  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85289",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Distributed throughout  the United States and to Puerto Rico. States include: CA, NC, OR, MI, GE, AL, SC, TX, NJ, IA, MS, IN, MA, NC, OH, AZ, LA, MA, NY, PA, KY, WV, MI, WA, CO, IL, VA, FL, KS, MD, ID, DE, NV, OK, AR, HI, WI, NY.",
      "recall_number": "Z-1752-2020",
      "product_description": "Opiate 300 Ng Reagent (OP 1 X 250) for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems OP 300 Urine Calibrators, Catalogue No.475024. The OP assay provides a rapid screening procedure for determining the presence of OP and its metabolites in urine.",
      "product_quantity": "7,951 units",
      "reason_for_recall": "Cross-reactivity testing may not meet the claims indicated in the Opiate 300 ng (OP)Instructions For Use. The presence of  certain compounds in urine may result in false positive or false negative patient test results.",
      "recall_initiation_date": "20190904",
      "center_classification_date": "20200420",
      "termination_date": "20210428",
      "report_date": "20200429",
      "code_info": "Catalog Number 475024; UDI:15099590222970 ; Lot Numbers: M904130 and older including  M802097, M803045, M804024, M805123, M806080, M808069, M809218, M810293, M901044, M902086, and M903175.                               ."
    }
  ]
}