609  AU01224 U102977805  AU01227 U464456001  AU01234 U464456002  AU01250 USM4199323  AU50080 USM4166921       AU50132 US26492218  AU50150 U217405807  AU50226 US78041309  AU50230 U126446617  AU50241 USN4516001  AU50259 U178608809  AU50368 U276410806  AU50395 USM3183515  AU50397 USM4046520  AU50442 USM4482805  AU50179 FI1007US22  AU50125 EE4029US32  AU50223 EE4029US33  AU01263 GON4424245  AU01235 VE95AU01235  AU00157 DK1018US37       AU00204 DK1030US57       AU00291 DK1009US29       AU00292 DK1022US55       AU00666 DK1083US69       AU00727 DK1047U148       AU00743 DK1047U149       AU00869 DK1083US67       AU00905 DK1039US32       AU01000 DK1044U153  AU01014 DK1042US41  AU01076 DK1077US37  AU01187 DK1066US71  AU50009 DK1047U152       AU50038 DK1106US21       AU50070 DK1051US28       AU50096 DK1066US67       AU50106 DK1066US68  AU50251 DK1026US84       AU50257 DK1026US85       AU50258 DK1026US86       AU50267 DK1077US39  AU50267 DK1077US39       AU50276 DK5005U154  AU01102 CZ4012US17  AU50219 CZ6028US10  AU00963 HR4100US04  AU50369 HR4252US01  AU00898 VE95AU00898      AU50172 VE95AU50172      AU50439 VIE95AU50439  AU01120 GON4408384  AU00643 416360VE95E  AU00670 416360VE95A  AU00674 519464VE95A  AU00678 416360VE95D  AU00725 709778VE95A  AU00735 416596VE95A  AU00894 416340VE95B  AU00897 514934VE95  AU00920 613737VE95  AU00947 416461VE95A  AU00955 519426VE95  AU01023 905639VE90  AU01026 416461VE95C  AU01035 514934VE95A  AU01040 705646VE95  AU01059 450375VE85  AU01074 905577VE95A  AU01103 709737VE95A  AU01105 905577VE95B  AU01107 450759VE95  AU01110 905521VE95A  AU01127 450371VE95A  AU01131 905577VE95D  AU01133 514495VE95A  AU01168 403270VE90B  AU01188 403943VE90A  AU01239 905895VE95A  AU01244 514934VE95B  AU01248 905895VE95B  AU01253 905895VE95D  AU01259 514934VE95C  AU50006 416360VE95B  AU50011 416360VE95C  AU50021 416360VE95F  AU50028 514412VE95  AU50030 604875VE90B  AU50089 416340VE95A  AU50095 450668VE95  AU50101 613938VE95  AU50144 416461VE95B  AU50144 416461VE95B  AU50187 905577VE95C  AU50265 709778AVE95A  AU50277 450468VE95B  AU50278 905527VE95A  AU50279 418656VE95A  AU50282 514340VE95A  AU50284 514340VE95B  AU50297 450468VE95A  AU50314 403270VE90A  AU50316 514338VE95A  AU50319 403270VE90C  AU50342 819346VE95A  AU50440 905895VE95C  AU50262 BG4004US05  AU50437 6073UL0290  AU01262 VE9AU01262  AU00889 210002US92       AU01154 030037US09  AU01220 030601US03  AU01237 030419US14  AU50104 210013US26  AU50115 549268US01  AU50357 030209US03  AU00296 AH5407US27       AU00686 TP7880US22       AU01130 AH5050US17  AU50112 AH7412US01  AU50337 AH4323US74  AU50459 TP7880US22  AU00136 83134499648  AU00278 83139590369  AU00358 83139203647  AU00536 83141504329  AU00539 83142373002  AU00673 83142913846  AU00890 83147389384  AU00922 83148231417  AU00930 83147509827  AU00934 83147389385  AU00943 83148231419  AU01061 83150626047  AU01142 83150943430  AU01212 83153077112  AU50110 83147639403  AU50173 83150938617  AU50185 83150939044  AU50193 83150938850  AU50194 83150938569  AU50213 83150972921  AU50221 83150626182  AU50221 83150626182  AU50229 83150626154  AU50229 83150626154  AU50232 83151090524  AU50238 83150972920  AU50242 83151032129  AU50249 83150981954  AU50254 83151090555  AU50287 83150943428  AU50296 83150943429  AU50355 83152817052  AU50378 83153077317  AU50452 83154259953  AU50452 83154259953  AU01003 VE95AU01003  AU50160 VE95AU50160",
      "more_code_info": ""
    }
  ]
}