{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82964",
      "recalling_firm": "Teleflex Medical",
      "address_1": "3015 Carrington Mill Blvd",
      "address_2": "N/A",
      "postal_code": "27560-5437",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide - US Nationwide Distribution",
      "recall_number": "Z-1763-2019",
      "product_description": "Hudson RCI Sheridan:  1) Uncuffed 6.0 mm, Product Code 5-10412  2) Uncuffed 6.5 mm, Product Code 5-10413  3) Uncuffed 7.0 mm, Product Code 5-10414    Product Usage:  Tracheal tube/airway management",
      "product_quantity": "14357",
      "reason_for_recall": "This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.",
      "recall_initiation_date": "20190524",
      "center_classification_date": "20190619",
      "termination_date": "20220215",
      "report_date": "20190626",
      "code_info": "1) Uncuffed 6.0 mm, Product Code 5-10412  Lot Numbers:  73J1600363  73K1600581  73L1600347  73L1600552  73L1600850  73M1600138  73A1700264  73B1700300  73C1700688  73E1700098  73E1700712  73G1700132  73H1700327  73J1700290  73K1700065  73K1700415  73L1700169  73L1700320  73M1700250  73A1800060  73B1800348  73B1800487  73C1800349  73C1800482  73D1800464  73E1800389  73F1800171  73G1800923    2) Uncuffed 6.5 mm, Product Code 5-10413  Lot Numbers:  73M1600139  73B1700301  73E1700472  73G1700127  73K1700416  73A1800061  73B1800349  73B1800488  73D1800634  73F1800544  73G1800924    3) Uncuffed 7.0 mm, Product Code 5-10414  Lot Numbers:  73A1700100  73F1700064  73L1800406"
    }
  ]
}