  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Winston Salem",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84822",
      "recalling_firm": "Wilson-Cook Medical Inc.",
      "address_1": "4900 Bethania Station Rd",
      "address_2": "N/A",
      "postal_code": "27105-1203",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution.  International distribution to Bahrain, Belarus, Canary Islands, Gibraltar, Guadeloupe, Guernsey, Iraq, Israel, Jersey, Jordan, Kuwait, Lebanon, Martinique, Monaco, Morocco, Oman, Pakistan, Qatar, Reunion, Saudi Arabia, Tunisia, United Arab Emirates, Unites States, Canada, Netherlands Antilles, Chile, Egypt, Japan, Mexico, Peru, Puerto Rico, Trinidad and Tobago, Australia, Brunei Dassalarum, Hong Kong, Korea, Myanmar, Macau, Malaysia, New Zealand, Philippines, Singapore, Thailand, Vietnam, Argentina, Brazil, Austria, Belgium, Bulgaria, Switzerland, Cyprus, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Greece, Hungary, Ireland, Iceland, Italy, Luxembourg, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Sweden, Slovenia, Slovakia, Turkey, Croatia",
      "recall_number": "Z-1745-2020",
      "product_description": "HEMO-7 Hemospray Endoscopic Hemostat",
      "product_quantity": "89,423 devices total",
      "reason_for_recall": "Complaints that the handle and/or activation knob on Hemostat devices have cracked or broken when the device is activated. In some cases, this has caused the carbon dioxide cartridge to exit the handle. Most of these complaints reported that the carbon dioxide cartridge exited the handle with minimal force. However, some complaints reported that the carbon dioxide cartridge exited the handle with the potential to project towards a user or a patient.",
      "recall_initiation_date": "20200204",
      "center_classification_date": "20200417",
      "termination_date": "20210115",
      "report_date": "20200429",
      "code_info": "All Lot numbers manufactured 01/16/2017 - 01/15/2020 of both the following:  HEMO-7 Order Number G56572   HEMO-10 Order Number G21049"
    }
  ]
}