  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Completed",
      "city": "Tempe",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79841",
      "recalling_firm": "Bard Peripheral Vascular Inc",
      "address_1": "1625 W 3rd St Bldg 1",
      "address_2": "N/A",
      "postal_code": "85281-2438",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-1768-2018",
      "product_description": "POWERPORT ClearVUE Slim with Smooth Septum, 6F ChronoFlex Silk, Polyurethane Catheter, REF 5676300",
      "product_quantity": "3196 units",
      "reason_for_recall": "The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions.",
      "recall_initiation_date": "20180208",
      "center_classification_date": "20180509",
      "report_date": "20180516",
      "code_info": "Lot Number Unique Device Identifier (UDI)  REAN0055 (01)00801741110771(17)170628(10)REAN0055  REAN0202 (01)00801741110771(17)170528(10)REAN0202  REAP0771 (01)00801741110771(17)170828(10)REAP0771  REAS1708 (01)00801741110771(17)171028(10)REAS1708  REAT0147 (01)00801741110771(17)171128(10)REAT0147  REAT1852 (01)00801741110771(17)171228(10)REAT1852  REAU1811 (01)00801741110771(17)171228(10)REAU1811  REAU2420 (01)00801741110771(17)171228(10)REAU2420  REAV2465 (01)00801741110771(17)180228(10)REAV2465  REAW1422 (01)00801741110771(17)180331(10)REAW1422  REAW1831 (01)00801741110771(17)180331(10)REAW1831  REAX1132 (01)00801741110771(17)180331(10)REAX1132  REAY1623 (01)00801741110771(17)180331(10)REAY1623  REAY2597 (01)00801741110771(17)180331(10)REAY2597  REBP0090 (01)00801741110771(17)180331(10)REBP0090  REBS1269 (01)00801741110771(17)181031(10)REBS1269  REBR1002 (01)00801741110771(17)181031(10)REBR1002  REBS1439 (01)00801741110771(17)181031(10)REBS1439  REBT1052 (01)00801741110771(17)181231(10)REBT1052  REBT2348 (01)00801741110771(17)181231(10)REBT2348  REBV0714 (01)00801741110771(17)190131(10)REBV0714  REBV1758 (01)00801741110771(17)190131(10)REBV1758  REBW1288 (01)00801741110771(17)190331(10)REBW1288  REBX1628 (01)00801741110771(17)190331(10)REBX1628  REBX2052 (01)00801741110771(17)190430(10)REBX2052"
    }
  ]
}