  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Valence Cedex 9",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71365",
      "recalling_firm": "BIOMET FRANCE S.A.R.L.",
      "address_1": "plateau de lautagne",
      "address_2": "valence cedex 9",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution to MD only",
      "recall_number": "Z-1744-2015",
      "product_description": "The Optigun Ratchet is a hand-held cement gun for use with Optivac.  The Optigun Ratchet is used to deliver bone cement from the Optivac.",
      "product_quantity": "3",
      "reason_for_recall": "The pin which maintains the knob button, may disconnect and become lost.  This may result in a delay of surgery to obtain a new Optigun and detect/retrieve the pin, and may necessitate revision surgery if the pin falls into the surgical site and is not detected.",
      "recall_initiation_date": "20150514",
      "center_classification_date": "20150610",
      "termination_date": "20151027",
      "report_date": "20150617",
      "code_info": "Catalog number: 419500  Lot number: 0101296012"
    }
  ]
}