  "results": [
    {
      "status": "Terminated",
      "city": "Beverly",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64989",
      "recalling_firm": "SpineFrontier, Inc.",
      "address_1": "500 Cummings Ctr Ste 3500",
      "address_2": "N/A",
      "postal_code": "01915-6516",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "USA Nationwide Distribution including the state of TX",
      "recall_number": "Z-1737-2013",
      "product_description": "KRD1 PedFuse Rods, Model:  IM80100-13 Rev B    Product Usage:  Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod sits with the pedicle screw tulip and they are locked into place with a set screw or locking cap. The rod may be placed along several pedicle screws. Component rods are used during implantation of pedicle screw system devices intended to provide immobilization and stabilization of spinal segments of skeletally mature patients.",
      "product_quantity": "2 rods",
      "reason_for_recall": "One of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles.",
      "recall_initiation_date": "20111230",
      "center_classification_date": "20130716",
      "termination_date": "20140108",
      "report_date": "20130724",
      "code_info": "Lot Number:  002876-007R"
    }
  ]
}