      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68331",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E Milham Ave",
      "address_2": "N/A",
      "postal_code": "49002-9704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide Distribution - US (nationwide)",
      "recall_number": "Z-1774-2014",
      "product_description": "Stryker 2.5mm x 3.6mm Neuro Drill Soft Touch  Sterile    Product Usage:  The devices are intended to be used to cut bone and bone related tissue during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial surgeries. The main procedures/conditions in which these products are expected to be used are Mastoidectomy/Cochlear implant, Craniotomy, Vertebral Fusion and Fixation and Craniosynostosis.",
      "product_quantity": "41",
      "reason_for_recall": "There may be missing notches, partial notches or no notches at all in some Cutting Accessories that are used with the Elite\" Attachments.",
      "recall_initiation_date": "20140522",
      "center_classification_date": "20140610",
      "termination_date": "20150506",
      "report_date": "20140618",
      "code_info": "Part Number 5820-107-25, Lot Number 13346017"
    }
  ]
}