  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82975",
      "recalling_firm": "Maquet Cardiovascular Us Sales, Llc",
      "address_1": "45 Barbour Pond Dr",
      "address_2": "N/A",
      "postal_code": "07470-2094",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide GA, IL, IN, MI, NC, OR, PA, VA    Foreign:   Australia, Canada, China, France, Germany, India, Japan, Korea, Netherlands, Saudi Arabia, South Africa, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, Ukraine",
      "recall_number": "Z-1777-2019",
      "product_description": "Getinge 86-Series-Medical Washer-Disinfector   Part Number: S-86662003     Product Usage:  Usage:  The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.",
      "product_quantity": "4 units",
      "reason_for_recall": "Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.",
      "recall_initiation_date": "20190520",
      "center_classification_date": "20190613",
      "termination_date": "20231215",
      "report_date": "20190619",
      "code_info": "Serail Numbers:  WAA088731, WAA089124 , WAA089128, WAA089586"
    }
  ]
}