  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82964",
      "recalling_firm": "Teleflex Medical",
      "address_1": "3015 Carrington Mill Blvd",
      "address_2": "N/A",
      "postal_code": "27560-5437",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide - US Nationwide Distribution",
      "recall_number": "Z-1759-2019",
      "product_description": "Hudson RCI Sheridan LTS, Product Code 5-11112    Product Usage:  Tracheal tube/airway management",
      "product_quantity": "3483 units",
      "reason_for_recall": "This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.",
      "recall_initiation_date": "20190524",
      "center_classification_date": "20190619",
      "termination_date": "20220215",
      "report_date": "20190626",
      "code_info": "Lot Numbers:  73K1600369, 73L1600011, 73A1700646, 73C1700238, 73D1700625, 73E1700475, 73F1700387, 73H1700041, 73H1700483, 73J1700292, 73L1700503, 73M1700453, 73A1800671, 73B1800350, 73D1800138, 73F1800174, 73G1800126, 73G1800806, 73J1800308, 73K1800173, 73L1800225, 73K1800172"
    }
  ]
}