  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84918",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution was made to AL, CA, CO, DE, FL, GA, IL, IN, LA, MA, MD, MI, NC, ND, NJ, NY, OH, OR, PA, SD, TN, TX, and WA.  There was government distribution and no military distribution.     Foreign distribution was made to Argentina, Australia, Canada, China, Czech Republic, Egypt, France, Germany, India, Italy, Republic of Korea, Kuwait, Malaysia, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Taiwan, Turkey, UAE, United Kingdom, and Viet Nam.",
      "recall_number": "Z-1741-2020",
      "product_description": "Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power Express Dynamic Inlet, catalog B36352.",
      "product_quantity": "291 units",
      "reason_for_recall": "Potential exposure to biohazard.    Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing.",
      "recall_initiation_date": "20190517",
      "center_classification_date": "20200417",
      "termination_date": "20220311",
      "report_date": "20200429",
      "code_info": "All serial numbers installed before 5/20/2019:    Serials numbers in the U.S.:   PW211409002, PW211501005, PW211501006, PW211501007, PW211501008, PW211501013, PW211505001, PW211508003, PW211509002, PW211510001, PW211510002, PW211511001, PW211511002, PW211511003, PW211511005, PW211512001, PW211601001, PW211601002, PW211602002, PW211602003, PW211602004, PW211603001, PW211603002, PW211603003, PW211604001, PW211604002, PW211604005, PW211604006, PW211604009, PW211604010, PW211605001, PW211605002, PW211605004, PW211605005, PW211605006, PW211605007, PW211606005, PW211606007, PW211607001, PW211607004, PW211607005, PW211607009, PW211607017, PW211607018, PW211608001, PW211608006, PW211610002, PW211610006, PW211610007, PW211610008, PW211610011, PW211611001, PW211611002, PW211611003, PW211611004, PW211611005, PW211612003, PW211612006, PW211612007, PW211612008, PW211612010, PW211612013, PW211701001, PW211701010, PW211701013, PW211703004, PW211706008, PW211706009, PW211707005, PW211707006, PW211707007, PW211707009, PW211707011, PW211707012, PW211708008, PW211708009, PW211708010, PW211709004, PW211709011, PW211710001, PW211710009, PW211710010, PW211710012, PW211711002, PW211711004, PW211711007, PW211712001, PW211802013, PW211804005, PW211804006, PW211807006, and PW211807008.    Serial numbers OUS: PW201607008, PW211409001, PW211501001, PW211501002, PW211501003, PW211501004, PW211501010, PW211501011, PW211501012, PW211506001, PW211506002, PW211506003, PW211506004, PW211506005, PW211506006, PW211506007, PW211507001, PW211507002, PW211507003, PW211507004, PW211508001, PW211508002, PW211508004, PW211509001, PW211510003, PW211511004, PW211512002, PW211512003, PW211601005, PW21160113, PW21160114, PW211602001,  PW211602005, PW211602006, PW211603004, PW211603005, PW211603006, PW211604003, PW211604004, PW211604007, PW211604008, PW211605008, PW211606001, PW211606002, PW211606003, PW211606004, PW211606006, PW211606008, PW211607002, PW211607003, PW211607006, PW211607007, PW211607010, PW211607011, PW211607012, PW211607013, PW211607014, PW211607015, PW211607016, PW211608002, PW211608003, PW211608004, PW211608005, PW211610003, PW211610004, PW211610005, PW211610009, PW211610010, PW211611006, PW211612001, PW211612002, PW211612005, PW211612009, PW211612012, PW211701002, PW211701003, PW211701004, PW211701005, PW211701006, PW211701007, PW211701008, PW211701011, PW211701012, PW211701014, PW211702001, PW211702002, PW211702003, PW211702004, PW211702005, PW211702006, PW211702007, PW211702008, PW211703001, PW211703002, PW211703003, PW211703005, PW211703006, PW211703007, PW211703008, PW211703009, PW211703010, PW211703011, PW211706001, PW211706002, PW211706003, PW211706004, PW211706005, PW211706006, PW211706007, PW211706010, PW211706011, PW211706012, PW211706013, PW211707001, PW211707002, PW211707003, PW211707004, PW211707010, PW211708001, PW211708002, PW211708003, PW211708004, PW211708005, PW211708006, PW211708007, PW211709001, PW211709002, PW211709003, PW211709005, PW211709006, PW211709007, PW211709009, PW211709010, PW211709012, PW211710002, PW211710003, PW211710004, PW211710005, PW211710006, PW211710007, PW211710008, PW211710011, PW211711003, PW211712002, PW211712003, PW211712004, PW211712005, PW211712006, PW211712007, PW211712008, PW211712009, PW211712010, PW211801001, PW211801002, PW211801005, PW211801006, PW211801007, PW211801008, PW211801009, PW211801010, PW211801011, PW211801012, PW211802001, PW211802002, PW211802003, PW211802004, PW211802005, PW211802008, PW211802009, PW211802012, PW211802014, PW211802015, PW211802016, PW211803001, PW211803002, PW211803003, PW211803004, PW211803005, PW211804001, PW211804002, PW211804003, PW211804004, PW211804010, PW211804011, PW211804012, PW211804013, PW211804014, PW211804015, PW211804016, PW211804017, PW211804018, PW211804019, PW211805001, PW211805004, PW211805005, PW211805006, PW211805007, PW211806007, and PW212612004."
    }
  ]
}