  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71134",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) including DC & Puerto Rico and the following countries: Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Cambodia, Cameroon, Cayman Islands, China, Colombia, Democratic Congo, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guatemala, Hong Kong, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Kyrgyzstan, LAO People's Dem. Rep., Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malawi, Malaysia, Malta, Mauritius, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Philippines, Poland, Portugal, R¿union, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia & Montenegro, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam, Yemen, and Zambia.",
      "recall_number": "Z-1771-2015",
      "product_description": "Ingenuity Core (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.",
      "product_quantity": "215 Units",
      "reason_for_recall": "The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.",
      "recall_initiation_date": "20150306",
      "center_classification_date": "20150611",
      "termination_date": "20170313",
      "report_date": "20150617",
      "code_info": "Model #: 728321; Serial #s: 52000, 52001, 52002, 52004, 52006, 52007, 52008, 52009, 52010, 52013, 52015, 52016, 52017, 52018, 52020, 52021, 52022, 52024, 52026, 52033, 52038, 52039, 52046, 52051, 52052, 52057, 310001, 310002, 310003, 310004, 310005, 310006, 310007, 310009, 310010, 310011, 310012, 310013, 310014, 310015, 310016, 310017, 310018, 310019, 310020, 310021, 310022, 310024, 310025, 310026, 310027, 310028, 310029, 310030, 310031, 310032, 310033, 310034, 310035, 310036, 310037, 310038, 310040, 310041, 310043, 310044, 310045, 310046, 310047, 310048, 310049, 310050, 310051, 310052, 310053, 310054, 310055, 310056, 310057, 310059, 310060, 310061, 310062, 310063, 310064, 310065, 310066, 310068, 310069, 310070, 310071, 310072, 310073, 310074, 310075, 310076, 310077, 310078, 310079, 310080, 310081, 310082, 310083, 310084, 310085, 310086, 310087, 310088, 310089, 310090, 310091, 310092, 310093, 310094, 310095, 310096, 310097, 310099, 310100, 310101, 310102, 310103, 310104, 310105, 310106, 310107, 310108, 310109, 310110, 310111, 310112, 310113, 310114, 310115, 310116, 310117, 310118, 310119, 310120, 310121, 310123, 310124, 310125, 310126, 310128, 310129, 310131, 310132, 310133, 310134, 310135, 310136, 310137, 310138, 310139, 310140, 310141, 310143, 310144, 310145, 310148, 310149, 310151, 310152, 310153, 310154, 310155, 310157, 310158, 310160, 310161, 310162, 310163, 310164, 310165, 310170, 310171, 333001, 333002, 333003, 333005, 333006, 333007, 333008, 333009, 333010, 333011, 333012, 333013, 333014, 333015, 333016, 333017, 333022, 333024, 333025, 333026, 333027, 333030, 333031, 333035, 333037, 333038, 333042, 333044, 333045, 333046, 333051, 333052, 333053, 333055, 333057, 333060, 333061, 333066."
    }
  ]
}