      "status": "Terminated",
      "city": "Neuville Sur Oise",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79797",
      "recalling_firm": "Hyphen BioMed",
      "address_1": "155 rue d Eragny",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "OH, MD, NC, NY, NH, Canada, Sweden",
      "recall_number": "Z-1744-2018",
      "product_description": "ZYMUTEST HIA IgGAM; REF Number: RK040D; UDI: 03663537018770 a qualitative screening assay intended for the global detection of heparin- dependent antibodies, whether the isotype is: IgG, IgM and IgA, in human plasma, by clinical laboratories.",
      "product_quantity": "281",
      "reason_for_recall": "Incorrect package insert indicating use of the kits with plasma, serum, and other biological fluids. The kits should only be used with plasma.",
      "recall_initiation_date": "20180205",
      "center_classification_date": "20180508",
      "termination_date": "20180816",
      "report_date": "20180516",
      "code_info": "Lot numbers: F1600568P1, F1600568P2, F1700415P1, F1700415P2, F1701369P1, F1701369P2"
    }
  ]
}