{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82964",
      "recalling_firm": "Teleflex Medical",
      "address_1": "3015 Carrington Mill Blvd",
      "address_2": "N/A",
      "postal_code": "27560-5437",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide - US Nationwide Distribution",
      "recall_number": "Z-1766-2019",
      "product_description": "Sheridan/HVT:  1) Novaplus 7.0 mm, Product Code V5-10314  2) Novaplus 7.5 mm, Product Code V5-10315  3) Novaplus 8.0 mm, Product Code V5-10316  4) Novaplus 8.5 mm, Product Code V5-10317    Product Usage:  Tracheal tube/airway management",
      "product_quantity": "1390944 units",
      "reason_for_recall": "This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.",
      "recall_initiation_date": "20190524",
      "center_classification_date": "20190619",
      "termination_date": "20220215",
      "report_date": "20190626",
      "code_info": "1) Novaplus 7.0 mm, Product Code V5-10314  Lot Numbers:  73J1600387  73J1600388  73K1600157  73K1600603  73K1600604  73K1600736  73L1600367  73L1600368  73L1600502  73L1600870  73M1600165  73M1600270  73A1700121  73A1700122  73B1700322  73B1700428  73C1700426  73C1700576  73C1700577  73C1700702  73C1700703  73E1700106  73E1700107  73E1700108  73E1700239  73F1700560  73G1700147  73G1700148  73G1700302  73G1700303  73G1700623  73G1700624  73H1700055  73H1700497  73H1700568  73H1700628  73J1700309  73K1700086  73K1700087  73K1700616  73A1800296  73A1800430  73B1800218  73B1800372  73B1800373  73B1800374  73B1800512  73B1800513  73C1800373  73C1800574  73C1800575  73D1800656  73D1800657  73E1800580  73E1800581  73E1800809  73F1800723  73G1800107  73H1800205  73H1800456  73H1800457  73H1800759  73H1800760  73J1800120  73J1800534  73J1800697    2) Novaplus 7.5 mm, Product Code V5-10315  Lot Numbers:  73J1600389  73J1600580  73J1600581  73K1600158  73L1600260  73L1600518  73L1600672  73M1600268  73A1700289  73A1700290  73A1700741  73B1700165  73B1700638  73C1700265  73D1700137  73D1700138  73D1700266  73D1700267  73E1700240  73F1700281  73F1700365  73F1700414  73G1700237  73G1700437  73H1700056  73H1700352  73H1700629  73H1700796  73J1700443  73K1700435  73K1700436  73K1700754  73L1700672  73L1700673  73A1800687  73A1800795  73B1800514  73D1800158  73D1800286  73E1800098  73E1800309  73E1800312  73E1800407  73E1800408  73F1800191  73F1800192  73F1800362  73F1800363  73G1800595  73G1800831  73H1800078  73H1800688  73H1800761  73J1800327  73K1800322    3) Novaplus 8.0 mm, Product Code V5-10316  Lot Numbers:  73J1600582  73J1600772  73K1600605  73L1600174  73L1600369  73L1600370  73L1600673  73L1600871  73A1700428  73A1700661  73B1700625  73D1700430  73D1700431  73D1700432  73D1700631  73F1700038  73F1700039  73F1700734  73G1700438  73G1700625  73H1700353  73H1700499  73H1700797  73J1700144  73J1700493  73J1700610  73K1700617  73M1700150  73A1800070  73B1800728  73B1800729  73C1800205  73C1800207  73D1800291  73D1800483  73D1800484  73E1800810  73G1800234  73G1800596  73H1800458  73J1800334  73J1800698  73K1800195    4) Novaplus 8.5 mm, Product Code V5-10317  Lot Numbers:  73L1600674  73B1700639  73C1700704  73D1700632  73F1700561  73G1700439  73H1700354  73M1700025  73B1800730  73F1800193  73G1800019  73H1800093  73K1800574"
    }
  ]
}