{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85448",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-1764-2020",
      "product_description": "System AXIOM Aristos FX - Product Usage: A dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems. The MPRS allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, extremities, and excluding mammography. Radiographic exposures may be taken with the patient in sitting, standing, or supine positions.",
      "product_quantity": "314 (wordwide)",
      "reason_for_recall": "It was determined that in rare cases persons can receive low dose of unintended radiation.  The system performed exposures without the exposure button being pressed. Exposure will be released for less than 500ms, therefore, the overall dose is estimated at 0.01mSv or less than 500 mGy in total.",
      "recall_initiation_date": "20200406",
      "center_classification_date": "20200505",
      "termination_date": "20201027",
      "report_date": "20200513",
      "code_info": "serail number 1192",
      "more_code_info": ""
    }
  ]
}