{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76862",
      "recalling_firm": "Draegar Medical Systems, Inc.",
      "address_1": "6 Tech Dr",
      "address_2": "N/A",
      "postal_code": "01810-2434",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of: AL, CA, GA, IL, LA, MA NY, OH, PA, TN, TX and VA and the countries of:  AR  AT  BE  BG  BN  BO  BY  CA  CH  CL  CO  CZ  DE  EG  ES  FI  FR  GB  GR  HR  HU  ID  IE  IL  IN  IT  LT  MX  MY  NL  NO  NP  PA  PE  PK  PL  PT  RO  RS  RU  SA  SE  SK  TH  TR  TW  UG  UZ  VN  ZA",
      "recall_number": "Z-1774-2017",
      "product_description": "Kappa, Catalog Number: MS18853 in combination with Scio, Scio Four, Scio Four Oxi plus, Scio Four Oxi, Scio Four plus.  Catalog Number(s): MS18597/MS18596/MS18853.    Product Usage:  The devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient\u0019s medical condition. The Infinity Delta Series (Delta/Delta XL/Kappa) monitors are intended to be used on adult, pediatric, and neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended for use in the adult and pediatric populations only; and tcpO2, which for the neonatal population, is to only be used when the patient is not under gas anesthesia. For combination with Scio gas module: Scio gas module samples breathing gases from adults and pediatrics. The gas module continuously measure the content of CO2, N2O, O2 and one of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to the In-finity monitors.",
      "product_quantity": "2,156 (total for all devices)",
      "reason_for_recall": "It was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit.",
      "recall_initiation_date": "20170328",
      "center_classification_date": "20170410",
      "termination_date": "20180917",
      "report_date": "20170419",
      "code_info": "VF10.0 software",
      "more_code_info": ""
    }
  ]
}