  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90757",
      "recalling_firm": "Fresenius Medical Care Holdings, Inc.",
      "address_1": "920 Winter St Bld 950",
      "address_2": "N/A",
      "postal_code": "02451-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.",
      "recall_number": "Z-1770-2022",
      "product_description": "AquaC UNO H [120V, 60Hz, 1700VA, US-Version- a reverse osmosis unit intended for use with hemodialysis systems.   Fresenius Part Number: 24-0970-1  Vivonic Part Number: G02000966-US",
      "product_quantity": "6 units",
      "reason_for_recall": "Internal ethemet cables are missing one or both ferrite cores therefore  the electromagnetic compatibility of the device is not ensured",
      "recall_initiation_date": "20220621",
      "center_classification_date": "20220919",
      "report_date": "20220928",
      "code_info": "UDI: (01)4251285200234(11)200422(21)0CHS3987 (01)4251285200234(11)200526(21)0CHS4108 (01)4251285200234(11)210623(21)1CHS5372 (01)4251285200234(11)210707(21)1CHS5405 (01)4251285200234(11)190305(21)9CHS2881 (01)4251285200234(11)190529(21)9CHS3090  Serial Numbers: 0CHS3987 0CHS4108 1CHS5372 1CHS5405 9CHS2881 9CHS3090"
    }
  ]
}