      "status": "Terminated",
      "city": "Billerica",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82722",
      "recalling_firm": "Conformis, Inc.",
      "address_1": "600 Technology Park Dr",
      "address_2": "N/A",
      "postal_code": "01821-4154",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-1775-2019",
      "product_description": "Conformis iTotal Hip Replacement System:  HBS-033-00xx-010101, (HIP-PATIENT SPECIFIC STEM LEFT)",
      "product_quantity": "128",
      "reason_for_recall": "Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants.",
      "recall_initiation_date": "20190329",
      "center_classification_date": "20190613",
      "termination_date": "20200428",
      "report_date": "20190619",
      "code_info": "HBS-033-0010-010101:  Serial Numbers:423398, 424244, 424392, 425455, 425607, 425755, 426309, 427186, 428901, 430404,430766, 431394,431844, 434871, 436520, 438363, 438925, 439064, 439561, 440369, 441662, 441734;    HBS-033-0011-010101:  Serial Numbers:422796, 424450, 424510, 426034, 426118, 430150, 430534, 431142, 431502, 431581, 432030, 433555, 434136, 434439, 435282, 436108, 436212, 437452, 437657, 437745, 437968, 440603, 442816;    HBS-033-0012-010101:  Serial Numbers: 425346, 425608, 425866, 427702, 427754, 429459, 429844, 430053, 430087, 432011, 432363, 436792, 437653, 438516, 439063, 440253;    HBS-033-0013-010101:  Serial Numbers: 422797, 424647, 425456, 426297, 427360, 427898, 428372, 428957, 429014, 429921, 429937, 430338, 430613, 430962, 433445, 434149, 434451, 434635, 435118, 436125, 436230, 436333, 437073, 438024, 438112, 439641, 439929, 441028, 441030;    HBS-033-0014-010101:  Serial Numbers: 423231, 425276, 426069, 426382, 427480, 427752, 427883, 428820, 428900, 431315, 431333, 431977, 432543, 432953, 434612, 435649, 435883, 435910, 436122, 437322, 437863, 438631, 440445;    HBS-033-0015-010101:  Serial Numbers: 429552, 432936, 433085, 437373, 437681, 437925, 439272  HBS-033-0016-010101 425418, 428545, 431199, 435657, 436790, 437425, 437656, 442745;    HBS-033-0016-010101:  Serial Numbers:425418, 428545, 431199, 435657, 436790, 437425, 437656, 442745"
    }
  ]
}