  "results": [
    {
      "status": "Terminated",
      "city": "Utica",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76482",
      "recalling_firm": "ConMed Corporation",
      "address_1": "525 French Rd",
      "address_2": "N/A",
      "postal_code": "13502-5945",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) Internationally to Australia, Canada, Germany, Italy, UK, Great Britain, and Ireland.",
      "recall_number": "Z-1760-2017",
      "product_description": "HYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED CORPORATION, 525 FRENCH ROAD UTICA, NY 13502. --- There is no problem with the Hyfrecator 2000 Electrosurgical Units or with cases of P/N 7-796-19 sold in cases as P/N 7-796-19BX or P/N 7-796-19CS.",
      "product_quantity": "N/A",
      "reason_for_recall": "For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile.  This could cause an end user to use a non-sterile sheath as if it were Sterile.",
      "recall_initiation_date": "20170228",
      "center_classification_date": "20170407",
      "report_date": "20170419",
      "code_info": "Affected Lot Numbers: 201511234, 201512074, 201512094, 201601064, 201603284",