{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hertford",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65630",
      "recalling_firm": "Del Mar Reynolds Medical, Ltd.",
      "address_1": "1-2 Harforde Ct., Foxholes",
      "address_2": "Business Park",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including South Carolina and internationally to Canada and Great Britian.",
      "recall_number": "Z-1771-2013",
      "product_description": "Spacelabs ARKON Anesthesia Workstation.     Spacelabs ARKON Anesthesia Workstation may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer.",
      "product_quantity": "9 units (4 in the US and 5 outside US)",
      "reason_for_recall": "Spacelabs ARKON Anesthesia System is recalled due to software defect.  The System Status Computer may incorrectly determine that a communication error has occurred. If this situation occurs, a buzzer will activate and a yellow triangle will be displayed on the System Status Computer screen that alerts the user of this error.",
      "recall_initiation_date": "20130607",
      "center_classification_date": "20130719",
      "termination_date": "20130829",
      "report_date": "20130731",
      "code_info": "ARKN-000001, ARKN-000005, ARKN-000010, ARKN-000015, ARKN-000016, ARKN-000017, ARKN-000018, ARKN-000020, ARKN-000021.",
      "more_code_info": ""
    }
  ]
}