  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Rd",
      "reason_for_recall": "The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.",
      "address_2": "",
      "product_quantity": "81,626",
      "code_info": "Serial Numbers:  US94348760  US53404426  US11479851  US53404552  US53404773  US65112716  US61307506  US63409602  US65113980  US11482453  US44801180  US53404420  US00552837  US00554294  US71218273  US53402710  US53402748  US53402489  US44800249  US73729592  US53404994  US44800361  US53406037  US53404431  US00554876  US00554879  US73728988  US52402192  US65112718  US65113706  US65112932  US73731015  US71216309  US63411735  US44801179  US53403403  US71217875  US03862052  US61306876  US53405461  US02656890  US53403885  US50901456  US53403109  US53405111  US65112438  US44800332  US00550542  US00550535  US71217878  US50901335  US71217882  US83233987  US53403371  US00551365  US83231550  US53405148  US00550937  US94348947  US11470201  US11467421  US90741654  US71217797  US63409492  US03863289  US11468736  US90740188  US03864848  US11480781  US11479368  US90739769  US90439076  US11477413  US11478904  US11468946  US71217881  US61307709  US11480777  US90739914  US11476769  US03863203  US11483122  US83231713  US11469041  US02656232  US03860015  US03865996  US73731254  US03863933  US11472503  US11166777  US94350014  US71216576  US73729003  US03859909  US03863291  US53405452  US00553108  US11468063  US91944592  US84137301  US03859856  US11471545  US91944588  US53402706  US53404811  US03863598  US11479477  US03865320  US71214310  US11480447  US84137955  US11485811  US03858616  US65112287  US91943592  US90740923  US84138344  US03863893  US71219972  US11482060  US11470580  US03862139  US03863412  US03864916  US11490636  US65112308  US63409285  US83231925  US63411840  US11473308  US61307371  US94347811  US71217367  US83233757  US11490472  US11478644  US44800327  US11472476  US11477950  US53406036  US53406000  US71217877  US63410601  US11470815  US94347131  US11469286  US94346807  US44800314  US44800308  US84135185  US94347833  US91944790  US73723262  US91944590  US11486554  US11481315  US03862872  US03860965  US84135017  US61307275  US90740547  US90740062  US11473124  US65112381  US00551266  US11485535  US71216220  US03865100  US03864729  US71214936  US03863998  US11489560  US11167106  US73728042  US11478910  US11479934  US11486663  US73723938  US02657041  US11471264  US11472800  US03357492  US90740273  US11472482  US11476030  US11166347  US11467659  US02656225  US91943134  US11472155  US11480160  US11486623  US00554484  US03357542  US03863752  US03863781  US11481066  US11486214  US11480516  US11480474  US11475074  US84138434  US11490133  US11481218  US73728650  US71218078  US11479054  US11481068  US84136805  US02656243  US84137560  US90739196  US03865718  US73728499  US73728495  US83231924  US00551964  US73727726  US03859609  US03859605  US03858607  US94347827  US03864925  US00550733  US00552123  US11476997  US11484514  US00552055  US00553504  US84136700  US11470581  US00556087  US11480901  US11481401  US11484023  US73722029  US90740864  US71221133  US00555732  US11482295  US53403117  US11475543  US00554690  US03864284  US11471762  US00554728  US73721317  US11482552  US00554733  US11166682  US00554527  US91942702  US03861295  US61307814  US61307808  US00550830  US03861983  US11476886  US11471550  US94348463  US11487298  US73726435  US03859506  US61306450  US53404761  US00553489  US50901430  US71216324  US94349457  US11485710  US71217466  US94347824  US11484434  US03357824  US11484745  US03861753  US11471764  US03865154  US61306879  US11470411  US52401928  US91944517  US84137735  US00550480  US03863174  US00552755  US03859941  US91943636  US90740178  US65113344  US11484858  US53404352  US11482298  US03859186  US11483180  US00553470  US83232484  US53404417  US53404868  US73726802  US11478905  US11471513  US11485068  US71218170  US11480460  US11473177  US71218283  US11480757  US91946382  US00553870  US00553857  US73726158  US90741159  US44800280  US83234532  US83234533  US84136212  US11469151  US03357193  US11167245  US11167240  US03860552  US73721324  US44800189  US94347999  US00550749  US53403504  US53403718  US63411200  US03862946  US91946416  US71220387  US73725323  US71220043  US71217666  US03865194  US73726777  US11468296  US11490653  US65113041  US53404012  US03863389  US00551960  US02656227  US00552029  US73724725  US11490426  US11166676  US11484544  US11468497  US73726397  US00551259  US00553698  US03859611  US00550867  US84136777  US11478913  US11485414  US71216522  US53406107  US61307394  US91942294  US03859535  US00552494  US71219905  US71217087  US71219682  US90741252  US11475624  US03858990  US11469147  US00552128  US11471780  US11475772  US11167258  US65113481  US84137737  US90741163  US94349464  US03863924  US00551475  US90740599  US11474258  US11481133  US11468475  US03357228  US03357196  US00552386  US63411986  US03864355  US73722958  US73728858  US03357183  US00553837  US91946355  US11483476  US91942315  US63410889  US65112388  US71217161  US00553697  US44801177  US83234253  US63409493  US94347556  US44800245  US91942369  US90741162  US11488731  US03357229  US00555989  US73727267  US00553503  US00552019  US00552518  US71215059  US73727951  US44800343  US11478903  US00550471  US02657113  US61307979  US61307980  US84138224  US73721875  US90740050  US03863935  US73726524  US73726521  US11471285  US63408887  US63408905  US73725939  US61306606  US03357980  US71216816  US00554486  US00551185  US73727026  US61308267  US11472596  US84137622  US03861198  US83234250  US44800274  US11471827  US11483105  US50901678  US50901436  US50901439  US11469152  US00554332  US00554300  US73727835  US71220252  US90739889  US90740847  US90740677  US11469497  US00551790  US94349448  US90740913  US63411732  US73730076  US53405729  US91945323  US11489979  US71215006  US84136553  US84136546  US91944139  US03866092  US94347534  US63409307  US00550730  US00551890  US91944481  US63411079  US11480684  US03860205  US90741161  US83232837  US83234002  US00550514  US11483321  US11477441  US00552621  US11478388  US94349622  US03865065  US84137290  US84137307  US11491040  US11471136  US11471174  US73731250  US11487366  US11473119  US03357720  US03358403  US03358410  US53404865  US53404589  US11482808  US91943765  US11483191  US11480365  US83234293  US11484659  US00553069  US03859377  US71214874  US03865356  US03865404  US03865388  US61307502  US03358539  US11470587  US03864412  US65112443  US63410132  US71219147  US03357907  US03357866  US03357860  US03357889  US03862356  US03862357  US84135364  US03358225  US65114003  US11476797  US84137761  US00550156  US63411208  US63411236  US11482042  US03358320  US61307799  US03862814  US73726915  US11483244  US53403285  US11467859  US00553320  US94349216  US71218196  US11467427  US94350127  US11470748  US03358503  US03357500  US11486221  US11471683  US11477423  US03860793  US00552616  US71217881  US84135857  US90739746  US63409486  US11479275  US11479320  US90739804  US63411147  US03357217  US11478396  US03859677  US83234363  US11481650  US91942482  US11479395  US11166417  US11468540  US73722566  US03858611  US90740881  US00553269  US11470224  US71217312  US73731425  US11485230  US65114104  US71219459  US71218101  US03863560  US84137976  US61307670  US73726305  US11470208  US94346680  US11478215  US84137118  US11477431  US71218849  US00552016  US11166358  US91943062  US00554288  US94348901  US71220368  US91946266  US91946264  US91945774  US71217669  US11480786  US73728691  US90740390  US11471093  US11485292  US11472320  US73724591  US73724581  US44800307  US11485324  US11473649  US00555726  US63410624  US00553192  US53402497  US61306702  US50901500  US53404812  US61307712  US61306753  US90741349  US71214398  US73725197  US73725220  US71215965  US11473031  US11470711  US03861346  US71221078  US90741591  US02656860  US71218459  US03357978  US11467767  US83231712  US63409142  US00550397  US11485682  US00551011  US00554176  US63410730  US00555795  US00554554  US00554565  US03860327  US53405700  US91943800  US65112521  US11479867  US50901206  US53404169  US11484646  US71220260  US83231934  US11471971  US11483248  US00556120  US11473765  US00551996  US11479435  US71216929  US03864292  US71214543  US53404111  US11473595  US11481941  US50901784  US03862039  US11483341  US02656795  US11470583  US91943457  US00551370  US11467858  US91944432  US11483069  US61306282  US03859597  US00553469  US00554265  US11482158  US73724790  US03866142  US73723734  US11473749  US65112998  US11482146  US11470814  US03861534  US11474584  US73724740  US00550779  US84137738  US53403435  US90741175  US03860310  US73729459  US73721840  US11467869  US84136164  US61306336  US11481864  US11167211  US63410757  US84136472  US00556065  US03862936  US11480992  US90739968  US73724098  US11474684  US00551225  US11478516  US00553403  US11471690  US03860880  US00553149  US00555397  US63410451  US00555399  US11472943  US91945068  US65112895  US71216312  US00551303  US90741593  US11479052  US44800313  US90740133  US11485601  US00550677  US11470009  US11475990  US11167120  US00551292  US00552660  US61307981  US73730156  US71214603  US00553847  US03860973  US03860229  US84135128  US11478911  US11478914  US11470419  US11489384  US71218677  US11467649  US11483141  US53404326  US00550546  US11475612  US00552121  US94347501  US03864976  US71216899  US73724738  US03859801  US71218482  US73724543  US91945914  US11487304  US44800451  US11473380  US11475202  US94349293  US84137555  US61307029  US11471100  US00550562  US91942821  US00555436  US83232017  US84137749  US63410568  US71220913  US91944832  US71220943  US84137965  US11475939  US71220400  US11469091  US73724577  US84137916  US73728531  US65113289  US91944852  US63411915  US73731161  US94349458  US63410457  US83232482  US73721300  US11481335  US11484509  US03357984  US63412195  US02656926  US91943142  US00551948  US73727318  US63411721  US03865041  US03859668  US03861986  US00552467  US03859711  US90741126  US53402484  US11483980  US00552078  US53405692  US84136117  US11480244  US71215694  US00550350  US71215686  US11489633  US03861418  US84137987  US73730971  US11472880  US03863376  US61308294  US73724939  US90739986  US03864281  US83232348  US83232347  US83233469  US11476777  US11489755  US03862625  US11470458  US73721979  US11467637  US11468060  US65113555  US73723873  US11467516  US11486476  US83232421  US03866323  US11473180  US11470585  US00554052  US53404282  US11478775  US11473876  US11476775  US11483521  US84137503  US84137661  US11473174  US84138455  US71216714  US03859558  US00551411  US03862612  US02656222  US83232710  US02656226  US84138471  US03861875  US11475784  US03862611  US11469150  US00551208  US11483245  US73721462  US02656900  US03858619  US11467583  US73726120  US71221114  US11481684  US11482446  US00555723  US11482159  US11474211  US00555778  US50901681  US50901688  US03864700  US00553611  US65112310  US11167098  US00552237  US02656898  US73724361  US11468579  US11489689  US03863591  US90741168  US11490336  US65112284  US63411714  US11483240  US00551943  US11482908  US11478199  US11477743  US11477737  US02656805  US02656816  US73728658  US11490150  US11470199  US11467852  US71219040  US03861195  US63411047  US91946398  US73727804  US11483558  US03862933  US11475786  US03862700  US52402162  US11486786  US91944798  US91944818  US03864619  US03862672  US63408669  US11476927  US03863056  US11485152  US11472631  US91942564  US73726200  US73726698  US52402202  US52402201  US00555482  US90740911  US63411990  US44800326  US50901643  US73726139  US90740192  US03865990  US11485772  US90741346  US00556155  US53403901  US11483706  US11471090  US00551845  US11481329  US11468062  US71215831  US11474987  US11474997  US11471417  US73728809  US90739897  US11480771  US11489763  US00551207  US11478818  US11478813  US71219067  US11470768  US84138669  US11483811  US84138027  US11482619  US00552867  US03863020  US90740047  US00550784  US11485256  US90739973  US11475780  US73727607  US00553201  US00551949  US03858885  US61308359  US03357799  US03866292  US00551197  US03358502  US11471666  US71214963  US00553571  US73725296  US11473171  US11476724  US11468051  US11476779  US53405600  US71218214  US11479757  US73722275  US11483744  US90741187  US73721932  US03862754  US83231634  US91942781  US61306405  US73723609  US11488632  US73722830  US65112557  US53404819  US73726585  US11471995  US73724788  US65113648  US73724525  US00552822  US44800315  US11478207  US94350113  US73728844  US11167101  US11485118  US11486468  US84138061  US11482489  US65112384  US11474884  US71215000  US03862337  US11467983  US53402676  US11471667  US03863586  US73724867  US83234586  US11480452  US71217817  US11469324  US91942521  US11472011  US11471951  US11483232  US71216381  US00551164  US91942915  US00551946  US90741312  US73726977  US73726955  US03357935  US03864926  US61307247  US03357463  US90740480  US71216032  US90740942  US00551897  US61306136  US83232578  US11468898  US84137102  US73721746  US84138445  US44800440  US44800480  US71214410  US00551921  US11167104  US00555763  US11481497  US52401989  US90740921  US11469560  US84138393  US11473994  US71218279  US53402814  US73722780  US03865664  US02656969  US03863617  US11476762  US63411613  US44800318  US71216412  US11490941  US03859224  US11480763  US03861959  US11467651  US73731512  US84138277  US71217453  US90741347  US71217228  US63409183  US71218853  US71218852  US65113979  US84135200  US91946494  US11468466  US91946493  US84134695  US84134687  US11488627  US94348345  US00555161  US00555166  US91945702  US91945670  US02657004  US71217212  US71214750  US84134872  US73724593  US11490910  US11490971  US11484875  US90739079  US03865500  US91942914  US11480339  US00554073  US03860355  US11479850  US03862764  US11476421  US91942287  US11476874  US03864277  US00551950  US03863385  US02656435  US11471402  US11485979  US11478667  US11473150  US11479055  US84136711  US91945059  US73726569  US00555811  US03357189  US00553821  US00553832  US00550722  US61306595  US65112583  US03859622  US11166604  US00555176  US03862614  US61306560  US02656928  US91944734  US11486229  US03858640  US73724828  US03864958  US00551938  US03357826  US50901256  US91942704  US00552539  US00552540  US11467428  US91944959  US71218129  US11483995  US73725894  US73726148  US71216492  US03861723  US53405959  US63410252  US11468378  US11470588  US00552156  US03863926  US03859533  US84137080  US71217953  US11471092  US73726604  US11480163  US11483555  US90740186  US73723946  US11468917  US71217541  US11485339  US83232411  US63412193  US11485165  US84137966  US44800401  US00552092  US03358029  US11483080  US84135561  US61306141  US11471406  US11471263  US02656477  US02656347  US02656361  US71215696  US71221027  US71219403  US71219343  US11467861  US11166683  US90741579  US83234544  US00554635  US83232235  US11483825  US11481296  US63410781  US52402161  US03862006  US03859956  US03859669  US11475534  US11472699  US73726014  US73726055  US71216021  US61307869  US00553915  US61307882  US11478175  US71215978  US65112344  US11473324  US11479570  US90740738  US11490992  US00550529  US71218278  US83234506  US65113836  US00551966  US11467706  US11478798  US00555195  US00553844  US73728790  US94349155  US11490436  US71218375  US03861463  US73726941  US11488772  US65113251  US53405924  US84138584  US11478921  US73725651  US11478158  US73723399  US94346775  US83233650  US00550376  US00550380  US11480318  US00550455  US83234023  US63410654  US00551876  US03860055  US11488628  US11474414  US03865178  US00554233  US91946470  US11478397  US61306404  US84135137  US03860367  US00551941  US00555974  US03861467  US11470574  US63409761  US90739835  US00552137  US65113232  US73729007  US73726322  US11475787  US91943953  US53403841  US73727797  US84137322  US11482218  US11478799  US90741172  US03357983  US11478909  US00553468  US11167102  US44800112  US90741167  US73726021  US00555725  US83233960  US11481683  US94349588  US03860106  US91943707  US71217876  US00556136  US11475938  US11480670  US83233133  US00553602  US73725223  US11486952  US71215454  US94348786  US11482445  US90739226  US00552862  US11468587  US00551601  US71219102  US73724270  US11481484  US11478804  US11473175  US11484581  US11477015  US90739656  US00552034  US84137936  US00550351  US03358249  US02656788  US03862849  US71218602  US00554472  US11471540  US03862862  US90740943  US11467862  US83232103  US90438813  US71217492  US00550745  US00551906  US00553171  US73729204  US11491039  US00554823  US11480764  US03358473  US11480373  US44800304  US03862332  US84138568  US91942986  US11481759  US90740377  US71214947  US11468571  US44800113  US53403945  US63409490  US91944867  US00555290  US11482392  US11468748  US00552056  US94347211  US71214941  US71216421  US11478768  US71216105  US65113168  US53405824  US00552017  US71220429  US03859411  US71220431  US53405864  US03863957  US73728869  US03358324  US71217722  US91942474  US03865977  US65112379  US11485207  US11474424  US91945582  US63411717  US00553391  US53405443  US91942477  US03858950  US11481832  US73721995  US11480982  US11472182  US73729465  US11471210  US11481003  US00553891  US71216742  US91942921  US11482733  US63408819  US11475776  US11481888  US73722155  US11473815  US90741477  US11485975  US00550180  US00553130  US00553642  US00553599  US11470780  US11483730  US11486803  US84137756  US90438913  US03858609  US53404580  US00552231  US11483561  US90740915  US73727665  US11471588  US73729670  US90741173  US73728808  US03358299  US03358284  US11474246  US71220894  US03863503  US63409488  US44800247  US44800238  US44800487  US52402214  US50901428  US11490641  US03858621  US63408736  US44800109  US00555507  US00555639  US11474893  US73724881  US91942984  US11467979  US03865821  US84136253  US11487464  US63410251  US65113302  US11475347  US03865087  US11476021  US73724443  US00552429  US00552435  US00555460  US00551914  US00555462  US03864283  US53404814  US94349159  US50901339  US02656546  US11470131  US00551901  US03358210  US73723737  US11166455  US71215044  US11479729  US11483835  US63409230  US53405119  US84136927  US50901235  US73728524  US63412011  US91946151  US73728415  US11469325  US11472487  US83234496  US53404791  US91942848  US61307360  US11472339  US11470059  US90740647  US65112933  US11472811  US11468986  US11469296  US65113293  US11472077  US52402171  US03864687  US11166796  US11167146  US03865165  US03863529  US61306874  US84137284  US73723933  US11470578  US52401858  US63409497  US73730070  US94349695  US50901723  US84138552  US03357216  US11469148  US73728112  US84135942  US00552865  US73725396  US11489501  US03861158  US11471543  US84135538  US03866091  US03862028  US71217165  US03862015  US73721242  US11482236  US83233020  US52401929  US03859277  US11482107  US03358520  US63409677  US03859673  US03866013  US90740883  US03865384  US73722315  US90739884  US02656233  US84138141  US11479144  US63411487  US11470426  US03860801  US03858691  US71219153  US11484523  US71220060  US61308087  US11488932  US91942453  US73728373  US11473610  US73728739  US11483038  US00551969  US11474944  US71218637  US61307249  US90741125  US00553259  US00553298  US11483155  US11482776  US73725594  US03858623  US03859152  US11479705  US84135898  US84135901  US02656497  US73728126  US11469672  US73727395  US11469673  US73729600  US11479739  US03859395  US73723461  US03859566  US03859564  US83233728  US52401976  US84137575  US73729858  US65113533  US00554248  US11474945  US02656518  US91944564  US00552900  US00554053  US03861508  US00555220  US11469292  US02656504  US11478655  US11484838  US53403120  US03861371  US03863786  US90740488  US61306865  US65112445  US90740077  US83232557  US83232554  US53404518  US03861445  US63409484  US00552347  US91944136  US00551875  US91944519  US73726914  US53405035  US91944942  US00555714  US91944953  US91943929  US11481823  US11471968  US00554207  US11480260  US03863638  US63410648  US84137718  US94347380  US03864844  US00553076  US84136594  US03358588  US11489386  US11477736  US94349444  US91945772  US91943939  US03865667  US03859962  US03865075  US00555088  US94348541  US90741565  US94348535  US03863273  US44800125  US84135818  US91944727  US00551925  US65112707  US03866010  US11482402  US61307673  US11470865  US50901234  US11485261  US00552603  US00553234  US63411844  US11476978  US03861036  US03859253  US00553302  US03863242  US73722322  US73722321  US53404849  US11480430  US11491046  US03862691  US11487034  US00553650  US02656838  US03862572  US73729454  US90740924  US03864854  US53402770  US03357499  US11482821  US63409073  US03863149  US73725231  US83231894  US03860718  US03357990  US03866206  US53403800  US00555382  US11476212  US83233696  US84137615  US00551967  US11478929  US00550716  US44800110  US03863816  US73721845  US73723709  US11485854  US03866193  US11467993  US03358273  US11484939  US11483822  US11471412  US03860579  US03861663  US73728957  US00551730  US11475549  US61306871  US03864961  US11484417  US73723315  US65113056  US61306135  US11472629  US02656684  US02656701  US73726206  US03358516  US84138373  US00552011  US73727955  US73731140  US00551847  US11468582  US91944026  US71221032  US73727195  US73728032  US73728036  US11473182  US00554489  US00551915  US03858667  US11468046  US03858728  US03861229  US53402529  US11471714  US11470351  US03865232  US94346659  US65114125  US00556160  US11472825  US53404638  US52402063  US73729653  US91945618  US11473638  US00550925  US61308370  US53404644  US11484107  US00556169  US00553950  US03862247  US65112282  US00551503  US83231931  US11470708  US84137962  US00553520  US73730198  US65112440  US63409839  US11476226  US53402298  US03865743  US11471154  US11469154  US00551844  US00550926  US83233461  US73722299  US71219562  US91942225  US83231966  US11476981  US11476982  US73726573  US11481827  US61308202  US71218007  US91942389  US91946439  US00555950  US00555930  US71219771  US11473079  US61307021  US53402274  US03863702  US03357185  US11469156  US11468153  US11486521  US03866255  US11489740  US00551774  US71218691  US00553903  US00551945  US91945205  US00550808  US91946447  US03863738  US11469904  US84135904  US65113278  US11484616  US61307757  US00555219  US00552009  US83233281  US83233272  US11480462  US11472134  US11471289  US03865936  US00551220  US91943421  US90741966  US65112798  US00551544  US11477313  US03357917  US03863952  US53405926  US11468482  US90741160  US91946530  US63409980  US84137982  US61306537  US53404686  US63409231  US91945143  US11475239  US11481513  US11469316  US50901627  US11485302  US61307932  US03859891  US11477952  US03859809  US02656224  US11471310  US11482169  US63410165  US11477579  US71220513  US94347297  US71216337  US11485243  US03863576  US03860582  US03865775  US00551128  US90741166  US73724530  US52401886  US00553799  US71218151  US90741598  US91943466  US11472157  US11472803  US91942371  US83232884  US61307591  US53404424  US61306999  US71215546  US71215543  US90741170  US63410134  US11166934  US84136628  US00552626  US11470468  US03859728  US00555288  US00552065  US63409935  US11167233  US73724684  US11472700  US50901600  US73729321  US90741131  US73726985  US11482830  US00551900  US63411894  US00551882  US02656869  US11485757  US11481325  US84138313  US00553532  US11480787  US11166375  US03858612  US03859878  US11480444  US94347545  US73727336  US73727345  US11470920  US03357777  US94347392  US90741946  US90741904  US90741220  US91942816  US73724540  US11471761  US00554641  US91944795  US63410416  US11482046  US73729138  US03859520  US84135056  US11472488  US00553275  US73726726  US65112615  US65112515  US73727908  US11473751  US11481679  US91944223  US63410942  US00552045  US03859332  US03859350  US03358309  US73724536  US11486033  US11471261  US02657049  US00551893  US61307380  US63411590  US53404225  US03861073  US73723580  US71216900  US11481921  US03864912  US53406058  US11483606  US83231632  US73725241  US11485385  US03860080  US11475492  US03862935  US00551841  US03859907  US11473911  US73725466  US03865715  US94349705  US94349700  US03357970  US00550846  US71214218  US11473590  US00551578  US00554133  US11471767  US00552975  US00552914  US03862409  US00555802  US11471719  US03861236  US11482883  US71219676  US03866019  US53404173  US73723749  US53403726  US00552250  US73721583  US84137741  US11467695  US00556015  US11482590  US11484688  US11472147  US63409494  US03358216  US00551170  US11481138  US71218296  US84134886  US84134903  US73725450  US11486154  US11480601  US84138513  US63409656  US11484517  US63410350  US11482161  US11469330  US03358228  US11472858  US00552500  US71217482  US00552217  US73727466  US00551242  US11476776  US53404383  US03864022  US00550811  US00553840  US03866143  US00551514  US00552482  US00554359  US71219690  US00551850  US11478623  US11470697  US00551843  US02656217  US61306878  US91944649  US03863835  US53403545  US90740658  US03863271  US11481158  US61306762  US03864193  US63410466  US11485321  US03860649  US11485016  US91943797  US03864734  US11478592  US84136077  US11472115  US11479796  US11482119  US03864076  US90740635  US03860959  US63409523  US00556005  US11471546  US03859562  US03860809  US83234524  US00553140  US84136939  US61308173  US61307865  US03863307  US03863475  US00553238  US00555089  US73728307  US61306798  US00552870  US90739166  US61307671  US61308232  US11470749  US84138228  US00550474  US03865892  US53404499  US73721603  US73722989  US03858622  US03860016  US11483984  US11478803  US11478228  US11471598  US73725537  US53405221  US73722328  US00554651  US91942635  US73730938  US11479427  US94349392  US11480681  US11166578  US00553195  US11476984  US73724722  US03865679  US00555450  US11471760  US50901271  US00555739  US91943131  US00554257  US52401939  US11467752  US00550850  US03860030  US83233712  US71214757  US61308223  US83233101  US11476977  US94348791  US90739991  US11474883  US03859364  US91943524  US11468761  US11485155  US11480156  US00553202  US71216196  US71216348  US00555906  US11482177  US00552053  US11484652  US90439067  US11484438  US03358217  US11483324  US63409056  US73723990  US00552013  US11481884  US11480603  US03860287  US03860279  US73731028  US71214538  US71214540  US11473296  US11486985  US11481702  US73725029  US11469513  US63411747  US63409487  US03860990  US00551939  US03357982  US53405423  US11478280  US11478916  US03861461  US03860576  US00553752  US73724538  US63410463  US00550961  US00551105  US11473176  US11475082  US90740856  US03862855  US94346617  US03866170  US94348159  US11490470  US84134600  US61306898  US03866244  US53405247  US71219593  US90739301  US11482796  US65113705  US03864074  US11166680  US91945585  US53405479  US73723747  US91944730  US03865777  US00550661  US11485317  US03861895  US03865313  US83233976  US83233975  US11472078  US11478387  US03358579  US71215549  US11167018  US11467398  US73728346  US73726261  US65112385  US03865179  US00550225  US03864608  US00554251  US65112937  US73721849  US00553375  US11166457  US94349775  US94349824  US71217472  US73723468  US53405057  US63409406  US11478812  US83234374  US63410465  US03860646  US11467546  US61307407  US90438992  US11476048  US90740096  US84137900  US11483278  US83233791  US73726729  US03862004  US63409427  US71217121  US65113733  US11468590  US63408697  US73723259  US94347134  US71220769  US90438840  US83233751  US63411548  US11482452  US03858628  US03357186  US11469318  US91943792  US11471766  US53403152  US11481825  US53405122  US11485381  US73731018  US53403633  US11489503  US00553103  US73722592  US90439075  US73730228  US11468578  US71221116  US91945445  US71218203  US83233112  US11473151  US00555928  US03860052  US11476354  US84135259  US11468198  US00551067  US11477945  US03859981  US63409496  US11472323  US63411249  US83233735  US00553977  US11468521  US53404777  US03860466  US94348276  US03861996  US94349002  US94348116  US91943701  US03357226  US73722590  US11474198  US94347815  US03859820  US03861555  US11480613  US03862040  US73727508  US84137359  US00553096  US11479554  US11482154  US11485541  US94347121  US02656283  US71220765  US71215134  US03357768  US00551862  US63409428  US11478181  US00554436  US91944742  US11482397  US11166936  US00552675  US94349224  US11473657  US83234224  US11477107  US03863353  US44800350  US94347382  US11479623  US91944740  US91944736  US53402933  US11471765  US94348887  US11485390  US11475789  US00552296  US94346971  US63410623  US63410621",
      "center_classification_date": "20200422",
      "distribution_pattern": "The products were distributed US nationwide.  Global distribution.",
      "state": "MA",
      "product_description": "TRx4841A 1.4 GHz IntelliVue Tele TRX, Model Number 862439;  Includes System   Codes 453564007261, 453564007271, 453564007281, 989803196951",
      "report_date": "20210203",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips North  America, LLC",
      "recall_number": "Z-1767-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "85241",
      "termination_date": "20210211",
      "more_code_info": "",
      "recall_initiation_date": "20200305",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}