      "status": "Terminated",
      "city": "Santa Clara",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65688",
      "recalling_firm": "Topcon Medical Laser Systems, Inc",
      "address_1": "3130 Coronado Dr",
      "address_2": "N/A",
      "postal_code": "95054-3205",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA and the countries of China, Hong Kong, Brazil, France, Japan, Netherlands, Pakistan, Malaysia, United Kingdom, South Korea, Thailand, Dubai, and South Africa.",
      "recall_number": "Z-1774-2013",
      "product_description": "PASCAL and PASCAL Streamline (532 nm.577 nm);    Treats the patient's eye.",
      "product_quantity": "7 US, 33 foreign",
      "reason_for_recall": "When using the \"Favorites\" function for saved settings, there is potential for a mismatch between displayed power and the actual power output. This may result in over or under treatment.",
      "recall_initiation_date": "20120327",
      "center_classification_date": "20130719",
      "termination_date": "20130719",
      "report_date": "20130731",
      "code_info": "PASCAL Standard 2R/G System v2.x.x and PASCAL Streamline 2R/G Software v2.x.x  Serial numbers:  United States  33057910  33060410  33063610  66078611  66084911  66085111  66085511  Out of United States  OMC-179  11138509  11141209  11147209  11147809  11148910  11149010  11149210  11149710 .  11149810  11150110  11150210  11150410  11150610  11150810  11150910  11151010  1 151110  11151211  33054310  33059310  33059510  33062810\"  66075211'  66076711  66077111  66078011  66079911  6608391'1  66084611  66084811  66085011  66085211",
      "more_code_info": ""
    }
  ]
}