      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67879",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI & WV.  and countries of Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Ecuador, France, French Polynesia, Gabon, Germany, India, Indonesia, Israel, Italy, Jordan, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Zealand, Oman, Palestine, Philippines, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syrian Arab Republic, Thailand, Ukraine &  United Kingdom.",
      "recall_number": "Z-1745-2014",
      "product_description": "Brilliance CT Big Bore Oncology Whole Body Computed Tomography X-Ray System  & Brilliance CT Big Bore Radiology Whole Body Computed Tomography X-Ray System.     Product  Usage:  The Brilliance CT Big Bore is a Whole Body Computed Tomography X-Ray System which produces cross-sectional body images by computer reconstruction of x-ray transmission data taken at different angles and planes.",
      "product_quantity": "454 Units",
      "reason_for_recall": "During internal testing, the Brilliance CT Big Bore was found to be out of tolerance for radio frequency emissions. Specifically, at the 48MHZ frequency, the testing indicated the Brilliance CT Big Bore was 3.5dB uV/meter higher than the applicable IEC 60601-1-2 standard specification.",
      "recall_initiation_date": "20131219",
      "center_classification_date": "20140621",
      "termination_date": "20150825",
      "report_date": "20140702",
      "code_info": "Brilliance CT Big Bore Oncology:  Model #728243; Serial #\u0019s: 7185, 7187-7189, 7191-7199, 7201,-7243, 7245-7251, 7253-7267, 7269, 7270, 7272-7297, 7299, 7300, 7302, 7304-7317, 7319-7374, 7376-7379, 7381, 7382, 7385, 7387-7389, 7391-7393, 7395, 7397-7401, 7403-7406, 7409-7429, 7431-7433, 7435-7438, 7440-7451, 7454-7457, 7459, 7462-7485, 7487-7509, 7511-7513, 7515-7519, 7521-7524, 7526-7528, 7530-7532, 7534-7537, 7539-7542, 7544, 7545, 7547, 7549-7556, 7558-7562, 7564-7566, 7568, 7569, 7573, 7574, 7577-7582, 7584, 7585, 7588-7593, 7596-7599, 7601-7603, 7605, 7608, 7609, 7611-7613, 7615-7618, 7620, 7622-7624, 7626-7630, 7634, 7635, 7637-7648 & 75000.    Brilliance CT Big Bore Radiology:  Model #728244; Serial #\u0019s: 7190, 7200, 7244, 7252, 7268, 7271, 7301, 7318, 7375, 7380, 7383, 7384, 7386, 7390, 7396, 7402, 7407, 7408, 7427, 7430, 7439, 7452, 7453, 7458, 7486, 7510, 7514, 7520, 7525, 7529, 7533, 7538, 7543, 7546, 7557, 7570-7572, 7575, 7576, 7583, 7586, 7587, 7595, 7600, 7604, 7607, 7610, 7614, 7619, 7621, 7625, 7631-7633 & 7636."
    }
  ]
}