{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Memphis",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71330",
      "recalling_firm": "Smith & Nephew, Inc.",
      "address_1": "1450 E Brooks Rd",
      "address_2": "N/A",
      "postal_code": "38116-1804",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of CA, FL, IL, IN, MI, NE, NY, TN, TX, VA, WI and the countries of  Australia, Switzerland, Chile, Canada, and Spain",
      "recall_number": "Z-1738-2015",
      "product_description": "R3(TM) 28 MM I.D., 60 MM O.D. CONSTRAINED LINER, REF 71339160, QTY: (1), STERILE EO,     Product Usage:  The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained liner is intended for single use only.",
      "product_quantity": "40 units",
      "reason_for_recall": "The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.",
      "recall_initiation_date": "20150506",
      "center_classification_date": "20150609",
      "termination_date": "20220401",
      "report_date": "20150617",
      "code_info": "Lot numbers:  16HM08087, 14HM08097, 14HM08133, 14HM08141, and 14JM21517"
    }
  ]
}