  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Plymouth",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84109",
      "recalling_firm": "Medivators, Inc.",
      "address_1": "14605 28th Ave N",
      "address_2": "N/A",
      "postal_code": "55447-4822",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of Australia, England.",
      "recall_number": "Z-1778-2020",
      "product_description": "Medivators  Endo carry-on Procedure Kit REF CPK-100222-0002 - Product Usage: is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.",
      "product_quantity": "53,790 units total",
      "reason_for_recall": "Medivators is recalling Endo SmartCap Tubing due to a manufacturing anomaly that may present during procedures as inadequate insufflation, presence of air bubbles inside the endo smartcap tubing, insufficient water to rinse the endoscope lens, or water leaking from the air/water valve.",
      "recall_initiation_date": "20191009",
      "center_classification_date": "20200423",
      "termination_date": "20201211",
      "report_date": "20200429",
      "code_info": "Lot # 1000028736  GTIN:  00677964086045   Use-by date:  210120"
    }
  ]
}