  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Rd",
      "reason_for_recall": "The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.",
      "address_2": "",
      "product_quantity": "21,672",
      "code_info": "Serial Numbers:  US00609707  US00609806  US00609799  US03912118  US94208614  US94208908  US02711406  US94208759  US03916432  US00610259  US03912755  US92008041  US84205167  US03914259  US90806906  US73802986  US03914440  US70901588  US03924349  US03916495  US94209012  US03921163  US03916102  US03918561  US70901531  US03917998  US03916097  US03921176  US03916272  US03918699  US03917268  US92007828  US92007308  US73804950  US73803082  US73802366  US73803410  US03916489  US03917066  US00610411  US03921981  US03913216  US03921046  US03915031  US02711384  US73803628  US03916003  US03914228  US03924137  US03918134  US92008416  US92008297  US70901861  US73803331  US73802704  US84205165  US03915683  US03917164  US03924434  US73803177  US70901697  US92007783  US00609787  US73802951  US73803101  US84205510  US03917261  US03918163  US73802473  US92007700  US03915528  US03919990  US03912545  US03923181  US03919097  US03913306  US73802943  US03918021  US03919848  US02711041  US73804777  US03912514  US03914237  US03912387  US03918469  US90506272  US03921778  US92008308  US02711172  US84205636  US02711200  US03922626  US73802999  US70901679  US03913845  US03913840  US03914512  US73802057  US73802023  US03921834  US03917937  US03913836  US03919098  US92007303  US92007299  US73803355  US03918859  US03915028  US03912536  US73804868  US73804877  US03914601  US03914718  US73804849  US03920179  US02711671  US84205604  US73803885  US92007514  US03915312  US73803872  US03914400  US00610073  US03921800  US03921161  US73802676  US73802860  US03912978  US00609816  US92007307  US03920160  US03920164  US02711604  US03921167  US03921851  US73804308  US73804983  US84205563  US70901608  US73804593  US90806413  US73803785  US03913696  US00610607  US03920760  US03917340  US73804101  US03913540  US03916590  US03913566  US00610692  US00610696  US92007269  US03924465  US03918565  US03913486  US70901458  US73803076  US73804734  US03919980  US03920690  US03921549  US94208633  US02410995  US03918258  US73803762  US92008228  US03917785  US94208699  US00609861  US03921625  US02711356  US03918788  US03921435  US03924361  US03919768  US03914833  US03921590  US03918342  US03919801  US03922171  US00609863  US03921327  US03922170  US03914842  US03923661  US03912184  US92007450  US70901685  US03924429  US00609882  US73803111  US73804965  US03917494  US00610661  US03920105  US03924065  US73804094  US90807034  US03917959  US03916092  US73802761  US02711207  US73803242  US03914253  US73802850  US00609580  US00609576  US03916064  US73803046  US00610665  US73804900  US03917409  US03920266  US03921817  US02411013  US02711119  US03924370  US03917500  US03921152  US03924215  US02711217  US00609671  US73802444  US00610533  US03915783  US02711523  US03921500  US03921980  US03922407  US90806443  US73802570  US03914523  US03920753  US02711281  US03914892  US84205043  US03920773  US03919799  US70901191  US03912788  US02711215  US84205237  US03917166  US00609875  US03923922  US00609870  US00609872  US03913313  US73804061  US03922883  US00610797  US73803904  US03912428  US73804985  US03914171  US73802182  US03913381  US73803176  US73802773  US92007660  US03920558  US73802729  US73802398  US73803146  US73803602  US73804722  US92008163  US70901212  US92007241  US03916035  US73802745  US92007499  US73804866  US03920255  US03922767  US70901955  US92008019  US03920157  US03917173  US90506264  US03914898  US03912517  US03920666  US73802751  US03922781  US03920441  US92007054  US03923178  US94208946  US03915074  US92007786  US03920173  US94209123  US00610163  US92008259  US03915889  US84206011  US92008247  US92007596  US92007592  US03920187  US03921164  US03919102  US03920186  US03921169  US73804210  US03914131  US84205371  US73803382  US94209428  US90806701  US03914953  US02711347  US03924097  US03921761  US73803387  US03916648  US03917308  US92008254  US03917700  US03924428  US92008091  US92008094  US02410999  US03921793  US00610508  US03918863  US03921154  US73803190  US92007266  US03921546  US94209122  US92008100  US03924371  US03918862  US03919447  US92008162  US02711167  US03917503  US03924431  US84205744  US03924437  US03916091  US03919995  US03912297  US03922899  US73802717  US92007358  US73803600  US03915499  US03913323  US73802472  US84205943  US73804238  US03912188  US02410880  US03913628  US03916463  US03924114  US03924201  US03912094  US03912457  US02711535  US92007393  US03918139  US02711665  US84205889  US92007192  US92007196  US94209042  US84206131  US03912845  US03915320  US03912607  US03913722  US03917407  US03915117  US73802193  US03916439  US00610162  US03916455  US03923407  US90806286  US03917085  US02410816  US02711577  US84205655  US03914339  US03917685  US94208869  US02711121  US03921547  US00609653  US03914431  US73802156  US84205407  US94208427  US00610581  US03921771  US03914929  US03912771  US03915032  US92007606  US73802093  US03911749  US92007491  US73802161  US03912843  US70902000  US03913627  US03917161  US73804714  US00610019  US03914951  US73803210  US73802857  US03918520  US03920159  US70901084  US92007181  US03914821  US73802341  US03913738  US03920755  US92008193  US03914781  US00609946  US70901554  US73803635  US03914158  US94209110  US70901622  US90806939  US73803604  US02711404  US02711405  US02711349  US03920023  US73803124  US73803522  US73803495  US73803591  US84205302  US03920158  US03913993  US00609993  US03916460  US03924199  US03920386  US00610258  US94208563  US92007132  US03917882  US84205322  US03918623  US73803127  US84206113  US03923163  US92007668  US00609623  US03919676  US03924163  US03913320  US03922769  US70901494  US03915703  US03924418  US02711615  US73804063  US03914257  US73803114  US03919641  US90806333  US00610793  US03913736  US03921166  US03922744  US94209044  US03920188  US00609932  US00609926  US92007173  US73803733  US94208827  US03923817  US03921781  US73803880  US03922750  US03918272  US03912758  US92007062  US03914219  US03914218  US73804846  US02711681  US02711683  US03912046  US90806965  US90806976  US03912039  US84205249  US03913143  US03912977  US03912430  US03913893  US70901992  US70901793  US73803856  US73802561  US73803544  US73802089  US03922764  US03917757  US73803195  US94208786  US92007801  US73802664  US00610738  US94209309  US03914029  US03918737  US03921542  US73802698  US70901129  US70901124  US03918509  US03913269  US94209043  US03917490  US90806313  US73804453  US03923861  US03917425  US03911966  US73804783  US94208482  US73802890  US03917238  US94209260  US03924298  US94209262  US03923952  US03914281  US03917255  US03919067  US03923883  US03921343  US03918645  US03917658  US03920134  US03923543  US03924025  US03919643  US03918630  US03923886  US94209243  US84205580  US03912182  US00610742  US84205864  US03916461  US03914034  US03920819  US03920565  US03917659  US03921760  US03914836  US94209268  US00610433  US03924203  US03918107  US03912786  US73802582  US03921769  US73802598  US02711209  US90807026  US73804167  US03924455  US00609760  US03921047  US94209453  US03920618  US00609604  US03921061  US03915416  US73802277  US73804303  US73804300  US03922326  US84205395  US92007263  US03923808  US03915360  US03912477  US03916850  US03919023  US94209285  US03917398  US03917055  US03917259  US03920656  US03916168  US03922010  US03915075  US92007077  US03924172  US92007139  US03916516  US70901163  US03916506  US03917497  US03913997  US90806985  US03917179  US84205336  US73802273  US02711579  US02411021  US03920503  US03914788  US00609994  US73802491  US03913228  US73802591  US03917908  US00610645  US03913954  US03912965  US94208726  US03924247  US03918288  US03917930  US03919784  US03920226  US94209176  US90806987  US03917495  US70901433  US00610003  US70901288  US03917163  US03917408  US73804833  US03916193  US03913917  US03913346  US03913978  US03912834  US92007143  US03919280  US03917468  US03920163  US03921232  US03919110  US03919715  US03914671  US70901109  US73802050  US03916093  US03920174  US03912483  US03916464  US92008217  US03916557  US03919203  US03924286  US73802515  US03917420  US92008115  US73802584  US03919956  US03921611  US73804281  US03921803  US03914472  US03923165  US03924171  US03924148  US90806543  US00609862  US03920224  US02711129  US00610164  US73803765  US03917550  US03922264  US94209115  US94209114  US03923965  US73804915  US03914388  US03917496  US02711166  US73804824  US03917402  US03921776  US73803528  US73804086  US03920849  US03917690  US03912187  US03917445  US90806459  US70901972  US00610260  US70901262  US02711708  US03919096  US73802163  US03913622  US03921359  US03924381  US03920316  US03917410  US00610690  US03913619  US03912804  US03916444  US73803584  US03913305  US73802416  US03913399  US00610601  US00610558  US92008244  US03914957  US03912529  US73803188  US84205178  US03916431  US03918177  US03915907  US03917554  US84205012  US03914376  US94208984  US02711319  US03912185  US03924047  US03919471  US73804209  US03916275  US73802390  US92007282  US03912186  US70901477  US03916430  US84205248  US03919100  US03924294  US03924083  US00610643  US92007072  US94209436  US94209189  US73804318  US03919439  US73802853  US73803873  US03915251  US70901497  US70901403  US73804332  US03921093  US03924216  US73803757  US03915982  US90806323  US73804790  US03919908  US03911874  US00609895  US03923976  US03919992  US70901255  US03913631  US03919907  US03915876  US03912600  US03922027  US94208894  US03912241  US03913982  US73802572  US73803044  US03914133  US03924318  US03916041  US94208895  US03912066  US03924139  US90806327  US02410951  US03921104  US03919448  US03917542  US03917280  US03917281  US03912475  US03921847  US03924285  US03920133  US73803466  US92008154  US03914529  US03921598  US03924258  US03912509  US03918246  US03914126  US73802998  US94208530  US03913304  US70901637  US03923517  US03920473  US02711442  US03915029  US03914025  US94208642  US03922254  US92007184  US84205933  US73803821  US73804391  US73804355  US03912957  US92007105  US03920325  US73802514  US92008122  US03918280  US03918596  US03918586  US70901291  US03917220  US03913461  US03921362  US84205166  US03920070  US03923871  US03915578  US03915575  US70901523  US03913188  US03914446  US73804140  US03917345  US03916830  US03924291  US03924068  US03921720  US03914882  US03917056  US03919429  US03921806  US03912292  US03920829  US03920933  US00610254  US00609723  US73802108  US03912272  US73802372  US03915316  US03915954  US92008063  US73803080  US03924072  US73804222  US00609658  US03916834  US03922366  US03918518  US03916197  US02711146  US84205099  US03915643  US03913408  US02711298  US73804064  US03919443  US03913791  US03912049  US94208806  US90806790  US03912268  US92007381  US03921645  US90806930  US90806944  US90806893  US70901774  US00610120  US73802088  US90506230  US03913626  US03924050  US73802424  US03912617  US03922409  US73802087  US00610005  US92007520  US03915754  US73803761  US92007516  US00609867  US00609584  US03922779  US73804905  US03912967  US94208886  US00609669  US03924134  US03916341  US94208549  US03914275  US03922634  US00610779  US70901691  US70901689  US00609990  US03924229  US70901499  US03919958  US73803632  US03915199  US03913608  US70901906  US94209002  US03917047  US03912347  US00609789  US03915314  US73804053  US73803658  US73803681  US02711400  US73804975  US94208816  US92008369  US92008124  US00610101  US03921633  US03914994  US00610100  US84206054  US90806876  US84206090  US03921763  US90806981  US94208710  US03917693  US03923797  US03913268  US00610156  US03914087  US00610174  US03921447  US02711293  US90806345  US03915077  US73802562  US73802560  US03924198  US03919504  US00610565  US03913433  US03916179  US70901092  US73803651  US03914778  US03913823  US03924013  US00610454  US00610460  US03912822  US03917304  US03921363  US03917250  US03919105  US94209297  US03920233  US73802371  US03911942  US90806786  US03922030  US73802040  US90806620  US03920823  US92008392  US03919991  US03924061  US03913440  US03917688  US03916965  US92008134  US92008131  US03921618  US94209424  US94209425  US00309534  US03919765  US03915073  US92008374  US03913555  US03913224  US03921788  US90506237  US03924211  US70901304  US84205290  US03920633  US73803843  US03914258  US84205783  US00610353  US94209251  US73804944  US73804725  US03917087  US92007742  US92007748  US92007761  US73802197  US03916664  US03916049  US03924019  US00609734  US03918199  US84205183  US90806807  US03915500  US00610056  US03912841  US92007464  US73803906  US03912192  US03922110  US03916398  US00610307  US03919812  US73804650  US03917174  US73803741  US94209030  US70901888  US03920797  US03913385  US92007995  US03912555  US84205023  US03913693  US73802610  US73803773  US73803782  US73803912  US94208497  US02711402  US73802524  US92007113  US00610098  US02711372  US73804365  US84205686  US00610635  US03912584  US94208818  US73804881  US03915020  US03918289  US03918729  US73802241  US73802217  US03915805  US00610774  US03913959  US03915799  US00610158  US84205754  US03915448  US03924276  US03920192  US94208595  US94208593  US03914873  US03922897  US70901857  US73804448  US03921531  US94208509  US73804415  US00610255  US00609686  US03919510  US03917509  US03919166  US03917390  US03917508  US03915513  US03911739  US03911740  US03915415  US03924056  US03919986  US03915156  US03915177  US90806395  US73804588  US03921301  US03922012  US73802990  US03916458  US92007119  US92007118  US70901668  US70901507  US03924133  US03924136  US92008275  US03913139  US92007638  US94208701  US02711271  US70901688  US03924020  US03923149  US94208822  US73804795  US73802600  US92007136  US92007135  US03922762  US03916844  US70901615  US70901564  US03913458  US00610141  US00610591  US03912800  US03924453  US73804073  US94209239  US73804353  US73804339  US03918658  US73802674  US03912017  US03914862  US03914222  US03917198  US03924334  US03916748  US03916956  US03922770  US94208700  US03921504  US02711608  US03918513  US00610739  US03919988  US73802503  US02411004  US73804004  US02711629  US02711291  US03916414  US03924205  US03920331  US03923862  US03919027  US73803000  US03916999  US03921147  US03917734  US03914462  US73803569  US73803474  US03914378  US03924076  US03917726  US03921790  US02711036  US03920121  US03920034  US90806707  US03921173  US94209011  US90806503  US03915214  US73802208  US84205343  US92007227  US02411016  US03924338  US90506182  US03915947  US92007259  US03920180  US73803156  US00610267  US00610422  US00610421  US03922028  US73804257  US03922404  US84205358  US73803139  US84205666  US03914436  US03921796  US94209088  US92007797  US03919464  US03917172  US94208829  US03915282  US03921149  US70901909  US03917670  US00610237  US00610242  US73802368  US73802386  US92007158  US73802022  US00609996  US03915272  US03913809  US73802862  US03923662  US03912776  US90806317  US03912125  US70901323  US70901659  US73803745  US03916082  US02711563  US03921653  US03924064  US73802449  US84206056  US03919767  US03921819  US03913600  US70901491  US03912981  US73804037  US03924315  US00610677  US03911905  US03922402  US03912748  US02711697  US73802685  US84205272  US03919473  US90806730  US03916225  US03920610  US03922016  US92008379  US70901360  US03920762  US03918220  US92007312  US03918609  US92007824  US03922333  US03921993  US03917270  US84205687  US73804538  US03924449  US03920502  US03914711  US70901399  US03919735  US84205235  US73802298  US73804397  US03921747  US94209338  US90806700  US03915424  US70901322  US03911867  US70901402  US90806714  US03921833  US84205271  US92007134  US70901487  US03913625  US73803612  US03923168  US03921777  US73802173  US03922022  US03914156  US00610517  US94208538  US03917067  US70901187  US94209027  US84206058  US03916738  US03914921  US00610505  US03913344  US94209292  US03914254  US03913971  US02711664  US84206035  US73803149  US92007082  US73802461  US03920341  US00609547  US03914552  US03924270  US90806511  US92007967  US92007968  US92008408  US03915403  US03913293  US02711537  US84205381  US84206003  US73803073  US03918987  US03917177  US73802170  US94208476  US92007362  US92007364  US00610644  US70901985  US03916600  US00309543  US90506185  US03920794  US90506216  US03912116  US90806742  US73803369  US94209028  US03913983  US03915612  US84205518  US92007658  US00610282  US03924452  US00610449  US03919461  US03919884  US73802098  US84205201  US03915529  US03917929  US70901461  US92008182  US03913623  US94208797  US03923960  US00610455  US73803180  US03917167  US03924289  US94209280  US00309525  US73803614  US03917404  US73804179  US73804678  US90806361  US03911945  US03918670  US90506222  US94209086  US73804781  US92008264  US03917343  US92007042  US90806507  US03922893  US03916870  US03923869  US03921616  US92008279  US03913161  US03911815  US70901902  US03915437  US73802653  US73802652  US03924063  US03917450  US00610764  US03924077  US90806949  US03920008  US73802055  US94209185  US03921585  US03921150  US03912854  US03924107  US70901064  US03924112  US03924101  US84205104  US73802670  US90806982  US73802666  US03913171  US73803866  US73802808  US94208693  US94208695  US94208877  US03921768  US70901112  US03915535  US70901050  US73803555  US03922267  US03915392  US03924240  US00610741  US03920227  US03913690  US00309536  US73802536  US90806480  US90806470  US03918075  US73803166  US03913138  US03914038  US03915546  US73804621  US73804627  US03921159  US73802438  US03923933  US03916061  US90806397  US03917403  US90806368  US90806365  US03918869  US03918868  US84205342  US73804854  US94208567  US03920340  US94209052  US90806408  US03914611  US92008191  US03920440  US03912309  US73803996  US92007488  US00610474  US94208896  US03915634  US92008334  US03920702  US03913107  US92008241  US73803515  US73803647  US70901714  US73802013  US03923565  US03919281  US03919475  US03919681  US03920860  US03924312  US03917076  US03921032  US03920039  US03914963  US03921532  US00610634  US03919251  US03919244  US03924414  US03912418  US03921189  US03914159  US03916418  US84205867  US70901425  US92007621  US92007618  US03916954  US03919279  US03917915  US03914752  US03920800  US03921361  US03915878  US02711564  US03919282  US03919286  US03914375  US03919172  US03913908  US03914912  US03919331  US03921380  US03923544  US00610092  US03917264  US03916568  US03916567  US92008128  US03920763  US84205031  US03920796  US03920176  US03912837  US70901960  US73803902  US73804787  US03922766  US03911873  US03921670  US73802708  US03915555  US70901294  US84206019  US03913491  US03915508  US84206101  US84206110  US00610441  US03916051  US73802430  US03922330  US03921160  US00609675  US70901315  US70901393  US03916167  US00610560  US03912780  US73803578  US92008185  US03913185  US03917691  US84205887  US73802719  US03924110  US03914649  US94208890  US03924271  US73804823  US03918644  US90806530  US73804793  US92007732  US70901266  US03917312  US92007253  US03912183  US73802451  US03916098  US03913689  US03917662  US73803794  US92007359  US70901215  US03914447  US03923541  US73802455  US03922406  US70901467  US84205931  US92007978  US03924200  US73802118  US73804888  US90806437  US03924218  US73803389  US73803863  US90806324  US70901282  US03919230  US03914101  US70901279  US90806929  US90806386  US73804013  US00309520  US03914809  US03914384  US03922322  US03920792  US73803107  US03914316  US03914310  US03914311  US03921801  US03924156  US73802330  US92008136  US90806963  US03921767  US73804227  US73803541  US73803925  US94208623  US94208617  US92008214  US92008213  US92008310  US03915350  US03920161  US73803690  US03921157  US03924436  US94208789  US84205420  US03914958  US03921762  US03912368  US03921758  US94209215  US92007378  US03921779  US00610215  US02711367  US03922981  US03923705  US90806289  US84206142  US73802296  US73803650  US73803937  US03920889  US03923160  US73804162  US03916281  US03920167  US03914804  US03924253  US03924333  US92008048  US84205511  US70901080  US03915652  US03919033  US03914264  US03919018  US90806504  US03924197  US03917322  US03918709  US84205089  US73803224  US90806377  US70901076  US73803042  US03915324  US03915440  US03923065  US03917776  US03922113  US02711211  US00610590  US03916200  US03916198  US90806314  US73802403  US03912398  US90807032  US84205987  US03915323  US70902006  US03924269  US73802164  US03922900  US84205119  US03920156  US03922025  US84205681  US03920438  US73803516  US00610585  US73803009  US03917342  US03917262  US03921964  US90806404  US73802127  US84205261  US70901754  US70901892  US02711296  US90806988  US03912100  US03915633  US03923170  US03914134  US03919999  US03922394  US73803893  US03913641  US84205719  US03916826  US03919663  US90806684  US84205726  US03919103  US00609948  US03913629  US73802988  US03917175  US03921596  US03917025  US03917180  US84205502  US03914819  US70901759  US03921765  US03924339  US03921099  US03917492  US03924105  US73803214  US73802039  US73802575  US84205069  US70901072  US00610275  US94208938  US92007826  US03915887  US73802716  US03914173  US92007267  US03924259  US03924239  US03920000  US84205728  US03917162  US84205495  US03913088  US90806429  US00610263  US90806980  US70901150  US03915197  US03922197  US73802211  US03920566  US73803700  US90806873  US03913343  US92008262  US03912190  US73804949  US03921165  US03918618  US02711241  US92008340  US03912395  US70901542  US94208639  US03915653  US90806467  US03916037  US90806461  US90806791  US90806801  US03913766  US70901709  US70901583  US03924144  US00610589  US70901411  US73804363  US73804940  US03915318  US73803508  US03914823  US03913621  US03921545  US73804399  US84206084  US94208764  US03917194  US92007777  US03913929  US03917505  US03923146  US03922760  US70901964  US03916459  US03920168  US73802709  US73803178  US84205779  US92008252  US03919431  US92008249  US03912727  US03920172  US94208650  US03915270  US70901025  US03920585  US03918042  US03920268  US03917912  US92008327  US03918230  US03918237  US03924301  US03917019  US03919815  US03920382  US03915325  US03916833  US03917943  US03917963  US03921569  US03923847  US03914256  US03920783  US03917171  US03923220  US94208504  US03916889",
      "center_classification_date": "20200422",
      "distribution_pattern": "The products were distributed US nationwide.  Global distribution.",
      "state": "MA",
      "product_description": "TRx4851A 2.4 GHz IntelliVue Tele TRX, Model Number 862231; Includes Service Numbers 453564052401, 453564052411, 453564052441, 453564052451, 453564166851, 453564166861",
      "report_date": "20210203",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips North  America, LLC",
      "recall_number": "Z-1766-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "85241",
      "termination_date": "20210211",
      "more_code_info": "",
      "recall_initiation_date": "20200305",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}